Optimizing Needle Direction During Transjugular Liver Biopsy Provides Superior Biopsy Specimens

  • Clinical Investigation
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose

Transjugular liver biopsy (TJLB) is commonly performed for staging of liver fibrosis and cirrhosis among patients with coagulopathy and/or ascites. We hypothesized that device orientation during needle firing influences hepatic tissue apposition with the specimen notch and specimen quality.

Methods

Needle biopsies were performed in ex vivo bovine livers with specimen notch of the biopsy device oriented at cranial, caudal, or lateral directions with respect to the guiding metal cannula. Biopsy specimen length was measured and evaluated for fragmentation using light microscopy. In addition, a consecutive cohort of patients (n = 50) who underwent TJLB with random (n = 22) or caudal (n = 28) needle orientation was retrospectively reviewed. The number of needle passes was documented, and pathology specimen adequacy was graded using an ordinal scale.

Results

A total of 400 biopsies were performed (100 in each orientation) in ex vivo bovine livers. Longer specimens were obtained with caudal orientation of the needle specimen notch (p < 0.0001, ANOVA and Kruskal–Wallis tests). There was no difference in the degree of fragmentation. In the retrospective clinical study, specimen adequacy was significantly higher among patients in the caudal orientation group (p = 0.0002, Mann–Whitney U test).

Conclusion

Caudal orientation of the needle specimen notch of the biopsy device during TJLB produces superior core biopsy specimens. This simple technical modification may assist in obtaining higher-quality biopsy specimens during TJLB.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Cholongitas E, Senzolo M, Standish R et al (2006) A systematic review of the quality of liver biopsy specimens. Am J Clin Pathol 125:710–721

    Article  PubMed  Google Scholar 

  2. Cholongitas E, Quaglia A, Samonakis D et al (2007) Transjugular liver biopsy in patients with diffuse liver disease: comparison of three cores with one or two cores for accurate histological interpretation. Liver Int 27:646–653

    Article  CAS  PubMed  Google Scholar 

  3. Vibhokorn S, Cholongitas A, Kalambokis G et al (2009) A comparison of four- versus three-pass transjugular biopsy using a 19-G Tru-Cut needle and a randomized study using a cassette to prevent biopsy fragmentation. Cardiovasc Intervent Radiol 32:508–513

    Article  Google Scholar 

  4. Gamble P, Colapinto RF, Stronell RD et al (1985) Transjugular liver biopsy: a review of 461 biopsies. Radiology 157:589–593

    Article  CAS  PubMed  Google Scholar 

  5. Lebrec D, Goldfarb G, Degott C et al (1982) Transvenous liver biopsy: an experience based on 1000 hepatic tissue samplings with this procedure. Gastroenterology 83:338–340

    CAS  PubMed  Google Scholar 

  6. Bruzzi JF, O’Connell MJ, Thakore H et al (2002) Transjugular liver biopsy: assessment of safety and efficacy of the quick-core biopsy needle. Abdom Imaging 27:711–715

    Article  CAS  PubMed  Google Scholar 

  7. Smith TP, Presson TL, Heneghan MA et al (2003) Transjugular biopsy of the liver in pediatric and adult patients using an 18-G automated core biopsy needle: a retrospective review of 410 consecutive procedures. AJR 180:167–172

    Article  PubMed  Google Scholar 

  8. Ishikawa T, Kamimura H, Tsuchiya A et al (2006) Comparison of a new aspiration needle device and the quick-core biopsy needle for transjugular liver biopsy. World J Gastroenterol 21(12):6339–6342

    Google Scholar 

  9. Bravo AA, Sheth SG, Chopra S (2001) Liver biopsy. N Engl J Med 344:495–500

    Article  CAS  PubMed  Google Scholar 

  10. Colloredo G, Guido M, Sonzogni A et al (2003) Impact of liver biopsy size on histological evaluation of chronic viral hepatitis: the smaller the sample, the milder the disease. J Hepatol 39:239–244

    Article  PubMed  Google Scholar 

  11. Bedossa P, Dargere D, Paradis V (2003) Sampling variability of liver fibrosis in chronic hepatitis C. Hepatology 38:1449–1457

    Article  PubMed  Google Scholar 

Download references

Conflict of interest

The authors declare that they have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Timothy W. I. Clark.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Clark, T.W.I., McCann, J.W., Salsamendi, J. et al. Optimizing Needle Direction During Transjugular Liver Biopsy Provides Superior Biopsy Specimens. Cardiovasc Intervent Radiol 37, 1540–1545 (2014). https://doi.org/10.1007/s00270-013-0819-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-013-0819-4

Keywords

Navigation