Abstract
Purpose
Transjugular liver biopsy (TJLB) is commonly performed for staging of liver fibrosis and cirrhosis among patients with coagulopathy and/or ascites. We hypothesized that device orientation during needle firing influences hepatic tissue apposition with the specimen notch and specimen quality.
Methods
Needle biopsies were performed in ex vivo bovine livers with specimen notch of the biopsy device oriented at cranial, caudal, or lateral directions with respect to the guiding metal cannula. Biopsy specimen length was measured and evaluated for fragmentation using light microscopy. In addition, a consecutive cohort of patients (n = 50) who underwent TJLB with random (n = 22) or caudal (n = 28) needle orientation was retrospectively reviewed. The number of needle passes was documented, and pathology specimen adequacy was graded using an ordinal scale.
Results
A total of 400 biopsies were performed (100 in each orientation) in ex vivo bovine livers. Longer specimens were obtained with caudal orientation of the needle specimen notch (p < 0.0001, ANOVA and Kruskal–Wallis tests). There was no difference in the degree of fragmentation. In the retrospective clinical study, specimen adequacy was significantly higher among patients in the caudal orientation group (p = 0.0002, Mann–Whitney U test).
Conclusion
Caudal orientation of the needle specimen notch of the biopsy device during TJLB produces superior core biopsy specimens. This simple technical modification may assist in obtaining higher-quality biopsy specimens during TJLB.
Similar content being viewed by others
References
Cholongitas E, Senzolo M, Standish R et al (2006) A systematic review of the quality of liver biopsy specimens. Am J Clin Pathol 125:710–721
Cholongitas E, Quaglia A, Samonakis D et al (2007) Transjugular liver biopsy in patients with diffuse liver disease: comparison of three cores with one or two cores for accurate histological interpretation. Liver Int 27:646–653
Vibhokorn S, Cholongitas A, Kalambokis G et al (2009) A comparison of four- versus three-pass transjugular biopsy using a 19-G Tru-Cut needle and a randomized study using a cassette to prevent biopsy fragmentation. Cardiovasc Intervent Radiol 32:508–513
Gamble P, Colapinto RF, Stronell RD et al (1985) Transjugular liver biopsy: a review of 461 biopsies. Radiology 157:589–593
Lebrec D, Goldfarb G, Degott C et al (1982) Transvenous liver biopsy: an experience based on 1000 hepatic tissue samplings with this procedure. Gastroenterology 83:338–340
Bruzzi JF, O’Connell MJ, Thakore H et al (2002) Transjugular liver biopsy: assessment of safety and efficacy of the quick-core biopsy needle. Abdom Imaging 27:711–715
Smith TP, Presson TL, Heneghan MA et al (2003) Transjugular biopsy of the liver in pediatric and adult patients using an 18-G automated core biopsy needle: a retrospective review of 410 consecutive procedures. AJR 180:167–172
Ishikawa T, Kamimura H, Tsuchiya A et al (2006) Comparison of a new aspiration needle device and the quick-core biopsy needle for transjugular liver biopsy. World J Gastroenterol 21(12):6339–6342
Bravo AA, Sheth SG, Chopra S (2001) Liver biopsy. N Engl J Med 344:495–500
Colloredo G, Guido M, Sonzogni A et al (2003) Impact of liver biopsy size on histological evaluation of chronic viral hepatitis: the smaller the sample, the milder the disease. J Hepatol 39:239–244
Bedossa P, Dargere D, Paradis V (2003) Sampling variability of liver fibrosis in chronic hepatitis C. Hepatology 38:1449–1457
Conflict of interest
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Clark, T.W.I., McCann, J.W., Salsamendi, J. et al. Optimizing Needle Direction During Transjugular Liver Biopsy Provides Superior Biopsy Specimens. Cardiovasc Intervent Radiol 37, 1540–1545 (2014). https://doi.org/10.1007/s00270-013-0819-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00270-013-0819-4