Abstract
The aim of this study was to assess the performance of a motion-map algorithm that automatically determines optimal reconstruction windows for dual-source coronary CT angiography. In datasets from 50 consecutive patients, optimal systolic and diastolic reconstruction windows were determined using the motion-map algorithm. For manual determination of the optimal reconstruction window, datasets were reconstructed in 5% steps throughout the RR interval. Motion artifacts were rated for each major coronary vessel using a five-point scale. Mean motion scores using the motion-map algorithm were 2.4 ± 0.8 for systolic reconstructions and 1.9 ± 0.8 for diastolic reconstructions. Using the manual approach, overall motion scores were significantly better (1.9 ± 0.5 and 1.7 ± 0.6, p < 0.05), but diagnostic image quality was reached in >90% of cases using either approach. Using the automated approach, there was a negative correlation between heart rate and motion scores for systolic reconstructions (ρ = −0.26, p < 0.05) and a positive correlation for diastolic reconstructions (ρ = 0.46, p < 0.01). For the manual approach, no significant correlation was found for systolic reconstructions (ρ = −0.1, p = 0.52), while there was a positive correlation for diastolic reconstructions (ρ = 0.48, p < 0.01). Thus, the motion-map algorithm is a useful tool to save time in finding an appropriate reconstruction window in patients with heart rates <70 bpm (diastolic reconstruction) and >80 bpm (systolic reconstruction).
Similar content being viewed by others
References
Herzog C, Zwerner PL, Doll JR, Nielsen CD, Nguyen SA, Savino G, Vogl TJ, Costello P, Schoepf UJ (2007) Significant coronary artery stenosis: comparison on per-patient and per-vessel or per-segment basis at 64-section CT angiography. Radiology 244:112–120
Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557
Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Suss C, Grasruck M, Stierstorfer K, Krauss B, Raupach R, Primak AN, Kuttner A, Achenbach S, Becker C, Kopp A, Ohnesorge BM (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268
Matt D, Scheffel H, Leschka S, Flohr T, Marincek B, Kaufmann P, Alkadhi H (2007) Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability. AJR 189:567–573
Johnson T, Nikolaou K, Wintersperger B, Leber A, von Ziegler F, Rist C, Buhmann S, Knez A, Reiser M, Becker C (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415
Scheffel H, Alkadhi H, Plass A, Vachenauer R, Desbiolles L, Gaemperli O, Schepis T, Frauenfelder T, Schertler T, Husmann L, Grunenfelder J, Genoni M, Kaufmann PA, Marincek B, Leschka S (2006) Accuracy of dual-source CT coronary angiography: First experience in a high pre-test probability population without heart rate control. Eur Radiol 16:2739–2747
Leschka S, Scheffel H, Desbiolles L, Plass A, Gaemperli O, Valenta I, Husmann L, Flohr TG, Genoni M, Marincek B, Kaufmann PA, Alkadhi H (2007) Image quality and reconstruction intervals of dual-source CT coronary angiography: recommendations for ECG-pulsing windowing. Invest Radiol 42:543–549
Herzog C, Arning-Erb M, Zangos S, Eichler K, Hammerstingl R, Dogan S, Ackermann H, Vogl T (2006) Multi-detector row CT coronary angiography: influence of reconstruction technique and heart rate on image quality. Radiology 238:75–86
Achenbach S, Ropers D, Kuettner A, Flohr T, Ohnesorge B, Bruder H, Theessen H, Karakaya M, Daniel WG, Bautz W, Kalender WA, Anders K (2006) Contrast-enhanced coronary artery visualization by dual-source computed tomography-initial experience. Eur J Radiol 57:331–335
Leschka S, Husmann L, Desbiolles LM, Gaemperli O, Schepis T, Koepfli P, Boehm T, Marincek B, Kaufmann PA, Alkadhi H (2006) Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT. Eur Radiol 16:1964–1972
Wintersperger B, Nikolaou K, von Ziegler F, Johnson T, Rist C, Leber A, Flohr T, Knez A, Reiser M, Becker C (2006) Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed. Invest Radiol 41:436–442
Manzke R, Kohler T, Nielsen T, Hawkes D, Grass M (2004) Automatic phase determination for retrospectively gated cardiac CT. Med Phys 31:3345–3362
Hoffmann MH, Lessick J, Manzke R, Schmid FT, Gershin E, Boll DT, Rispler S, Aschoff AJ, Grass M (2006) Automatic determination of minimal cardiac motion phases for computed tomography imaging: initial experience. Eur Radiol 16:365–373
Seifarth H, Wienbeck S, Pusken M, Juergens KU, Maintz D, Vahlhaus C, Heindel W, Fischbach R (2007) Optimal systolic and diastolic reconstruction windows for coronary CT angiography using dual-source CT. AJR Am J Roentgenol 189:1317–1323
Brodoefel H, Burgstahler C, Tsiflikas I, Reimann A, Schroeder S, Claussen CD, Heuschmid M, Kopp AF (2008) Dual-source CT: effect of heart rate, heart rate variability, and calcification on image quality and diagnostic accuracy. Radiology 247:346–355
Rixe J, Rolf A, Conradi G, Elsaesser A, Moellmann H, Nef HM, Bachmann G, Hamm CW, Dill T (2008) Image quality on dual-source computed-tomographic coronary angiography. Eur Radiol 18:1857–1862
Hoffmann M, Shi H, Manzke R, Schmid F, De Vries L, Grass M, Brambs H, Aschoff A (2005) Noninvasive coronary angiography with 16-detector row CT: effect of heart rate. Radiology 234:86–97
Kachelriess M, Sennst D, Maxlmoser W, Kalender W (2002) Kymogram detection and kymogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart. Med Phys 29:1489–1503
Ertel D, Pflederer T, Achenbach S, Kachelriess M, Steffen P, Kalender WA (2008) Validation of a raw data-based synchronization signal (kymogram) for phase-correlated cardiac image reconstruction. Eur Radiol 18:253–262
Seifarth H, Maintz D, Jürgens K, Bruder H, Heindel W, Fischbach R (2006) Electrocardiogram-independent image reconstruction in cardiac multidetector computed tomography using retrospective motion synchronization. Invest Radiol 41:898–903
McCollough CH, Primak AN, Saba O, Bruder H, Stierstorfer K, Raupach R, Suess C, Schmidt B, Ohnesorge BM, Flohr TG (2007) Dose performance of a 64-channel dual-source CT scanner. Radiology 243:775–784
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Seifarth, H., Puesken, M., Wienbeck, S. et al. Automatic selection of optimal systolic and diastolic reconstruction windows for dual-source CT coronary angiography. Eur Radiol 19, 1645–1652 (2009). https://doi.org/10.1007/s00330-009-1329-2
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-009-1329-2