Abstract
Purpose
To measure the test-retest reproducibility of an automated system for quantifying whole body and compartmental muscle volumes using wide bore 3 T MRI.
Materials and methods
Thirty volunteers stratified by body mass index underwent whole body 3 T MRI, two-point Dixon sequences, on two separate occasions. Water-fat separation was performed, with automated segmentation of whole body, torso, upper and lower leg volumes, and manually segmented lower leg muscle volumes.
Results
Mean automated total body muscle volume was 19·32 L (SD9·1) and 19·28 L (SD9·12) for first and second acquisitions (Intraclass correlation coefficient (ICC) = 1·0, 95 % level of agreement -0·32–0·2 L). ICC for all automated test-retest muscle volumes were almost perfect (0·99–1·0) with 95 % levels of agreement 1.8–6.6 % of mean volume. Automated muscle volume measurements correlate closely with manual quantification (right lower leg: manual 1·68 L (2SD0·6) compared to automated 1·64 L (2SD 0·6), left lower leg: manual 1·69 L (2SD 0·64) compared to automated 1·63 L (SD0·61), correlation coefficients for automated and manual segmentation were 0·94–0·96).
Conclusion
Fully automated whole body and compartmental muscle volume quantification can be achieved rapidly on a 3 T wide bore system with very low margins of error, excellent test-retest reliability and excellent correlation to manual segmentation in the lower leg.
Key Points
• Sarcopaenia is an important reversible complication of a number of diseases.
• Manual quantification of muscle volume is time-consuming and expensive.
• Muscles can be imaged using in and out of phase MRI.
• Automated atlas-based segmentation can identify muscle groups.
• Automated muscle volume segmentation is reproducible and can replace manual measurements.
Similar content being viewed by others
References
Berger MJ, Doherty TJ (2010) Sarcopenia: Prevalence, Mechanisms, and Functional Consequences. In: Mobbs CV, Hof PR (eds) Interdiscip. Top. Gerontol. KARGER, Basel, pp 94–114
Summers GD, Deighton CM, Rennie MJ, Booth AH (2008) Rheumatoid cachexia: a clinical perspective. Rheumatology 47:1124–1131
Toda Y, Segal N, Toda T et al (2000) A decline in lower extremity lean body mass per body weight is characteristic of women with early phase osteoarthritis of the knee. J Rheumatol 27:2449–2454
Morley JE, Thomas DR, Wilson M-MG (2006) Cachexia: pathophysiology and clinical relevance. Am J Clin Nutr 83:735–743
von Haehling S, Anker SD (2010) Cachexia as a major underestimated and unmet medical need: facts and numbers. J Cachex Sarcopenia Muscle 1:1–5
Cooney JK, Law R-J, Matschke V et al (2011) Benefits of Exercise in Rheumatoid Arthritis. J Aging Res. doi:10.4061/2011/681640
HajGhanbari B, Hamarneh G, Changizi N et al (2011) MRI-based 3D shape analysis of thigh muscles patients with chronic obstructive pulmonary disease versus healthy adults. Acad Radiol 18:155–166
Zanetti M, Gerber C, Hodler J (1998) Quantitative assessment of the muscles of the rotator cuff with magnetic resonance imaging. Investig Radiol 33:163–170
Preininger B, Schmorl K, von Roth P et al (2012) The sex specificity of hip-joint muscles offers an explanation for better results in men after total hip arthroplasty. Int Orthop 36:1143–1148
Roubenoff R, Walsmith J, Lundgren N et al (2002) Low physical activity reduces total energy expenditure in women with rheumatoid arthritis: implications for dietary intake recommendations. Am J Clin Nutr 76:774–779
Lemmey AB, Marcora SM, Chester K et al (2009) Effects of high-intensity resistance training in patients with rheumatoid arthritis: A randomized controlled trial. Arthritis Care Res 61:1726–1734
Marcora SM, Lemmey AB, Maddison PJ (2005) Can progressive resistance training reverse cachexia in patients with rheumatoid arthritis? Results of a pilot study. J Rheumatol 32:1031–1039
Fortin M, Battié MC (2012) Quantitative Paraspinal Muscle Measurements: Inter-Software Reliability and Agreement Using OsiriX and ImageJ. Phys Ther 92:853–864
Hoyte L, Ye W, Brubaker L et al (2011) Segmentations of MRI images of the female pelvic floor: A study of inter- and intra-reader reliability. J Magn Reson Imaging 33:684–691
Brunner G, Nambi V, Yang E et al (2011) Automatic quantification of muscle volumes in magnetic resonance imaging scans of the lower extremities. Magn Reson Imaging 29:1065–1075
Mattei JP, Fur YL, Cuge N et al (2006) Segmentation of fascias, fat and muscle from magnetic resonance images in humans: the DISPIMAG software. Magn Reson Mater Phys Biol Med 19:275–279
Wald D, Teucher B, Dinkel J et al (2012) Automatic quantification of subcutaneous and visceral adipose tissue from whole‐body magnetic resonance images suitable for large cohort studies. J Magn Reson Imaging 36:1421–1434
Broderick BJ, Dessus S, Grace PA, ÓLaighin G (2010) Technique for the computation of lower leg muscle bulk from magnetic resonance images. Med Eng Phys 32:926–933
Baudin P-Y, Azzabou N, Carlier PG, Paragios N (2012) Prior Knowledge, Random Walks and Human Skeletal Muscle Segmentation. In: Ayache N, Delingette H, Golland P, Mori K (eds) Med. Image Comput. Comput.-Assist. Interv. – MICCAI 2012. Springer, Berlin, pp 569–576
Engstrom CM, Fripp J, Jurcak V et al (2011) Segmentation of the quadratus lumborum muscle using statistical shape modeling. J Magn Reson Imaging 33:1422–1429
Gubern-Mérida A, Kallenberg M, Martí R, Karssemeijer N (2012) Segmentation of the Pectoral Muscle in Breast MRI Using Atlas-Based Approaches. In: Ayache N, Delingette H, Golland P, Mori K (eds) Med. Image Comput. Comput.-Assist. Interv. – MICCAI 2012. Springer, Berlin, pp 371–378
Walter SD, Eliasziw M, Donner A (1998) Sample size and optimal designs for reliability studies. Stat Med 17:101–110
Dixon WT (1984) Simple proton spectroscopic imaging. Radiology 153:189–194
Rydell J, Johansson A, Leinhard OD, et al (2008) Three dimensional phase sensitive reconstruction for water/fat separation in MR imaging using inverse gradient. Proc. Int. Soc. Magn. Reson. Med. Annu. Meet. ISMRM’08. p 1521
Rydell J, Knutsson H, Pettersson J et al (2007) Phase Sensitive Reconstruction for Water/Fat Separation in MR Imaging Using Inverse Gradient. In: Ayache N, Ourselin S, Maeder A (eds) Med. Image Comput. Comput.-Assist. Interv. – MICCAI 2007. Springer, Berlin, pp 210–218
Dahlqvist Leinhard O, Johansson A, Lundberg P (2008) Water-fat shift displacement artifact correction in two-point Dixon imaging. Proc Intl Soc Mag Reson Med. p 1384
Leinhard OD, Johansson A, Rydell J, et al (2008) Quantitative abdominal fat estimation using MRI. 19th Int. Conf. Pattern Recognit. 2008 ICPR 2008. pp 1–4
Romu T, Borga M, Dahlqvist O (2011) MANA - Multi scale adaptive normalized averaging. 2011 I.E. Int. Symp. Biomed. Imaging Nano Macro. pp 361–364
Zijdenbos AP, Dawant BM, Margolin RA, Palmer AC (1994) Morphometric analysis of white matter lesions in MR images: method and validation. IEEE Trans Med Imaging 13:716–724
Commean PK, Tuttle LJ, Hastings MK et al (2011) Magnetic resonance imaging measurement reproducibility for calf muscle and adipose tissue volume. J Magn Reson Imaging 34:1285–1294
Hiba B, Richard N, Hébert LJ et al (2012) Quantitative assessment of skeletal muscle degeneration in patients with myotonic dystrophy type 1 using MRI. J Magn Reson Imaging 35:678–685
Dahlqvist Leinhard O, Romu T, Kihlberg J, et al (2012) Validation of whole--‐body adipose tissue quantification using air displacement plethysmometry
Barra V, Boire J-Y (2002) Segmentation of fat and muscle from MR images of the thigh by a possibilistic clustering algorithm. Comput Methods Prog Biomed 68:185–193
Acknowledgments
The scientific guarantor of this publication is Dr Andoni Toms. The authors of this manuscript, Magnus Borga, Olof Leinhard and Johannes Rosander, declare relationships with the following company: Advanced MR Analytics AB, Linköping, Sweden. The authors state that this work has not received any funding. One of the authors has significant statistical expertise. Institutional Review Board approval was obtained. Written informed consent was obtained from all subjects (patients) in this study. Methodology: prospective, experimental, multicenter study.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Thomas, M.S., Newman, D., Leinhard, O.D. et al. Test-retest reliability of automated whole body and compartmental muscle volume measurements on a wide bore 3T MR system. Eur Radiol 24, 2279–2291 (2014). https://doi.org/10.1007/s00330-014-3226-6
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-014-3226-6