Abstract

The purpose was to compare global left-ventricular (LV) function parameters measured with cine MRI with results from multiphase dual-source CT (DSCT) using 10 and 20 reconstruction phases. Twenty-eight patients with suspected or known CAD underwent DSCT coronary angiography. LV end-diastolic (EDV), end-systolic (ESV) and stroke volumes (SV), and ejection fraction (EF) were determined using LV segmentation and selection of specific phases from DSCT image sets reconstructed either at 5% or 10% steps through the R-R interval. Cine MRI served as the reference investigation. Threshold-based 3D-segmentation was feasible in all DSCT data sets. EDV and ESV were underestimated by DSCT, but showed excellent correlation (Pearson’s correlation coefficient 0.95/0.97) to values obtained with MRI. Using data from 5% DSCT image reconstructions instead of 10% phase reconstructions, the position of the ED and ES phase was changed in 16 of 28 patients; ESVs were to found to be slightly smaller, whereas EDV were slightly larger, resulting in a systematic overestimation of LV EF by 1.9% (p = 0.56). Threshold-based 3D segmentation enables accurate and reliable DSCT determination of global LV function with excellent correlation to cine MRI. Minor differences in LV EF indicate that both modalities are virtually interchangeable, even if the number of reconstructed phases is limited to 10% phase reconstructions.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

Explore related subjects

Discover the latest articles, news and stories from top researchers in related subjects.

References

  1. White HD, Norris RM, Brown MA, Brandt PWT, Whitlock RML, Wild CJ (1987) Left ventricular end-systolic volume as the major determination of survival after recovery from myocardial infarction. Circulation 76:44–51

    PubMed  CAS  Google Scholar 

  2. Hammermeister KE, DeRouen TA, Dodge HT (1979) Variables predictive of survival in patients with coronary disease. Selection by univariate and multivariate analyses from the clinical, electrocardiographic, exercise, arteriographic, and quantitative angiographic evaluations. Circulation 59:421–430

    PubMed  CAS  Google Scholar 

  3. Shah PK, Maddahi J, Staniloff HM, Ellrodt AG, Pichler M, Swan HJ, Berman DS (1986) Variable spectrum and prognostic implications of left and right ventricular ejection fraction in patients with and without clinical heart failure after acute myocardial infarction. Am J Cardiol 58:387–393

    Article  PubMed  CAS  Google Scholar 

  4. Dirksen MS, Bax JJ, de Roos A, Jukema JW, van der Geest RJ, Geleijns K, Boersma E, van der Wall EE, Lamb HJ (2002) Usefulness of dynamic multislice computed tomography of left ventricular function in unstable angina pectoris and comparison with echocardiography. Am J Cardiol 90:1157–1160

    Article  PubMed  Google Scholar 

  5. Grude M, Juergens KU, Wichter T, Paul M, Fallenberg EM, Muller JG, Heindel W, Breithardt G, Fischbach R (2003) Evaluation of global left ventricular myocardial function with electrocardiogram-gated multidetector computed tomography: comparison with magnetic resonance imaging. Invest Radiol 38:653–661

    Article  PubMed  Google Scholar 

  6. Koch K, Oellig F, Kunz P, Bender P, Oberholzer K, Mildenberger P, Hake U, Kreitner KF, Thelen M (2004) Assessment of global and regional left ventricular function with a 16-slice Spiral-CT using two different software tools for quantitative analysis and qualitative evaluation of wall motion changes in comparison with magnetic resonance imaging. Fortschr Röntgenstr 176:1786–1793

    Article  CAS  Google Scholar 

  7. Juergens KU, Grude M, Maintz D, Fallenberg EM, Wichter T, Heindel W, Fischbach R (2004) Multi-detector row CT of left ventricular function with dedicated analysis software versus MR imaging: initial experience. Radiology 230:403–410

    Article  PubMed  Google Scholar 

  8. Heuschmid M, Rothfuss J, Schröder S, Küttner A, Fenchel M, Stauder N, Mahnken AH, Burgstahler C, Miller S, Claussen CD, Kopp AF (2005) Left ventricular functional parameters: comparison of 16-slice spiral CT with MRI. Fortschr Röntgenstr 177:60–66

    Article  CAS  Google Scholar 

  9. Mahnken AH, Koos R, Katoh M, Spuentrup E, Busch P, Wildberger JE, Kühl HP, Günther RW (2005) Sixteen-slice spiral CT versus MR imaging for the assessment of left ventricular function in acute myocardial infarction. Eur Radiol 15:714–720

    Article  PubMed  Google Scholar 

  10. Yamamuro M, Tadamura E, Kubo S, Toyoda H, Nishina T, Ohba M, Hosokawa R, Kimura T, Tamaki N, Komeda M, Kita T, Konishi J (2005) Cardiac functional analysis with multi-detector row CT and segmental reconstruction algorithm: comparison with echocardiography, SPECT, and MR imaging. Radiology 234:381–390

    Article  PubMed  Google Scholar 

  11. Schlosser T, Pagonidis K, Herborn CU, Hunold P, Waltering KU, Lauenstein TC, Barkhausen J (2005) Assessment of left ventricular parameters using 16-MDCT and new software for endocardial and epicardial border delineation. AJR Am J Roentgenol 184:765–773

    PubMed  Google Scholar 

  12. Heuschmid M, Rothfuss JK, Schroeder S, Fenchel M, Stauder N, Burgstahler C, Franow A, Kuzo RS, Kuettner A, Miller S, Claussen CD, Kopp AF (2006) Assessment of left ventricular myocardial function using 16-slice multidetector-row computed tomography: comparison with magnetic resonance imaging and echocardiography. Eur Radiol 16:551–559

    Article  PubMed  Google Scholar 

  13. Mahnken AH, Mühlenbruch G, Koos R, Stanzel S, Busch PS, Niethammer M, Günther RW, Wildberger JE (2006) Automated vs. manual assessment of left ventricular function in cardiac multidetector row computed tomography: comparison with magnetic resonance imaging. Eur Radiol 16:1416–1423

    Article  PubMed  Google Scholar 

  14. Belge B, Coche E, Pasquet A, Vanoverschelde JL, Gerber BL (2006) Accurate estimation of global and regional cardiac function by retrospectively gated multidetector row computed tomography: comparison with cine magnetic resonance imaging. Eur Radiol 16:1424–1433

    Article  PubMed  Google Scholar 

  15. Juergens KU, Seifarth H, Maintz D, Grude M, Ozgun M, Wichter T, Heindel W, Fischbach R (2006) Left ventricular function determination with multidetector-row CT of the heart: Are short-axis image reformations necessary? AJR Am J Roentgenol 186:S371–S378

    Article  PubMed  Google Scholar 

  16. Dewey M, Muller M, Teige F, Hamm B (2006) Evaluation of a semiautomatic software tool for left ventricular function analysis with 16-slice computed tomography. Eur Radiol 16:25–31

    Article  PubMed  Google Scholar 

  17. Sugeng L, Mor-Avi V, Weinert L, Niel J, Ebner C, Steringer-Mascherbauer R, Schmidt F, Galuschky C, Schummers G, Lang RM, Nesser HJ (2006) Quantitative assessment of left ventricular size and function. Side-by-side comparison of real-time three-dimensional echocardiography and computed tomography with magnetic resonance reference. Circulation 114:654–661

    Article  PubMed  Google Scholar 

  18. Dewey M, Müller M, Eddicks S, Schnapauff D, Teige F, Rutsch W, Borges AC, Hamm B (2006) Evaluation of global and regional left ventricular function with 16-slice computed tomography, biplane cineventriculography, and two-dimensional transthoracic echocardiography. J Am Coll Cardiol 48:2034–2044

    Article  PubMed  Google Scholar 

  19. Mühlenbruch G, Das M, Hohl C, Wildberger JE, Rinck D, Flohr TG, Koos R, Knackstedt C, Günther RW, Mahnken AH (2006) Global left ventricular function in cardiac CT. Evaluation of an automated 3D region-growing segmentation algorithm. Eur Radiol 16:1117–1123

    Article  PubMed  Google Scholar 

  20. Fischbach R, Juergens KU, Ozgun M, Maintz D, Grude M, Seifarth H, Heindel W, Wichter T (2007) Assessment of regional left ventricular function with multidetector-row computed tomography versus magnetic resonance imaging. Eur Radiol 17:1009–1017

    Article  PubMed  Google Scholar 

  21. Miller S, Simonetti OP, Carr J, Kramer U, Finn JP (2002) MR imaging of the heart with cine true fast imaging with steady-state precession: influence of spatial and temporal resolutions on left ventricular functional parameters. Radiology 223:263–269

    Article  PubMed  Google Scholar 

  22. Kopp AF, Heuschmid M, Reimann A, Kuettner A, Beck T, Ohmer M, Burgstahler C, Brodoefel H, Claussen CD, Schroeder S (2005) Evaluation of cardiac function and myocardial viability with 16- and 64-slice multidetector computed tomography. Eur Radiol 15:D15–D20

    Article  PubMed  Google Scholar 

  23. Henneman MM, Schuijf JD, Jukema JW, Holman ER, Lamb HJ, de Roos A, van der Wall EE, Bax JJ (2006) Assessment of global and regional left ventricular function and volumes with 64-slice MSCT: a comparison with 2D echocardiography. J Nucl Cardiol 13:480–487

    Article  PubMed  Google Scholar 

  24. Schepis T, Gaemperli O, Koepfli P, Valenta I, Strobel K, Brunner A, Leschka S, Desbiolles L, Husmann L, Alkadhi H, Kaufmann PA (2006) Comparison of 64-slice CT with gated SPECT for evaluation of left ventricular function. J Nucl Med 47:1288–1294

    PubMed  Google Scholar 

  25. Schlosser T, Mohrs OK, Magedanz A, Voigtländer T, Schmermund A, Barkhausen J (2007) Assessment of left ventricular function and mass in patients undergoing computed tomography (CT) coronary angiography using 64-detector-row CT: comparison to magnetic resonance imaging. Acta Radiol 48:30–35

    Article  PubMed  CAS  Google Scholar 

  26. Rist C, Johnson TR, Becker A, Leber AW, Huber A, Busch S, Becker CR, Reiser MF, Nikolaou K (2007) Dual-source cardiac CT imaging with improved temporal resolution: impact on image quality and analysis of left ventricular function. Radiologe 47:287–294

    Article  PubMed  Google Scholar 

  27. Mahnken AH, Bruder H, Suess C, Mühlenbruch G, Bruners P, Hohl C, Guenther RW, Wildberger JE (2007) Dual-source computed tomography for assessing cardiac function: a phantom study. Invest Radiol 42:491–498

    Article  PubMed  Google Scholar 

  28. Alfakih K, Plein S, Thiele H, Jones T, Ridgway JP, Sivananthan MU (2003) Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences. J Magn Reson Imaging 17:323–329

    Article  PubMed  Google Scholar 

  29. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310

    PubMed  CAS  Google Scholar 

  30. Busch S, Johnson TR, Wintersperger BJ, Minaifar N, Bhargava A, Rist C, Reiser MF, Becker C, Nikolaou K (2007) Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings. Eur Radiol: DOI 10.1007/s00330-007-0767-y

  31. Brodoefel H, Kramer U, Reimann A, Burgstahler C, Schroeder S, Kopp A, Heuschmid M (2007) Dual-source CT with improved temporal resolution in assessment of left ventricular function: a pilot study. AJR Am J Roentgenol 189:1064–1070

    Article  PubMed  Google Scholar 

  32. van der Geest RJ, Buller VG, Jansen E, Lamb HJ, Baur LH, van der Wall EE, de Roos A, Reiber JH (1997) Comparison between manual and semiautomated analysis of left ventricular volume parameters from short axis MR images. J Comput Assist Tomogr 21:756–765

    Article  PubMed  Google Scholar 

  33. Vogel-Claussen J, Finn JP, Gomes AS, Hundley GW, Jerosch-Herold M, Pearson G, Sinha S, Lima JA, Bluemke DA (2006) Left ventricular papillary muscle mass: relationship to left ventricular mass and volumes by magnetic resonance imaging. J Comput Assist Tomogr 30:426–432

    Article  PubMed  Google Scholar 

  34. Sievers B, Kirchberg S, Bakan A, Franken U, Trappe H-J (2004) Impact of papillary muscles in ventricular volume and ejection fraction assessment by cardiovascular magnetic resonance. J Cardiovasc Magn Reson 6:9–16

    Article  PubMed  Google Scholar 

  35. Van der Vleuten PA, Willems TP, Götte MJ, Tio RA, Greuter MJ, Zijlstra F, Oudkerk M (2006) Quantification of global left ventricular function: comparison of multidetector computed tomography and magnetic resonance imaging: a meta-analysis and review of the current literature. Acta Radiol 47:1049–1057

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

This work was partly supported by grants from the Deutsche Forschungsgemeinschaft (DFG), Sonderforschungsbereich 656 MoBil Münster, Germany (project C2)

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kai Uwe Juergens.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Puesken, M., Fischbach, R., Wenker, M. et al. Global left-ventricular function assessment using dual-source multidetector CT: effect of improved temporal resolution on ventricular volume measurement. Eur Radiol 18, 2087–2094 (2008). https://doi.org/10.1007/s00330-008-0982-1

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-008-0982-1

Keywords

Navigation