Skip to main content
Log in

Inferior Mesenteric Artery Embolization Before Endovascular Aortic Aneurysm Repair Using Amplatzer Vascular Plug Type 4

  • Clinical Investigation
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose

This study was designed to evaluate the efficacy and safety of Amplatzer Vascular Plug type 4 (AVP-4) for embolization of the inferior mesenteric artery (IMA) before endovascular aneurysm repair (EVAR) of the abdominal aorta to prevent endoleaks.

Methods

A single-center retrospective review of 31 patients who underwent IMA embolizations before EVAR using the AVP-4 was performed. We analyzed the insertion and detachment procedure, the technical success, and the final position of the plug. Technical success was defined as complete occlusion of the IMA. To compare the incidence of IMA-related type II endoleaks in patients with and without preoperative IMA embolization, we additionally reviewed the course of 43 patients with a preoperatively patent IMA who underwent no IMA embolization.

Results

Plugs with a diameter of 5, 6, and 8 mm were used in 5 (16.1 %), 21 (67.7 %), and 5 (16.1 %) patients, respectively (50–100 % oversizing). In 29 of 31 patients (93.5 %), we observed complete occlusion of the IMA within 10 min (mean 5.1 min). Precise placement of the plug in the proximal segment of the IMA without occlusion of the first IMA branches was achievable in all patients. The distance between the AVP-4 and the first branches was on average 12 (range 2–57) mm. Preoperative IMA embolization with AVP-4 significantly reduced the incidence of complex IMA-lumbar type II endoleaks after EVAR (0/31 vs. 11/43; p = 0.002).

Conclusions

The AVP-4 is a safe, feasible, and technically effective embolization device for IMA embolization before EVAR.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Veith FJ, Baum RA, Ohki T et al (2002) Nature and significance of endoleaks and endotension: summary of opinions expressed at an international conference. J Vasc Surg 35:1029–1035

    Article  PubMed  Google Scholar 

  2. Velazquez OC, Baum RA, Carpenter JP et al (2000) Relationship between preoperative patency of the inferior mesenteric artery and subsequent occurrence of type II endoleak in patients undergoing endovascular repair of abdominal aortic aneurysms. J Vasc Surg 32(4):777–788

    Article  CAS  PubMed  Google Scholar 

  3. Jones JE, Atkins MD, Brewster DC et al (2007) Persistent type 2 endoleak after endovascular repair of abdominal aortic aneurysm is associated with adverse late outcomes. J Vasc Surg 46(1):1–8

    Article  PubMed  Google Scholar 

  4. Sampaio SM, Panneton JM, Mozes GI et al (2005) Aneurysm sac thrombus load predicts type II endoleaks after endovascular aneurysm repair. Ann Vasc Surg 19(3):302–309

    Article  PubMed  Google Scholar 

  5. Fan CM, Rafferty EA, Geller SC et al (2001) Endovascular stent-graft in abdominal aortic aneurysms: the relationship between patent vessels that arise from the aneurysmal sac and early endoleak. Radiology 218(1):176–182

    Article  CAS  PubMed  Google Scholar 

  6. Abularrage CJ, Crawford RS, Conrad MF et al (2010) Preoperative variables predict persistent type 2 endoleak after endovascular aneurysm repair. J Vasc Surg 52(1):19–24

    Article  PubMed  Google Scholar 

  7. AbuRahma AF, Mousa AY, Campbell JE et al (2011) The relationship of preoperative thrombus load and location to the development of type II endoleak and sac regression. J Vasc Surg 53(6):1534–1541

    Article  PubMed  Google Scholar 

  8. Brountzos E, Karagiannis G, Panagiotou I et al (2012) Risk factors for the development of persistent type II endoleaks after endovascular repair of infrarenal abdominal aortic aneurysms. Diagn Interv Radiol 18(3):307–313

    PubMed  Google Scholar 

  9. Axelrod DJ, Lookstein RA, Guller J et al (2004) Inferior mesenteric artery embolization before endovascular aneurysm repair: technique and initial results. J Vasc Interv Radiol 15(11):1263–1267

    Article  PubMed  Google Scholar 

  10. Bonvini R, Alerci M, Antonucci F et al (2003) Preoperative embolization of collateral side branches: a valid means to reduce type II endoleaks after endovascular AAA repair. J Endovasc Ther 10(2):227–232

    Article  PubMed  Google Scholar 

  11. Muthu C, Maani J, Plank LD et al (2007) Strategies to reduce the rate of type II endoleaks: routine intraoperative embolization of the inferior mesenteric artery and thrombin injection into the aneurysm sac. J Endovasc Ther 14(5):661–668

    Article  PubMed  Google Scholar 

  12. Nevala T, Biancari F, Manninen H et al (2010) Inferior mesenteric artery embolization before endovascular repair of an abdominal aortic aneurysm: effect on type II endoleak and aneurysm shrinkage. J Vasc Interv Radiol 21(2):181–185

    Article  PubMed  Google Scholar 

  13. Parry DJ, Kessel DO, Robertson I et al (2002) Type II endoleaks: predictable, preventable, and sometimes treatable? J Vasc Surg 36(1):105–110

    Article  PubMed  Google Scholar 

  14. Ward TJ, Cohen S, Fischman AM et al (2013) Preoperative inferior mesenteric artery embolization before endovascular aneurysm repair: decreased incidence of type II endoleak and aneurysm sac enlargement with 24-month follow-up. J Vasc Interv Radiol 24(1):49–55

    Article  PubMed  Google Scholar 

  15. Buttarelli L, Capocasale E, Marcato C et al (2011) Embolization of pancreatic allograft arteriovenous fistula with the Amplatzer Vascular Plug 4: case report and literature analysis. Transplant Proc 43(10):4044–4047

    Article  CAS  PubMed  Google Scholar 

  16. Ferro C, Rossi UG, Bovio G et al (2010) The Amplatzer Vascular Plug 4: preliminary experience. Cardiovasc Interv Radiol 33(4):844–848

    Article  Google Scholar 

  17. Mordasini P, Szucs-Farkas Z, Do DD et al (2010) Use of a latest-generation vascular plug for peripheral vascular embolization with use of a diagnostic catheter: preliminary clinical experience. J Vasc Interv Radiol 21(8):1185–1190

    Article  PubMed  Google Scholar 

  18. Ng EH, Comin J, David E et al (2012) Amplatzer vascular plug 4 for proximal splenic artery embolization in blunt trauma. J Vasc Interv Radiol 23(7):976–979

    Article  PubMed  Google Scholar 

  19. Pech M, Mohnike K, Wieners G et al (2011) Advantages and disadvantages of the Amplatzer Vascular Plug IV in visceral embolization: report of 50 placements. Cardiovasc Interv Radiol 34(5):1069–1073

    Article  Google Scholar 

  20. Rossi UG, Seitun S, Ferro C (2013) Endovascular embolization of a third jejunal artery aneurysm: isolation technique using the Amplatzer Vascular Plug 4. Catheter Cardiovasc Interv 81(6):1049–1052

    Article  PubMed  Google Scholar 

  21. Yildiz AE, Peynircioglu B, Cil BE (2012) Applications of the Amplatzer Vascular Plug 4. Diagn Interv Radiol 18(2):225–230

    PubMed  Google Scholar 

  22. Sacks D, McClenny TE, Cardella JF et al (2003) Society of interventional radiology clinical practice guidelines. J Vasc Interv Radiol 14(9 Pt 2):S199–S202

    Article  PubMed  Google Scholar 

  23. Michels NA, Siddharth P, Kornblith PL et al (1965) The variant blood supply to the descending colon, rectosigmoid and rectum based on 400 dissections. Its importance in regional resections—a review of medical literature. Dis Colon Rectum 8:251–278

    Article  CAS  PubMed  Google Scholar 

  24. Pech M, Kraetsch A, Wieners G et al (2009) Embolization of the gastroduodenal artery before selective internal radiotherapy: a prospectively randomized trial comparing platinum-fibered microcoils with the Amplatzer Vascular Plug II. Cardiovasc Interv Radiol 32(3):455–461

    Article  Google Scholar 

Download references

Conflict of interest

The authors declare that they have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to René Müller-Wille.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Müller-Wille, R., Uller, W., Gößmann, H. et al. Inferior Mesenteric Artery Embolization Before Endovascular Aortic Aneurysm Repair Using Amplatzer Vascular Plug Type 4. Cardiovasc Intervent Radiol 37, 928–934 (2014). https://doi.org/10.1007/s00270-013-0762-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-013-0762-4