Metal artifact reduction MRI of total ankle arthroplasty implants

  • Cesar de Cesar Netto
  • Lucas F. Fonseca
  • Benjamin Fritz
  • Steven E. Stern
  • Esther Raithel
  • Mathias Nittka
  • Lew C. Schon
  • Jan Fritz
Musculoskeletal

Abstract

Objectives

To assess high-bandwidth and compressed sensing-(CS)-SEMAC turbo spin echo (TSE) techniques for metal artifact reduction MRI of total ankle arthroplasty (TAA) implants.

Methods

Following institutional approval and consent, 40 subjects with TAA implants underwent 1.5-T MRI prospectively. Evaluations included bone-implant interfaces, anatomical structures, abnormal findings and differential diagnoses before and after MRI. AUCs of P-P plots were used to determine superiority. Statistical differences were evaluated with McNemar and chi-square tests. P-values ≤ 0.05 were considered significant.

Results

CS-SEMAC TSE was superior to high-bandwidth TSE in showing the bone-implant interfaces (AUC=0.917), periprosthetic bone, tendons and joint capsule (AUC=0.337–0.766), bone marrow oedema (43 % difference, p=0.041), interface osteolysis (63 %, p=0.015), tendinopathy (62 %, p=0.062), periprosthetic fractures (60 %, p=0.250), synovitis (43 %, p=0.250), as well as reader confidence for bone marrow oedema (p=<0.001), fracture (p=0.001), interface osteolysis (p=0.003), synovitis (p=0.027) and tendinopathy (p=0.034). The number of differential diagnoses in symptomatic subjects after the MRI with CS-SEMAC decreased from 3 (1–4) to 1 (1–2) (p<0.001).

Conclusions

MRI of TAA implants with CS-SEMAC improves the diagnosis of interface osteolysis, periprosthetic bone marrow oedema, fractures and tendinopathy when compared to high-BW TSE, and has a positive effect on patient management.

Key Points

High-bandwidth TSE and compressed sensing SEMAC improve MRI of ankle arthroplasty implants.

Compressed sensing SEMAC improves bone-implant interfaces, periprosthetic bone, tendons and joint capsule visibility.

Compressed sensing SEMAC improves the diagnosis of osteolysis, tendinopathy, fractures and synovitis.

MRI decreases the number of clinical differential diagnoses of painful ankle arthroplasty implants.

Keywords

Magnetic resonance imaging Ankle Arthroplasty Artifacts Joint prosthesis 

Abbreviations and acronyms

AUC

Area under the curve

BW

Receiver bandwidth

CI

Confidence interval

CS

Compressed sensing

ICC

Intraclass correlation coefficient

IW

Intermediate-weighted

LCD

Liquid crystal display

MAVRIC

Multiacquisition Variable-Resonance Image Combination

MRI

Magnetic resonance imaging

P-P

Percent–percent

SEMAC

Slice Encoding for Metal Artifact Correction

STIR

Short tau inversion recovery

T

Tesla

TAA

Total ankle arthroplasty

TSE

Turbo spin echo

Notes

Acknowledgements

The authors would like to thank Christoph Forman, Jens Wetzl and Michael Zenge for their work on the image reconstruction framework.

Funding

This study has received funding by Siemens AG.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is Jan Fritz.

Conflict of interest

The authors of this manuscript declare relationships with the following companies: Siemens AG.

Statistics and biometry

One of the authors has significant statistical expertise.

Ethical approval

Institutional Review Board approval was obtained.

Informed consent

Written informed consent was obtained from all subjects in this study.

Methodology

prospective

experimental

performed at one institution

Supplementary material

330_2017_5153_MOESM1_ESM.tif (7.4 mb)
ESM 1 Bone-implant interface zones. The tibial and talar implant surfaces facing the host bone were divided into lateral, central and medial thirds, as well as five sectors in anterior to posterior direction, equating to a total of 15 tibial and 15 talar bone-implant interface zones. (tif 7.37 MB)
330_2017_5153_MOESM2_ESM.tif (2 mb)
ESM 2 P-P Plots for Compressed sensing (CS) slice Encoding for Metal Artifact Correction (SEMAC) versus high bandwidth (BW) turbo spin echo (TSE) magnetic resonance imaging (MRI) evaluations summarizing the overall results of the technical, implant and anatomical analyses. In the technical analysis, CS-SEMAC TSE images were superior in 35% of the readings, while high-BW TSE images were superior in 34% with an overall minimal preference for CS-SEMAC TSE images (Area under the curve (AUC) = 0.062). In the implant analysis, CS-SEMAC TSE images were superior in 97% of the readings, while high-BW images were superior in only 1% (AUC = 0.917). In the anatomical analysis, CS-SEMAC TSE images were superior in 42% of the readings, while high-BW was superior in 24% with a minimal preference for CS-SEMAC TSE images (AUC = 0.153). (tif 2.00 MB)

References

  1. 1.
    Lefrancois T, Younger A, Wing K, Penner MJ, Dryden P, Wong H et al (2017) A Prospective Study of Four Total Ankle Arthroplasty Implants by Non-Designer Investigators. J Bone Joint Surg Am 99:342–348CrossRefPubMedGoogle Scholar
  2. 2.
    Tan EW, Maccario C, Talusan PG, Schon LC (2016) Early Complications and Secondary Procedures in Transfibular Total Ankle Replacement. Foot Ankle Int 37:835–841CrossRefPubMedGoogle Scholar
  3. 3.
    Hsu AR, Haddad SL, Myerson MS (2015) Evaluation and management of the painful total ankle arthroplasty. J Am Acad Orthop Surg 23:272–282CrossRefPubMedGoogle Scholar
  4. 4.
    Bestic JM, Bancroft LW, Peterson JJ, Kransdorf MJ (2008) Postoperative imaging of the total ankle arthroplasty. Radiol Clin N Am 46:1003–1015 v-viCrossRefPubMedGoogle Scholar
  5. 5.
    Bestic JM, Peterson JJ, DeOrio JK, Bancroft LW, Berquist TH, Kransdorf MJ (2008) Postoperative evaluation of the total ankle arthroplasty. AJR Am J Roentgenol 190:1112–1123CrossRefPubMedGoogle Scholar
  6. 6.
    Kim DR, Choi YS, Potter HG, Li AE, Chun KY, Jung YY et al (2016) Total Ankle Arthroplasty: An Imaging Overview. Korean J Radiol 17:413–423CrossRefPubMedCentralPubMedGoogle Scholar
  7. 7.
    Mulcahy H, Chew FS (2015) Current Concepts in Total Ankle Replacement for Radiologists: Features and Imaging Assessment. AJR Am J Roentgenol 205:1038–1047CrossRefPubMedGoogle Scholar
  8. 8.
    Cadden AR (2012) Imaging in total ankle replacement. Semin Musculoskelet Radiol 16:205–216CrossRefPubMedGoogle Scholar
  9. 9.
    Fritz J, Lurie B, Miller TT (2013) Imaging of hip arthroplasty. Semin Musculoskelet Radiol 17:316–327CrossRefPubMedGoogle Scholar
  10. 10.
    Fritz J, Lurie B, Miller TT, Potter HG (2014) MR imaging of hip arthroplasty implants. Radiographics 34:E106–E132CrossRefPubMedGoogle Scholar
  11. 11.
    Fritz J, Lurie B, Potter HG (2015) MR Imaging of Knee Arthroplasty Implants. Radiographics 35:1483–1501CrossRefPubMedCentralPubMedGoogle Scholar
  12. 12.
    Jungmann PM, Agten CA, Pfirrmann CW, Sutter R (2017) Advances in MRI around metal. J Magn Reson ImagingGoogle Scholar
  13. 13.
    Lu W, Pauly KB, Gold GE, Pauly JM, Hargreaves BA (2009) SEMAC: Slice Encoding for Metal Artifact Correction in MRI. Magn Reson Med 62:66–76CrossRefPubMedCentralPubMedGoogle Scholar
  14. 14.
    Koch KM, Lorbiecki JE, Hinks RS, King KF (2009) A multispectral three-dimensional acquisition technique for imaging near metal implants. Magn Reson Med 61:381–390CrossRefPubMedGoogle Scholar
  15. 15.
    Agten CA, Del Grande F, Fucentese SF, Blatter S, Pfirrmann CW, Sutter R (2015) Unicompartmental knee arthroplasty MRI: impact of slice-encoding for metal artefact correction MRI on image quality, findings and therapy decision. Eur Radiol 25:2184–2193CrossRefPubMedGoogle Scholar
  16. 16.
    Agten CA, Sutter R, Dora C, Pfirrmann CW (2017) MR imaging of soft tissue alterations after total hip arthroplasty: comparison of classic surgical approaches. Eur Radiol 27:1312–1321CrossRefPubMedGoogle Scholar
  17. 17.
    Hilgenfeld T, Prager M, Heil A, Schwindling FS, Nittka M, Grodzki D, et al (2017) PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging. Eur RadiolGoogle Scholar
  18. 18.
    Kretzschmar M, Nardo L, Han MM, Heilmeier U, Sam C, Joseph GB et al (2015) Metal artefact suppression at 3 T MRI: comparison of MAVRIC-SL with conventional fast spin echo sequences in patients with Hip joint arthroplasty. Eur Radiol 25:2403–2411CrossRefPubMedGoogle Scholar
  19. 19.
    Sutter R, Ulbrich EJ, Jellus V, Nittka M, Pfirrmann CW (2012) Reduction of metal artifacts in patients with total hip arthroplasty with slice-encoding metal artifact correction and view-angle tilting MR imaging. Radiology 265:204–214CrossRefPubMedGoogle Scholar
  20. 20.
    Ai T, Padua A, Goerner F, Nittka M, Gugala Z, Jadhav S et al (2012) SEMAC-VAT and MSVAT-SPACE sequence strategies for metal artifact reduction in 1.5T magnetic resonance imaging. Investig Radiol 47:267–276CrossRefGoogle Scholar
  21. 21.
    Deligianni X, Bieri O, Elke R, Wischer T, Egelhof T (2015) Optimization of Scan Time in MRI for Total Hip Prostheses: SEMAC Tailoring for Prosthetic Implants Containing Different Types of Metals. Röfo 187:1116–1122PubMedGoogle Scholar
  22. 22.
    Mansson S, Muller GM, Wellman F, Nittka M, Lundin B (2015) Phantom based qualitative and quantitative evaluation of artifacts in MR images of metallic hip prostheses. Phys Med 31:173–178CrossRefPubMedGoogle Scholar
  23. 23.
    Hargreaves BA, Chen W, Lu W, Alley MT, Gold GE, Brau AC et al (2010) Accelerated slice encoding for metal artifact correction. J Magn Reson Imaging 31:987–996CrossRefPubMedCentralPubMedGoogle Scholar
  24. 24.
    Hayter CL, Koff MF, Shah P, Koch KM, Miller TT, Potter HG (2011) MRI after arthroplasty: comparison of MAVRIC and conventional fast spin-echo techniques. AJR Am J Roentgenol 197:W405–W411CrossRefPubMedGoogle Scholar
  25. 25.
    Worters PW, Sung K, Stevens KJ, Koch KM, Hargreaves BA (2013) Compressed-sensing multispectral imaging of the postoperative spine. J Magn Reson Imaging 37:243–248CrossRefPubMedGoogle Scholar
  26. 26.
    Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195CrossRefPubMedGoogle Scholar
  27. 27.
    Fritz J, Ahlawat S, Demehri S, Thawait GK, Raithel E, Gilson WD et al (2016) Compressed Sensing SEMAC: 8-fold Accelerated High Resolution Metal Artifact Reduction MRI of Cobalt-Chromium Knee Arthroplasty Implants. Investig Radiol 51:666–676CrossRefGoogle Scholar
  28. 28.
    Fritz J, Fritz B, Thawait GK, Raithel E, Gilson WD, Nittka M et al (2016) Advanced metal artifact reduction MRI of metal-on-metal hip resurfacing arthroplasty implants: compressed sensing acceleration enables the time-neutral use of SEMAC. Skelet Radiol 45:1345–1356CrossRefGoogle Scholar
  29. 29.
    Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMedGoogle Scholar
  30. 30.
    Wilk MB, Gnanadesikan R (1968) Probability plotting methods for the analysis of data. Biometrika 55:1–17PubMedGoogle Scholar
  31. 31.
    Holmgren EB (1991) The P-P Plot as a Method for Comparing Treatment Effects. J Am Stat Assoc 90:360–365CrossRefGoogle Scholar
  32. 32.
    Pagenstert G, Horisberger M, Leumann AG, Wiewiorski M, Hintermann B, Valderrabano V (2011) Distinctive pain course during first year after total ankle arthroplasty: a prospective, observational study. Foot Ankle Int 32:113–119CrossRefPubMedGoogle Scholar
  33. 33.
    Mulcahy H, Chew FS (2015) Current Concepts in Total Ankle Replacement for Radiologists: Complications. AJR Am J Roentgenol 205:1244–1250CrossRefPubMedGoogle Scholar
  34. 34.
    Daniels TR, Mayich DJ, Penner MJ (2015) Intermediate to Long-Term Outcomes of Total Ankle Replacement with the Scandinavian Total Ankle Replacement (STAR). J Bone Joint Surg Am 97:895–903CrossRefPubMedGoogle Scholar
  35. 35.
    Mason LW, Wyatt J, Butcher C, Wieshmann H, Molloy AP (2015) Single-photon-emission computed tomography in painful total ankle replacements. Foot Ankle Int 36:635–640CrossRefPubMedGoogle Scholar
  36. 36.
    Kotnis R, Pasapula C, Anwar F, Cooke PH, Sharp RJ (2006) The management of failed ankle replacement. J Bone Joint Surg (Br) 88:1039–1047CrossRefGoogle Scholar
  37. 37.
    Ulbrich EJ, Sutter R, Aguiar RF, Nittka M, Pfirrmann CW (2012) STIR sequence with increased receiver bandwidth of the inversion pulse for reduction of metallic artifacts. AJR Am J Roentgenol 199:W735–W742CrossRefPubMedGoogle Scholar
  38. 38.
    Koch KM, Brau AC, Chen W, Gold GE, Hargreaves BA, Koff M et al (2011) Imaging near metal with a MAVRIC-SEMAC hybrid. Magn Reson Med 65:71–82CrossRefPubMedGoogle Scholar

Copyright information

© European Society of Radiology 2017

Authors and Affiliations

  1. 1.Department of Orthopedic SurgeryMedStar Union Memorial HospitalBaltimoreUSA
  2. 2.Russell H. Morgan Department of Radiology and Radiological ScienceJohns Hopkins University School of MedicineBaltimoreUSA