Abstract

Introduction

The usefulness of carbon dioxide (CO2) gas digital subtraction angiography (DSA) has been reported for patients with renal insufficiency and allergy to iodinated contrast agents. However, CO2 gas cannot replace the iodinated contrast agent in all cases owing to some disadvantages. We describe balloon-occluded CO2 DSA (B-CO2 DSA) as an improved CO2 DSA procedure for interventions in the internal iliac artery (IIA) region and compare the quality of images obtained using conventional CO2 DSA and B-CO2 DSA.

Materials and Methods

B-CO2 DSA-guided embolization was performed for one case of genital bleeding with an acute anaphylactic reaction to the iodinated contrast agent and for three cases of type II endoleaks after endovascular abdominal aortic aneurysm repair with renal dysfunction. A 9-mm occlusion balloon catheter was placed just after the orifice of the IIA. Then, 10–15 ml of CO2 gas was injected manually via the catheter with and without balloon occlusion. The quality of sequential digital subtraction angiograms was analyzed based on a scoring criterion.

Results

In all four cases, image quality was improved with B-CO2 DSA; the poor quality of images without balloon occlusion was because of reflux of the CO2 gas.

Conclusions

B-CO2 DSA improves the image quality of CO2 DSA in the IIA region and is useful for vascular intervention.

Level of Evidence

Level IV.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

References

  1. Hawkins IF. Carbon dioxide digital subtraction arteriography. Am J Roentgenol. 1982;139:19–24.

    Article  CAS  Google Scholar 

  2. Kessel DO, Robertson I, Patel J, Peters K, Taylor EJ, Denton LJ, et al. Carbon-dioxide-guided vascular interventions: technique and pitfalls. Cardiovasc Intervent Radiol. 2002;25:476–83.

    Article  PubMed  Google Scholar 

  3. Fujihara M, Kawasaki D, Shintani Y, Fukunaga M, Nakama T, Koshida R, et al. Endovascular therapy by CO2 angiography to prevent contrast-induced nephropathy in patients with chronic kidney disease: a prospective multicenter trial of CO2 angiography registry. Catheter Cardiovasc Interv. 2015;85:870–7.

    Article  PubMed  Google Scholar 

  4. Rundback JH, Shah PM, Wong J, Babu SC, Rozenblit G, Poplausky MR. Livedo reticularis, rhabdomyolysis, massive intestinal infarction, and death after carbon dioxide arteriography. J Vasc Surg. 1997;26:337–40.

    Article  PubMed  CAS  Google Scholar 

  5. Caridi JG, Hawkins IF. CO2 digital subtraction angiography: potential complications and their prevention. J Vasc Interv Radiol. 1997;8:383–91.

    Article  PubMed  CAS  Google Scholar 

  6. Huang SG, Woo K, Moos JM, Han S, Lew WK, Chao A, et al. A prospective study of carbon dioxide digital subtraction versus standard contrast arteriography in the detection of endoleaks in endovascular abdominal aortic aneurysm repairs. Ann Vasc Surg. 2013;27:38–44.

    Article  PubMed  Google Scholar 

  7. Cho KJ. Carbon dioxide angiography: scientific principles and practice. Vasc Specialist Int. 2015;31:67–80.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Song K, Cho D, Shinn K, Charlton E, Cho K. Gas dynamics in CO2 angiography: in vitro evaluation in a circulatory system model. Invest Radiol. 1999;34:151–5.

    Article  PubMed  CAS  Google Scholar 

  9. Taylor FC, Smith DC, Watkins GE, Kohne RE, Suh RD. Balloon occlusion versus wedged hepatic venography using carbon dioxide for portal vein opacification during TIPS. Cardiovasc Interv Radiol. 1999;22:150–1.

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Mitsuhiro Kishino.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical Approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Informed Consent

Informed consent was obtained from all individual participants included in the study.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kishino, M., Nakaminato, S., Kitazume, Y. et al. Balloon-Occluded Carbon Dioxide Gas Angiography for Internal Iliac Arteriography and Intervention. Cardiovasc Intervent Radiol 41, 1106–1111 (2018). https://doi.org/10.1007/s00270-018-1957-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-018-1957-5

Keywords

Navigation