Fatal outcome in atrial migration of the tempofilter

  • Clinical Investigations
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose: To report the risk of fatal atrial migration with the Tempofilter.

Methods: Among temporary filters, the high safety profile Tempofilter has been marketed as offering protection for up to 6 weeks. We implanted about 60 Tempofilters to prevent pulmonary embolism. The main indications were temporary thromboembolic risk, recurrent pulmonary embolism, and contraindication to or failure of anticoagulant therapy. Follow-up was performed regularly by plain abdominal film and Doppler ultrasound. Filters were removed about 4 weeks after placement.

Results: We encountered three cases (5%) of atrial migration and one case of 5-cm cephalad displacement of the filter. Of the three patients with atrial migration, two died within 3 days of implantation, one from a massive pulmonary embolism and the other with cardiac tamponade. One patient did not show any serious complications.

Conclusions: The Tempofilter may actively migrate cranially and become dangerous in the case of migration within the heart.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Dorfman GS (1990) Percutaneous inferior vena cava filters. Radiology 174:987–992

    PubMed  CAS  Google Scholar 

  2. Grassi CJ (1991) Inferior vena caval filters: Analysis of five currently available devices. AJR 156:813–821

    PubMed  CAS  Google Scholar 

  3. Millward SF,Marsh JI, Peterson RA, Rasuli P, French G, Wilson CM, Tennenhouse JE, Morris DC, Simons ME, Formoso AF (1991) LGM (Vena Tech) vena cava filter: Clinical experience in 64 patients. J Vasc Interv Radiol 2:429–433

    Article  PubMed  CAS  Google Scholar 

  4. Becker DM, Philbrick JT, Selby JB (1992) Inferior vena cav filters: Indications, safety, effectiveness. Arch Intern Med 152:1985–1994

    Article  PubMed  CAS  Google Scholar 

  5. McCowan TC, Ferris EJ, Carver DK, Molpus WM (1992) Complications of the nitinol vena cava filter. J Vasc Interv Radiol 3:401–408

    Article  PubMed  CAS  Google Scholar 

  6. Ferris EJ, McCowan TC, Carver DK, McFarland DR (1993) Percutaneous inferior vena cava filters: Follow-up of seven designs in 320 patients. Radiology 188:851–856

    PubMed  CAS  Google Scholar 

  7. Vos LD, Tielbeek AV, Bom EP, Gooszen HC, Vroegindeweij D (1997) The Gunther temporary inferior vena cava filter for short-term protection against pulmonary embolism. Cardiovasc Intervent Radiol 20:91–97

    Article  PubMed  CAS  Google Scholar 

  8. Buecker A, Neuerburg J, Schmitz-Rode T, Vorwerk D, Guenther RW (1997) In vitro evaluation of a rheolytic thrombectomy system for clot removal from five different temporary vena cava filters. Cardiovasc Intervent Radiol 20:448–451

    Article  PubMed  CAS  Google Scholar 

  9. Dick A, Neuerburg J, Schmitz-Rode T, Alliger H, Günther RW (1998) Declotting of embolized temporary vena cava filter by ultrasound and the Angiojet: Comparative experimental in vitro studies. Invest Radiol 33:91–97

    Article  PubMed  CAS  Google Scholar 

  10. Kuszyk BS, Venbrux AC, Samphilipo MA Jr, Magee CA, Olson JL, Osterman FA Jr (1995) Subcutaneously tethered temporary filter: Pathologic effects in swine. J Vasc Interv Radiol 6:895–902

    Article  PubMed  CAS  Google Scholar 

  11. Bovyn G, Gory P, Reynaud P, Ricco JB (1997) The Tempofilter: A multicenter study of a new temporary caval filter implantable for up to six weeks. Ann Vasc Surg 11:520–528

    Article  PubMed  CAS  Google Scholar 

  12. Roehm JOGF, Johnsrude IS, Barth KH, Gianturco C (1988) The Bird’s Nest inferior vena cava filter: Progress report. Radiology 168:745–749

    PubMed  Google Scholar 

  13. Vorwerk D, Schmitz-Rode T, Schurmann K, Tacke J, Guenther RW (1995) Use of a temporary caval filter to assist percutaneous iliocaval thrombectomy: Experimental results. J Vasc Interv Radiol 6:737–740

    Article  PubMed  CAS  Google Scholar 

  14. Magnant JG, Walsh DB, Juravsky LI, Cronenwett JL (1992) Current use of inferior vena cava filters. J Vasc Surg 16:701–706

    Article  PubMed  CAS  Google Scholar 

  15. Cope C, Baum RA, Duszak (1996) Temporary use of a Bird’s Nest filter during iliocaval thrombolysis. Radiology 198:765–767

    PubMed  CAS  Google Scholar 

  16. Théry C, Asseman P, Amrouni N, Becquart J, Lesenne M, Legghe R, Marache P (1990) Use of a new removable vena cava filter in order to prevent pulmonary embolism in patients submitted to thrombolysis. Eur Heart J 11:334–341

    PubMed  Google Scholar 

  17. Nakagawa N, Cragg AH, Smith TP, Castañeda F, Barnhart WH, De Jong SC (1994) A retrievable nitinol vena cava filter: Experimental and initial clinical results. J Vasc Interv Radiol 5:507–512

    Article  PubMed  CAS  Google Scholar 

  18. Rogoff PA, Hilgenberg AD, Miller SL, Stepham SM (1992) Cephalic migration of the Bird’s Nest inferior vena caval filter: Report of two cases. Radiology 184:819–823

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Rossi, P., Arata, F.M., Bonaiuti, P. et al. Fatal outcome in atrial migration of the tempofilter. Cardiovasc Intervent Radiol 22, 227–231 (1999). https://doi.org/10.1007/s002709900372

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002709900372

Key words

Navigation