Abstract
The effect of the z-flying focal spot (zFFS) technology was evaluated by simulations and measurements with respect to resolution and artifact behavior for a 64-slice spiral cone-beam computed tomography (CT) scanner. The zFFS alternates between two z-positions of the X-ray focal spot, acquiring two slices per detector row, which results in double sampling in the z-direction. We implemented a modified reconstruction that is able to obtain images as they would be without zFFS. A delta phantom equipped with a thin gold disc was used to measure slice sensitivity profiles (SSP), and a high-contrast bar phantom was used to quantify the resolution in the x/z-plane with and without zFFS. The zFFS decreases the full width at half maximum (FWHM) of the SSPs by a factor of about 1.4. The double z-sampling allows the separation of 0.4 mm bars in the z-direction compared with 0.6 mm in the case without zFFS. The zFFS effectively reduces windmill artifacts in the reconstructed images while maintaining the transverse resolution, even at the largest available pitch value of 1.5.
Similar content being viewed by others
References
Silver MD, Taguchi K, Hein IA, Han HS, Kazama M, Mori I (2003) Windmill artifact in multislice CT. Proc SPIE 5032:1918–1927
Taguchi K, Aradate H, Saito Y (2004) The cause of the artifact in 4-slice helical computed tomography. Med Phys 31(7):2033–2037
Kalender WA (2005) Computed Tomography, 2nd edn. Wiley-VCH, New York
Hsieh J (2003) Analytical models for multi-slice helical CT performance parameters. Med Phys 30(2):169–178
Schardt P, Deuringer J, Freudenberger J, Hell E, Knuepfer W, Mattern D, Schild M (2004) New x-ray tube performance in computed tomography by introducing the rotating envelope tube. Med Phys 32(9):2699–2706
Kachelrieß M, Knaup M, Penßel C, Kalender WA (2005) Flying focal spot (FFS) in cone-beam CT. In: Records of the 2004 IEEE Medical Imaging Conference pp 3759–3763
Flohr T, Stierstorfer K, Ulzheimer S, Bruder H, Primak AN, McCollough CH (2005) Image reconstruction and image quality evaluation of a 64-slice CT scanner with focal spot. Med Phys 32(8):2536–2547
Kachelrieß M, Schaller S, Kalender WA (2000) Advanced single-slice rebinning in cone-beam spiral CT”. Med Phys 27:754–772
Kachelrieß M, Kalender WA (2005) Presampling, algorithm factors, and noise in CT. Med Phys 32(5):1321–1334
Taguchi T, Aradate H (1998) Algorithm for image reconstruction in multislice helical CT. Med Phys 25(4):550–561
Acknowledgement
We gratefully acknowledge support from Siemens Medical Solutions who provided a Sensation 64 CT scanner to our institute for experimental and clinical work.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kyriakou, Y., Kachelrieβ, M., Knaup, M. et al. Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner. Eur Radiol 16, 1206–1215 (2006). https://doi.org/10.1007/s00330-005-0118-9
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-005-0118-9