Abstract
Purpose
Hip arthroplasty with metal-on-metal bearings like hip resurfacing results in the release of metallic ions. In parallel, like every metallic implant, knee arthroplasty implants undergo passive corrosion. We analyzed blood levels of cobalt and chromium ions in patients who have a hip resurfacing arthroplasty and compared them to patients who have undergone knee arthroplasty at a minimum follow-up of one year. The hypothesis was that there is no difference in the ion release between hip resurfacing and knee arthroplasty.
Methods
Sixty-three patients who underwent knee arthroplasty were compared to a cohort of 132 patients who underwent hip resurfacing. The blood levels of cobalt and chromium ions were determined preoperatively and at six and 12 months postoperatively and then compared between groups. We analyzed the relationship between ion release and the change in clinical outcome scores (Harris Hip score, Oxford Hip score, Merle D’Aubigné Postel score, Oxford Knee score, International Knee Society score), the BMI, sex, physical activity, implant size and inclination of the acetabular implant (hip resurfacing patients only). Mixed linear models were used to assess the changes in ion blood levels over time.
Results
The cobalt blood levels were higher in the first 6 months in the resurfacing group (0.87 ug/L vs 0.67 ug/L; p = 0.011), while it was higher in the knee arthroplasty group at 12 months (1.20 ug/L vs 1.41 ug/L; p = 0.0008). There were no significant differences in chromium levels during the follow-up period.
Conclusion
The increase in metal ion release after knee arthroplasty is as high as after hip resurfacing at the one year follow-up. The monitoring of this parameter probably should not be recommended in case of good clinicals outcomes.
Similar content being viewed by others
References
Brown SR, Davies WA, DeHeer DH, Swanson AB (2002) Long-term survival of McKee-Farrar total hip prostheses. Clin Orthop Rel Res 402:157–163. https://doi.org/10.1097/00003086-200209000-00013
Drummond J, Tran P, Fary C (2015) Metal-on-metal hip arthroplasty: a review of adverse reactions and patient management. J Funct Biomater 6(3):486–499. https://doi.org/10.3390/jfb6030486
Jantsch S, Schwägerl W, Zenz P, Semlitsch M, Fertschak W (1991) Long-term results after implantation of McKee-Farrar total hip prostheses. Arch Orthop Trauma Surg 110(5):230–237. https://doi.org/10.1007/BF00572878
European Commission, Directorate General for Health & Consumers. Opinion on the safety of metal-on-metal joint replacements with a particular focus on hip implants. Available at: http://dx.publications.europa.eu/10.2772/76284. Accessed February 2015.
Health C for D and R. Information for Orthopaedic Surgeons. FDA 2019. Available at; https://www.healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2012/15835a-eng.php. Accessed May 2012.
Guyen O, Tissot C (2016) Suivi des patients avec arthroplastie de hanche métal-métal et stratégie de prise en charge des complications. Rev Med Suisse 12(543):2156–2163
Lons A, Putman S, Pasquier G, Migaud H, Drumez E, Girard J (2017) Metallic ion release after knee prosthesis implantation: a prospective study. Int Orthop 41(12):2503–2508. https://doi.org/10.1007/s00264-017-3528-9
Lützner J, Günther K-P, Postler A, Morlock M (2020) Metal ion release after hip and knee arthroplasty - causes, biological effects and diagnostics. Z Orthop Unfall 158(4):369–382. https://doi.org/10.1055/a-0929-8121
Merle D’Aubigné R (1990) Numerical classification of the function of the hip. Rev Chir Orthop Reparatrice Appar Mot 76(6):371–374
Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51(4):737–755
Delaunay C, Epinette J-A, Dawson J, Murray D, Jolles B-M (2009) Cross-cultural adaptations of the Oxford-12 HIP score to the French speaking population. Orthop Traumatol Surg Res 95(2):89–99. https://doi.org/10.1016/j.otsr.2009.01.003
Jenny J-Y, Diesinger Y (2011) Validation of a French version of the Oxford knee questionnaire. Orthop Traumatol Surg Res 97(3):267–271. https://doi.org/10.1016/j.otsr.2010.07.009
Scuderi GR, Bourne RB, Noble PC, Benjamin JB, Lonner JH, Scott WN (2012) The new Knee Society Knee Scoring System. Clin Orthop Relat Res 470(1):3–19. https://doi.org/10.1007/s11999-011-2135-0
Tian J, Sun L, Hu R, Han W, Tian X (2018) Long-term results of primary hip arthroplasty with cup inclination angle bigger than fifty degrees. J Clin Orthop Trauma 9(2):133–136. https://doi.org/10.1016/j.jcot.2017.03.007
Vassiliou K, Elfick APD, Scholes SC, Unsworth A (2006) The effect of ‘running-in’ on the tribology and surface morphology of metal-on-metal Birmingham hip resurfacing device in simulator studies. Proc Inst Mech Eng H 220(2):269–277. https://doi.org/10.1243/09544119JEIM63
Heisel C, Streich N, Krachler M, Jakubowitz E, Kretzer JP (2008) Characterization of the running-in period in total hip resurfacing arthroplasty: an in vivo and in vitro metal ion analysis. J Bone Joint Surg 90(3):125–133. https://doi.org/10.2106/JBJS.H.00437
Elkins JM, Callaghan JJ, Brown TD (2014) Stability and trunnion wear potential in large-diameter metal-on-metal total hips: a finite element analysis. Clin Orthop Relat Res 472(2):529–542. https://doi.org/10.1007/s11999-013-3244-8
Kim Y-H, Park J-W, Kim J-S (2019) There is no significant difference in fretting and corrosion at the trunnion of metal and ceramic heads. Orthopedics 42(1):99–103. https://doi.org/10.3928/01477447-20181206-03
Ridon PE, Putman S, Migaud H, Berton C, Pasquier G, Girard J (2019) Long-term comparative study of large-diameter metal-on-metal bearings: resurfacing versus total arthroplasty with large-diameter Durom™ bearing. Orthop Traumatol Surg Res 105(5):943–948. https://doi.org/10.1016/j.otsr.2019.04.006
Sawalha S, Ravikumar R, Chowdhury EA, Massraf A (2022) The effect of obesity on blood metal ion levels after hip resurfacing and metal-on-metal total hip replacement. Hip Int 22(1):107–112. https://doi.org/10.5301/HIP.2012.9034
Ray GS, Laaksonen I, Galea VP, Madanat R, Muratoglu O, Malchau H (2020) Obesity is not associated with hip failure in patients with articular surface replacement of the hip. Hip Int 30(4):78–86. https://doi.org/10.1177/1120700019828705
De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br 90(10):1291–1297. https://doi.org/10.1302/0301-620X.90B10.20533
Le Duff MJ, Ebramzadeh E, Amstutz HC (2017) Contact patch to rim distance: the quintessential tool for metal-on-metal bearing in vivo performance analysis - a review. Hip Int 27(3):220–225. https://doi.org/10.5301/hipint.5000511
Yoon JP, Le Duff MJ, Johnson AJ, Takamura KM, Ebramzadeh E, Amstutz HC (2013) Contact patch to rim distance predicts metal ion levels in hip resurfacing. Clin Orthop Relat Res 471(5):1615–1621. https://doi.org/10.1007/s11999-012-2711-y
Girard J, Pansard E, Ouahes R, Migaud H, Delay C, Vasseur L (2014) Correlation between groin pain and cup design of hip-resurfacing implants: a prospective study. Int Orthop 38(5):923–928. https://doi.org/10.1007/s00264-013-2226-5
Reiner T, Sorbi R, Müller M, Nees T, Kretzer JP, Rickert M, Moradi B (2020) Blood metal ion release after primary total knee arthroplasty : a prospective study. Orthop Surg 12(2):396–403. https://doi.org/10.1111/os.12591
Tanoğlu O, Say F, Yücens M, Alemdaroğlu KB, İltar S, Aydoğan NH (2020) Titanium alloy intramedullary nails and plates affect serum metal ion levels within the fracture healing period. Biol Trace Elem Res 196(1):60–65. https://doi.org/10.1007/s12011-019-01913-1
Rylander LS, Milbrandt JC, Armington E, Wilson M, Olysav DJ (2010) Trace metal analysis following locked volar plating for unstable fractures of the distal radius. Iowa Orthop J 30:89–93
Zeh A, Becker C, Planert M, Lattke P, Wohlrab D (2009) Time-dependent release of cobalt and chromium ions into the serum following implantation of the metal-on-metal Maverick type artificial lumbar disc (Medtronic Sofamor Danek). Arch Orthop Trauma Surg 129(6):741–746. https://doi.org/10.1007/s00402-008-0677-8
Kim Y-J, Kassab F, Berven SH, Zurakowski D, Hresko MT, Emans JB, Kasser JR (2008) Serum levels of nickel and chromium after instrumented posterior spinal arthrodesis. Spine 30(8):923–926. https://doi.org/10.1097/01.brs.0000158872.42802.be
Peters RM, Willemse P, Rijk PC, Hoogendoorn M, Zijlstra WP (2017) Fatal cobalt toxicity after a non-metal-on-metal total hip arthroplasty. Case Rep Orthop:9123684. https://doi.org/10.1155/2017/9123684
Lecoanet P, Blangis M, Garcia M, Legallois Y, Fabre T (2019) Chromium-cobalt intoxication with intense systemic complications following total hip revision after per-operative ceramic fracture. Case Rep Orthop:4209796. https://doi.org/10.1155/2019/4209796
Agence nationale de sécurité du médicament et des produits de santé. Modalités de suivi des patients porteurs de prothèse totale de hanche à couple de frottement métal-métal. 2014. Available at: https://ansm.sante.fr/var/ansm_site/storage/original/application/0adb46da3096f9ce463611b830d170d1.pdf. Accessed December 2014.
Author information
Authors and Affiliations
Contributions
Simon Lukas: data acquisition, analysis and interpretation, and writing of the article. Pierre Martinot: writing of the article and critical review of the article for its intellectual content. Adrien Lons: critical review of the article for its intellectual content. Elodie Drumez: statistics. Julien Girard: critical review of the article for its intellectual content. Sophie Putman: critical review of the article for its intellectual content. Henri Migaud and Julien Girard: critical review of the article for its intellectual content and final approval of the version to be submitted.
Corresponding author
Ethics declarations
Ethics approval
According to French law, we did not provide an IRB because the carry out of a retrospective study involving a classic patient follow-up does not fall within the framework of an ethics committee.
Consent to participate
Informed consent was obtained from all individual participants included in the study.
Consent for publication
None.
Competing interests
The other authors declare no competing interests. Ortherwise, Henri Migaud is an education and research consultant for Zimmer, Corin, SERF, MSD and chief editor for Orthopaedics & Traumatology: Surgery & Research (Elsevier), Sophie Putman is an education and research consultant for Corin, Julien Girard is an education and research consultant for Microport, Smith & Nephew and Zimmer-Biomet.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Level of evidence: III (case-control study)
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Lukas, S., Martinot, P., Putman, S. et al. Metal ion release after hip resurfacing arthroplasty and knee arthroplasty: a retrospective study of one hundred ninety-five cases. International Orthopaedics (SICOT) 48, 119–126 (2024). https://doi.org/10.1007/s00264-023-05915-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-023-05915-6