Abstract
Purpose
To investigate liver stiffness changes—evaluated by point shear wave elastography (pSWE)—in controlled hyperthermia and microwave ablation (MWA) in an ex vivo animal model.
Materials and Methods
Five samples of ex vivo bovine liver were uniformly heated to temperatures ranging from 40 to 100 °C. B-mode ultrasound imaging and pSWE were acquired simultaneously, and shear wave velocity (SWV) was measured in a region of interest (ROI). The threshold value of SWV at 60 °C (avg60) was identified. Subsequently, MWA was performed in 11 liver samples at 60 W until avg60 + 0.5 m/s was reached. SWV was measured in ROIs at 10–40 mm from the antenna feed. The correlation of mean values of SWV with location (within, border, or outside necrotic area) at gross pathology was evaluated.
Results
In controlled hyperthermia experiments, a steep transition in liver stiffness was observed at 63.0 ± 2.4 °C (SWV 3.54 ± 0.68 m/s). Avg60 was of 2.5 m/s. In 8/9 MWA experiments, interrupted when SWV of 3 m/s was measured, the ROI was at the inner side of the necrotic area border at pathology (accuracy 89%). No correlation between SWV values for outside, border, and within necrosis could be identified.
Conclusions
pSWE can provide a velocity threshold predictive of the presence of coagulation necrosis during MWA in ex vivo liver model. However, pSWE is not able to reliably capture changes in stiffness within, at the border, and outside the necrotic zone in this experimental model.
Similar content being viewed by others
References
Ahmed M, Solbiati L, Brace CL, et al. International working group on image-guided tumor ablation; interventional oncology Sans Frontières expert panel; technology assessment committee of the society of interventional radiology; standard of practice committee of the cardiovascular and interventional radiological society of Europe. Image-guided tumor ablation: standardization of terminology and reporting criteria—a 10-year update. J Vasc Interv Radiol. 2014;25(11):1691–705.
Zhou Z, Wu W, Wu S, et al. A survey of ultrasound elastography approaches to percutaneous ablation monitoring. Proc Inst Mech Eng H. 2014;228(10):1069–82.
Brosses ES, Pernot M, Tanter M. The link between tissue elasticity and thermal dose in vivo. Phys Med Biol. 2011;56(24):7755–65.
Barr RG, Ferraioli G, Palmeri ML, et al. Elastography assessment of liver fibrosis: society of radiologists in ultrasound consensus conference statement. Radiology. 2015;276(3):845–61.
Wu T, Felmlee JP, Greenleaf JF, et al. Assessment of thermal tissue ablation with MR elastography. Magn Reson Med. 2001;45(1):80–7.
Brosses ES, Gennisson JL, Pernot M, et al. Temperature dependence of the shear modulus of soft tissues assessed by ultrasound. Phys Med Biol. 2010;55(6):1701–18.
Mariani A, Kwiecienski W, Pernot M. Real time shear waves elastography monitoring of thermal ablation: in vivo evaluation in pig livers. J Surg Res. 2014;188(1):37–43.
Ferraioli G, Tinelli C, Lissandrin R, et al. Point shear wave elastography method for assessing liver stiffness. World J Gastroenterol. 2014;20(16):4787–96.
Kennedy P, Wagner M, Castéra L, et al. Quantitative elastography methods in liver disease: current evidence and future directions. Radiology. 2018;286(3):738–63.
Rossmann C, Garrett-Mayer E, Rattay F, et al. Dynamics of tissue shrinkage during ablative temperature exposures. Physiol Meas. 2014;35(1):55–67.
Lepetit J. A theoretical approach of the relationships between collagen content, collagen cross-links and meat tenderness. Meat Sci. 2007;76(1):147–59.
Jiang J, Varghese T, Brace TL, et al. Young’s modulus reconstruction for radio-frequency ablation electrode-induced displacement fields: a feasibility study. IEEE Trans Med Imaging. 2009;28(8):1325–34.
Zhu Y, Hall TJ, Jiang J. A finite-element approach for Young’s modulus reconstruction. IEEE Trans Med Imaging. 2003;22(7):890–901.
Arnal B, Pernot M, Tanter M. Monitoring of thermal therapy based on shear modulus changes: II. Shear wave imaging of thermal lesions. IEEE Trans Ultrason Ferroelectr Freq Control. 2011;58(8):1603–11.
Sugimoto K, Oshiro H, Ogawa S, et al. Radiologic-pathologic correlation of three-dimensional shear-wave elastographic findings in assessing the liver ablation volume after radiofrequency ablation. World J Gastroenterol. 2014;20(33):11850–5.
Dewall RJ, Varghese T, Brace TL. Visualizing ex vivo radiofrequency and microwave ablation zones using electrode vibration elastography. Med Phys. 2012;39(11):6692–700.
Bo XW, Li XL, Xu HX, et al. 2D shear-wave ultrasound elastography (SWE) evaluation of ablation zone following radiofrequency ablation of liver lesions: is it more accurate? Br J Radiol. 2016;89(1060):20150852.
Correa-Gallego C, Karkar AM, Monette S, et al. Intraoperative ultrasound and tissue elastography measurements do not predict the size of hepatic microwave ablations. Acad Radiol. 2014;21(1):72–8.
Farina L, Weiss N, Nissenbaum Y, et al. Characterisation of tissue shrinkage during microwave thermal ablation. Int J Hyperth. 2014;30(7):419–28.
Brace CL, Diaz TA, Hinshaw JL, et al. Tissue contraction caused by radiofrequency and microwave ablation: a laboratory study in liver and lung. J Vasc Interv Radiol. 2010;21:1280–6.
Yang D, Converse MC, Mahvi DM, et al. Measurement and analysis of tissue temperature during microwave liver ablation. IEEE Trans Biomed Eng. 2007;54:150–5.
Yang D, Converse MC, Mahvi DM, et al. Expanding the bioheat equation to include tissue internal water evaporation during heating. IEEE Trans Biomed Eng. 2007;54:1382–8.
Tsui PH, Wang CY, Zhou Z, et al. Monitoring radiofrequency ablation using ultrasound envelope statistics and shear wave elastography in the periablation period: an in vitro feasibility study. PLoS ONE. 2016;11(9):e0162488.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Electronic Supplementary Material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Crocetti, L., Calcagni, F., Gherarducci, G. et al. Monitoring of Thermal-Induced Changes in Liver Stiffness During Controlled Hyperthermia and Microwave Ablation in an Ex Vivo Bovine Model Using Point Shear Wave Elastography. Cardiovasc Intervent Radiol 42, 744–750 (2019). https://doi.org/10.1007/s00270-018-02152-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-018-02152-y