Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner

  • Computer Tomography
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

The effect of the z-flying focal spot (zFFS) technology was evaluated by simulations and measurements with respect to resolution and artifact behavior for a 64-slice spiral cone-beam computed tomography (CT) scanner. The zFFS alternates between two z-positions of the X-ray focal spot, acquiring two slices per detector row, which results in double sampling in the z-direction. We implemented a modified reconstruction that is able to obtain images as they would be without zFFS. A delta phantom equipped with a thin gold disc was used to measure slice sensitivity profiles (SSP), and a high-contrast bar phantom was used to quantify the resolution in the x/z-plane with and without zFFS. The zFFS decreases the full width at half maximum (FWHM) of the SSPs by a factor of about 1.4. The double z-sampling allows the separation of 0.4 mm bars in the z-direction compared with 0.6 mm in the case without zFFS. The zFFS effectively reduces windmill artifacts in the reconstructed images while maintaining the transverse resolution, even at the largest available pitch value of 1.5.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11

Similar content being viewed by others

References

  1. Silver MD, Taguchi K, Hein IA, Han HS, Kazama M, Mori I (2003) Windmill artifact in multislice CT. Proc SPIE 5032:1918–1927

    Article  Google Scholar 

  2. Taguchi K, Aradate H, Saito Y (2004) The cause of the artifact in 4-slice helical computed tomography. Med Phys 31(7):2033–2037

    Article  PubMed  Google Scholar 

  3. Kalender WA (2005) Computed Tomography, 2nd edn. Wiley-VCH, New York

    Google Scholar 

  4. Hsieh J (2003) Analytical models for multi-slice helical CT performance parameters. Med Phys 30(2):169–178

    Article  PubMed  Google Scholar 

  5. Schardt P, Deuringer J, Freudenberger J, Hell E, Knuepfer W, Mattern D, Schild M (2004) New x-ray tube performance in computed tomography by introducing the rotating envelope tube. Med Phys 32(9):2699–2706

    Article  Google Scholar 

  6. Kachelrieß M, Knaup M, Penßel C, Kalender WA (2005) Flying focal spot (FFS) in cone-beam CT. In: Records of the 2004 IEEE Medical Imaging Conference pp 3759–3763

  7. Flohr T, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough CH (2005) Image reconstruction and image quality evaluation of a 64-slice CT scanner with focal spot. Med Phys 32(8):2536–2547

    Article  PubMed  CAS  Google Scholar 

  8. Kachelrieß M, Schaller S, Kalender WA (2000) Advanced single-slice rebinning in cone-beam spiral CT”. Med Phys 27:754–772

    Article  PubMed  Google Scholar 

  9. Kachelrieß M, Kalender WA (2005) Presampling, algorithm factors, and noise in CT. Med Phys 32(5):1321–1334

    Article  PubMed  Google Scholar 

  10. Taguchi T, Aradate H (1998) Algorithm for image reconstruction in multislice helical CT. Med Phys 25(4):550–561

    Article  PubMed  CAS  Google Scholar 

Download references

Acknowledgement

We gratefully acknowledge support from Siemens Medical Solutions who provided a Sensation 64 CT scanner to our institute for experimental and clinical work.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Willi A. Kalender.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kyriakou, Y., Kachelrieβ, M., Knaup, M. et al. Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner. Eur Radiol 16, 1206–1215 (2006). https://doi.org/10.1007/s00330-005-0118-9

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-005-0118-9

Keywords

Navigation