Abstract
Purpose
In emergency endovascular repair (EVAR) of ruptured aneurysms of the aorta (rAAA), anatomical suitability must be determined. Semiautomatic three-dimensional assessment of the aortoiliac arteries has the potential to standardise measurements. This study assesses the fitness for purpose of such a semiautomatic approach for rAAA and determined interobserver agreement on suitability.
Materials and Methods
Interobserver study with six trained observers (4 vascular surgeons, 2 radiologists) blindly assessing preoperative computed tomography angiography scans of 50 consecutive patients with rAAA. A central lumen line (CLL) was generated, and perpendicular diameters, length along the CLL, and EVAR suitability were determined using dedicated sizing software (3mensio; 3mensio Vascular; Bilthoven, The Netherlands). Success of generating a CLL, time of assessment, and interobserver agreement was determined.
Results
In the majority of the patients (median 76 %, range 64–78 %), a CLL was semiautomatically generated. The median duration of CLL generation and performance measurements was 7.5 min (interquartile range 5.5–10.6). Agreement on suitability was moderate for the entire group (Fleiss’ κ = 0.55, confidence interval 0.48–0.62) and ranged from moderate to good (Cohen’s κ = 0.40–0.72) between observer pairs.
Conclusion
Assessing EVAR suitability of rAAA patients using dedicated sizing software is possible in the majority of patients. The measurements can be performed in a reasonable amount of time, and the agreement of suitability for EVAR in patients with rAAA is moderate. Improvements and additional research are necessary to replace the current axial measurement.
Similar content being viewed by others
References
Richards T, Goode SD, Hinchliffe R, Altaf N, Macsweeney S, Braithwaite B (2009) The importance of anatomical suitability and fitness for the outcome of endovascular repair of ruptured abdominal aortic aneurysm. Eur J Vasc Endovasc Surg 38(3):285–290
van Keulen JW, Moll FL, van Herwaarden JA (2010) Tips and techniques for optimal stent graft placement in angulated aneurysm necks. J Vasc Surg 52(4):1081–1086
Ghatwary TM, Patterson BO, Karthikesalingam A, Hinchliffe RJ, Loftus IM, Morgan R et al (2013) A systematic review of protocols for the three-dimensional morphologic assessment of abdominal aortic aneurysms using computed tomographic angiography. Cardiovasc Intervent Radiol 36(1):14–24
Moll FL, Powell JT, Fraedrich G, Verzini F, Haulon S, Waltham M et al (2011) Management of abdominal aortic aneurysms clinical practice guidelines of the European Society for Vascular Surgery. Eur J Vasc Endovasc Surg 41(Suppl 1):S1–S58
Chaikof EL, Fillinger MF, Matsumura JS, Rutherford RB, White GH, Blankensteijn JD et al (2002) Identifying and grading factors that modify the outcome of endovascular aortic aneurysm repair. J Vasc Surg 35(5):1061–1066
Amsterdam Acute Aneurysm trial Collaborators (2000) Amsterdam acute aneurysm trial: background, design, and methods. Vascular 14(3):130–135
Reimerink JJ, Hoornweg LL, Vahl AC, Wisselink W, van den Broek TA, Legemate DA et al (2013) Endovascular repair versus open repair of ruptured abdominal aortic aneurysms: a multicenter randomized controlled trial. Ann Surg 258:248–256
van Keulen JW, Moll FL, Tolenaar JL, Verhagen HJ, van Herwaarden JA (2010) Validation of a new standardized method to measure proximal aneurysm neck angulation. J Vasc Surg 51(4):821–828
Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86(2):420–428
Sim J, Wright CC (2005) The kappa statistic in reliability studies: use, interpretation, and sample size requirements. Phys Ther 85(3):257–268
Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33(1):159–174
Diehm N, Baumgartner I, Silvestro A, Herrmann P, Triller J, Schmidli J et al (2005) Automated software supported versus manual aorto-iliac diameter measurements in CT angiography of patients with abdominal aortic aneurysms: assessment of inter- and intraobserver variation. Vasa 34(4):255–261
Baumueller S, Nguyen TD, Goetti RP, Lachat M, Seifert B, Pfammatter T et al (2011) Maximum diameter measurements of aortic aneurysms on axial CT images after endovascular aneurysm repair: sufficient for follow-up? Cardiovasc Intervent Radiol 34(6):1182–1189
Hoornweg LL, Wisselink W, Vahl AC, Reekers JA, van Delden OM, Legemate DA et al (2008) Interobserver and intraobserver variability of interpretation of CT angiography in patients with a suspected abdominal aortic aneurysm rupture. Eur J Vasc Endovasc Surg 35(3):295–300
Rayt H, Lambert K, Bown M, Fishwick G, Morgan R, McCarthy M et al (2008) Can surgeons assess CT suitability for endovascular repair (EVAR) in ruptured abdominal aortic aneurysm? Implications for a ruptured EVAR trial. Cardiovasc Intervent Radiol 31(5):865–869
Ghatwary TM, Patterson BO, Karthikesalingam A, Hinchliffe RJ, Loftus IM, Morgan R et al (2013) A systematic review of protocols for the three-dimensional morphologic assessment of abdominal aortic aneurysms using computed tomographic angiography. Cardiovasc Intervent Radiol 36(1):14–24
Acknowledgments
This study was performed in accordance with the Declaration of Helsinki. Funding was provided by the Academic Medical Centre of Amsterdam Foundation. They did not assist in the study design, the interpretation of the data, or the writing of the manuscript. The 3mensio software was provided to H. Marquering without cost for evaluation purposes.
Conflicts of interest
There were no conflicts of interest to be reported.
Author information
Authors and Affiliations
Corresponding author
Appendix
Appendix
Scanning protocol suspected rAAA.
Centre A
Sliding gantry Siemens Somatom Sensation 64, Erlangen, Germany. Slice thickness 3 mm, collimation 0.6 mm, reconstruction increment 0.75 mm, pitch 1.2. Contrast protocol: power injection 100 ml Ultravist 300 (Bayer Healthcare Pharmaceuticals, Germany), 5 ml/s, bolus tracking, 50 ml NaCl saline chase 5 ml/s.
Centre B
Siemens Somatom Sensation 64, Erlangen, Germany slice thickness 1.5 mm, collimation 0.6 mm, reconstruction increment 1 mm, pitch 1.2. Contrast protocol: power injection 100 ml Ultravist (Bayer Healthcare Pharmaceuticals, Germany) 300, 5 ml/s, bolus tracking.
Rights and permissions
About this article
Cite this article
Reimerink, J.J., Marquering, H.A., Vahl, A. et al. Semiautomatic Sizing Software in Emergency Endovascular Aneurysm Repair for Ruptured Abdominal Aortic Aneurysms. Cardiovasc Intervent Radiol 37, 623–630 (2014). https://doi.org/10.1007/s00270-013-0757-1