Abstract
Objective
To assess reduced volumes of contrast agent on image quality for coronary computed tomography angiography (CCTA) by using single-beat cardiac imaging with 320-slice CT.
Materials and methods
Forty consecutive male patients (mean age: 55.8 years) undergoing CCTA with body weight ≤85 kg, heart rate ≤65 bpm, and ejection fraction ≥55% were included. Image acquisition protocol was standardized (120 kV, 400 mA, and prospective ECG-triggered single-beat nonspiral CCTA). Patients were randomly assigned to one of four groups (G1: received 40 ml, G2: 50 ml, G3: 60 ml, G4: 70 ml). Groups were compared with respect to aortic attenuation, image noise, and image quality.
Results
CT values (mean ± standard deviation) in the aortic root were measured as 423 ± 38 HU in G1, and 471 ± 68, 463 ± 60, and 476 ± 78 HU in G2-4, respectively. There were no statistically significant differences in attenuation among the groups (P > 0.068). All 40 CT datasets were rated diagnostic, and image noise and image quality were not statistically different among groups.
Conclusion
Using 320-slice volume CT, diagnostic image quality can be achieved with 40 ml of contrast material in CCTA in patients with normal body weight, cardiac function, and low heart rate.
Similar content being viewed by others
References
Roberts WT, Bax JJ, Davies LC (2008) Cardiac CT and CT coronary angiography: technology and application. Heart 94:781–792
Cademartiri F, Mollet NR, Runza G et al (2006) Improving diagnostic accuracy of MDCT coronary angiography in patients with mild heart rhythm irregularities using ECG editing. AJR Am J Roentgenol 186:634–638
Cademartiri F, Nieman K, van der Lugt A et al (2004) Intravenous contrast material administration at 16-detector row helical CT coronary angiography: test bolus versus bolus-tracking technique. Radiology 233:817–823
Becker CR, Hong C, Knez A et al (2003) Optimal contrast application for cardiac 4-detector-row computed tomography. Invest Radiol 38:690–694
Rist C, Nikolaou K, Kirchin MA et al (2006) Contrast bolus optimization for cardiac 16-slice computed tomography: comparison of contrast medium formulations containing 300 and 400 milligrams of iodine per milliliter. Invest Radiol 41:460–467
Miller JM, Dewey M, Vavere AL et al (2009) Coronary CT angiography using 64 detector rows: methods and design of the multi-centre trial CORE-64. Eur Radiol 19:816–828
Tatsugami F, Kanamoto T, Nakai G et al (2009) Reduction of the total injection volume of contrast material with a short injection duration in 64-detector row CT coronary angiography. Br J Radiol. doi:10.1259/bjr/23203690
From AM, Bartholmai BJ, Williams AW et al (2008) Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc 83:1095–1100
Toprak O (2007) Conflicting and new risk factors for contrast induced nephropathy. J Urol 178:2277–2283
Thomsen HS (2008) Current evidence on prevention and management of contrast-induced nephropathy. Eur Radiol 17(Suppl 6):33–37
Thomsen HS, Morcos SK, Barrett BJ (2008) Contrast-induced nephropathy: the wheel has turned 360 degrees. Acta Radiol 49:646–657
McCullough PA (2008) Contrast-induced acute kidney injury. J Am Coll Cardiol 51:1419–1428
Johnson PT, Pannu HK, Fishman EK (2009) IV contrast infusion for coronary artery CT angiography: literature review and results of a nationwide survey. AJR Am J Roentgenol 192:W214–221
Husmann L, Herzog BA, Burkhard N et al (2009) Body physique and heart rate variability determine the occurrence of stair-step artefacts in 64-slice CT coronary angiography with prospective ECG-triggering. Eur Radiol 19:1698–1703
Aldrovandi A, Maffei E, Palumbo A et al (2009) Prognostic value of computed tomography coronary angiography in patients with suspected coronary artery disease: a 24-month follow-up study. Eur Radiol 19:1653–1660
Lund GK, Wegian E, Saeed M et al (2009) 64-Slice spiral computed tomography of the coronary arteries: dose reduction using an optimized imaging protocol including individual weight-adaptation of voltage and current-time product. Eur Radiol 19:1132–1138
Bae KT, Seeck BA, Hildebolt CF et al (2008) Contrast enhancement in cardiovascular MDCT: effect of body weight, height, body surface area, body mass index, and obesity. AJR 190:777–784
Nakaura T, Awai K, Yauaga Y et al (2008) Contrast injection protocols for coronary computed tomography angiography using a 64-detector scanner: comparison between patient weight-adjusted- and fixed iodine-dose protocols. Invest Radiol 43:512–519
Yamamuro M, Tadamura E, Kanao S et al (2007) Coronary angiography by 64-detector row computed tomography using low dose of contrast material with saline chaser: influence of total injection volume on vessel attenuation. J Comput Assist Tomogr 31:272–280
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Hein, P.A., May, J., Rogalla, P. et al. Feasibility of contrast material volume reduction in coronary artery imaging using 320-slice volume CT. Eur Radiol 20, 1337–1343 (2010). https://doi.org/10.1007/s00330-009-1692-z
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-009-1692-z