Abstract
Objectives
Based on the features of the three-dimensional spatial structure of the proximal femoral trabeculae, we developed a bionic triangular supporting intramedullary nail (TSIN) for the treatment of the femoral intertrochanteric fracture. The current study aimed to compare the mechanical stability and restoration of mechanical conduction between proximal femoral nail antirotation (PFNA) and TSIN to fix the intertrochanteric fractures.
Methods
Firstly, five sets of PFNA and TSIN with the same size were selected and fixed on a biomechanical testing machine, and strain gauges were pasted on the main nail, lag screw, and supporting screw to load to the vertical load to 600 N, and the displacement and strain values were recorded. Secondly, formalin-preserved femurs were selected, and the left and right femurs of the same cadaver were randomly divided into two groups to prepare intertrochanteric femur fractures (AO classification 31–A1), which were fixed with PFNA (n = 15) and TSIN (n = 15), respectively. Sixteen sites around the fracture line were chosen to paste strain gauges and loaded vertically to 600 N, and then, the fracture fragment displacement and strain values were recorded. Finally, a 10,000-cycle test ranging from 10 to 600 N was conducted, and the cycle number and displacement value were recorded.
Results
The overall displacement of PFNA was 2.17 ± 0.18 mm, which was significantly greater than the displacement of the TSIN group (1.66 ± 0.05 mm, P < 0.05) under a vertical load of 600 N. The strain below the PFNA lag screw was 868.29 ± 147.85, which was significantly greater than that of the TSIN (456.02 ± 35.06, P < 0.05); the strain value at the medial side of the PFNA nail was 444.00 ± 34.23, which was significantly less than that of the TSIN (613.57 ± 108.00, P < 0.05). Under the vertical load of 600 N, the displacement of the fracture fragments of the PFNA group was 0.95 ± 0.25 mm, which was significantly greater than that of the TSIN group (0.41 ± 0.09 mm, P < 0.05). The femoral specimens in the PFNA group showed significantly greater strains at the anterior (1, 2, and 4), lateral (7, 9, and 10), posterior (11), and medial (15 and 16) sites than those in the TSIN group (all P < 0.05). In the cyclic compression experiment, the displacements of the PFNA group at 2000, 4000, 6000, 8000, and 10,000 cycles were 1.38 mm, 1.81 mm, 2.07 mm, 2.64 mm, and 3.58 mm, respectively, which were greater than the corresponding displacements of the TSIN group: 1.01 mm, 1.48 mm, 1.82 mm, 2.05 mm, and 2.66 mm (P8000 = 0.012, P10000 = 0.006).
Conclusions
The current study showed that TSIN had apparent advantages in stability and stress conduction. TSIN enhanced the stability of intertrochanteric fractures, particularly in superior fracture fragments, improved stress conduction, reduced the stress in the anterior and medial femur, and restored the biomechanical properties of the femur.
Similar content being viewed by others
Change history
19 January 2022
A Correction to this paper has been published: https://doi.org/10.1007/s00264-021-05297-7
References
Chen W, Lv H, Liu S, Liu B, Zhu Y, Chen X, Yang G, Liu L, Zhang T, Wang H, Yin B, Guo J, Zhang X, Li Y, Smith D, Hu P, Sun J, Zhang Y (2017) National incidence of traumatic fractures in China: a retrospective survey of 512 187 individuals. Lancet Glob Health 5:e807–e817. https://doi.org/10.1016/S2214-109X(17)30222-X
Cummings SR, Melton LJ (2002) Epidemiology and outcomes of osteoporotic fractures. Lancet 359:1761–1767. https://doi.org/10.1016/S0140-6736(02)08657-9
Bergstrm U, Jonsson H, Gustafson Y, Pettersson U, Svensson O (2009) The hip fracture incidence curve is shifting to the right. Acta Orthop 80(5):520–524. https://doi.org/10.3109/17453670903278282
Cooper C, Campion G, Melton LJ 3rd (1992) Hip fractures in the elderly: a world-wide projection. Osteoporos Int 2:285–289. https://doi.org/10.1007/BF01623184
Cheung CL, Ang SB, Chadha M, Chow ES, Chung YS, Hew FL, Jaisamrarn U, Ng H, Takeuchi Y, Wu CH, Xia W, Yu J, Fujiwara S (2018) An updated hip fracture projection in Asia: the Asian Federation of Osteoporosis Societies study. Osteoporos Sarcopenia 4:16–21. https://doi.org/10.1016/j.afos.2018.03.003
Kaufer H (1980) Mechanics of the treatment of hip injuries. Clin Orthop Relat Res 146:53–61
Wegrzyn J, Roux JP, Loriau C, Bonin N, Pibarot V (2018) The tridimensional geometry of the proximal femur should determine the design of cementless femoral stem in total hip arthroplasty. Int Orthop 42:2329–2334. https://doi.org/10.1007/s00264-018-3843-9
Mahomed N, Harrington I, Kellurn J, Maistrelli G, Hearn T, Vroemen J (1994) Biomechanical analysis of the Gamma nail and sliding hip screw. Clin Orthop Relat Res 304:280–288
Wang J, Ma XL, Ma JX, Xing D, Yang Y, Zhu SW, Ma BY, Chen Y, Feng R, Jia HB (2014) Biomechanical analysis of four types of internal fixation in subtrochanteric fracture models. Orthop Surg 6:128–136
Bartoníček J, Rammelt S (2014) The history of internal fixation of proximal femur fractures Ernst Pohl-the genius behind. Int Orthop 38:2421–2426
Kasha S, Yalamanchili RK (2019) Management of subtrochanteric fractures by nail osteosynthesis: a review of tips and tricks. Int Orthop 44:645–653. https://doi.org/10.1007/s00264-019-04404-z
Siavashi B, Aalirezaei A, Moosavi M, Golbakhsh MR, Savadkoohi D, Zehtab MJ (2015) A comparative study between multiple cannulated screws and dynamic hip screw for fixation of femoral neck fracture in adults. Int Orthop 39:2069–2071
Glassner PJ, Tejwani NC (2011) Failure of proximal femoral locking compression plate: a case series. J Orthop Trauma 25:76–83. https://doi.org/10.1097/BOT.0b013e3181e31ccc
Jin WJ, Dai LY, Cui YM, Zhou Q, Jiang LS, Lu H (2005) Reliability of classification systems for intertrochanteric fractures of the proximal femur in experienced orthopaedic surgeons. Injury 36:858–861. https://doi.org/10.1016/j.injury.2005.02.005
Horner NS, Samuelsson K, Solyom J, Bjorgul K, Ayeni OR, Ostman B (2017) Implant-related complications and mortality after use of short or long gamma nail for intertrochanteric and subtrochanteric fractures: a prospective study with minimum 13-year follow-up. JB JS Open Access 2:e0026. https://doi.org/10.2106/JBJS.OA.17.00026
Lakho MT, Jatoi AA, Azfar MK, Ali A, Javed S, Bhatti A, Karim M (2019) Functional and radiological outcome of unstable intertrochanteric fracture post dynamic hip screw fixation. Cureus 11:e4360. https://doi.org/10.7759/cureus.4360
Yaozeng X, Dechun G, Huilin Y, Guangming Z, Xianbin W (2010) Comparative study of trochanteric fracture treated with the proximal femoral nail anti-rotation and the third generation of gamma nail. Injury 41:1234–1238. https://doi.org/10.1016/j.injury.2010.03.005
Liu Y, Tao R, Liu F, Wang Y, Zhou Z, Cao Y, Wang H (2010) Mid-term outcomes after intramedullary fixation of peritrochanteric femoral fractures using the new proximal femoral nail antirotation (PFNA). Injury 41:810–817. https://doi.org/10.1016/j.injury.2010.03.020
Fensky F, Nuchtern JV, Kolb JP, Huber S, Rupprecht M, Jauch SY, Sellenschloh K, Puschel K, Morlock MM, Rueger JM, Lehmann W (2013) Cement augmentation of the proximal femoral nail antirotation for the treatment of osteoporotic pertrochanteric fractures–a biomechanical cadaver study. Injury 44:802–807. https://doi.org/10.1016/j.injury.2013.03.003
Kwak DK, Kim WH, Lee SJ, Rhyu SH, Jang CY, Yoo JH (2018) Biomechanical comparison of three different intramedullary nails for fixation of unstable basicervical intertrochanteric fractures of the proximal femur: experimental studies. Biomed Res Int 2018:7618079. https://doi.org/10.1155/2018/7618079
Aminian A, Gao F, Fedoriw WW, Zhang LQ, Kalainov DM, Merk BR (2007) Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques. J Orthop Trauma 21:544–548. https://doi.org/10.1097/BOT.0b013e31814b822e
Hollensteiner M, Sandriesser S, Bliven E, von Ruden C, Augat P (2019) Biomechanics of osteoporotic fracture fixation. Curr Osteoporos Rep 17:363–374. https://doi.org/10.1007/s11914-019-00535-9
Wu X, Wang Z, Li H, Li Y, Wang H, Tian W (2019) Biomechanical evaluation of osteoporotic fracture: metal fixation versus absorbable fixation in Sawbones models. Injury 50:1272–1276. https://doi.org/10.1016/j.injury.2019.05.023
Nie S, Li M, Ji H, Li Z, Li W, Zhang H, Licheng Z, Tang P (2020) Biomechanical comparison of medial sustainable nail and proximal femoral nail antirotation in the treatment of an unstable intertrochanteric fracture. Bone Joint Res 9:840–847. https://doi.org/10.1302/2046-3758.912.BJR-2020-0284.R1
Frost HM (1987) Bone “mass” and the “mechanostat”: a proposal. Anat Rec 219:1–9. https://doi.org/10.1002/ar.1092190104
Kouvidis GK, Sommers MB, Giannoudis PV, Katonis PG, Bottlang M (2009) Comparison of migration behavior between single and dual lag screw implants for intertrochanteric fracture fixation. J Orthop Surg Res 4:16. https://doi.org/10.1186/1749-799X-4-16
Li M, Zhao K, Ding K, Cui YW, Cheng XD, Yang WJ, Hou ZY, Zhang YZ, Chen W, Hu P, Zhu YB (2021) Titanium alloy gamma nail versus biodegradable magnesium alloy bionic gamma nail for treating intertrochanteric fractures: a finite element analysis. Orthop Surg 13:1513–1520. https://doi.org/10.1111/os.12973
Sverdlova N (2011) Tensile trabeculae–myth or reality? J Musculoskelet Neuronal Interact 11:1–7
Cun Y, Dou C, Tian S, Li M, Zhu Y, Cheng X, Chen W (2020) Traditional and bionic dynamic hip screw fixation for the treatment of intertrochanteric fracture: a finite element analysis. Int Orthop 44:551–559. https://doi.org/10.1007/s00264-019-04478-9
Nawathe S, Nguyen BP, Barzanian N, Akhlaghpour H, Bouxsein ML, Keaveny TM (2015) Cortical and trabecular load sharing in the human femoral neck. J Biomech 48:816–822
Ding K, Yang W, Zhu J, Cheng X, Wang H, Hao D, Yinuo S, Zhu Y, Zhang Y, Chen W, Zhang Q (2021) Titanium alloy cannulated screws and biodegradable magnesium alloy bionic cannulated screws for treatment of femoral neck fractures: a finite element analysis. J Orthop Surg Res 16:511. https://doi.org/10.1186/s13018-021-02665-2
Hsu CE, Shih CM, Wang CC, Huang KC (2013) Lateral femoral wall thickness. A reliable predictor of post-operative lateral wall fracture in intertrochanteric fractures. Bone Joint J 95-B:1134–1138. https://doi.org/10.1302/0301-620X.95B8.31495
Sun LL, Li Q, Chang SM (2016) The thickness of proximal lateral femoral wall. Injury 47:784–785. https://doi.org/10.1016/j.injury.2016.01.002
Ehrnthaller C, Olivier AC, Gebhard F, Durselen L (2017) The role of lesser trochanter fragment in unstable pertrochanteric A2 proximal femur fractures - is refixation of the lesser trochanter worth the effort? Clin Biomech (Bristol, Avon) 42:31–37. https://doi.org/10.1016/j.clinbiomech.2016.12.013
Funding
National Natural Science Foundation of China (82072447); Hebei Province fund for Distinguished Young Scholars (H2021206329).
This study conforms to the provisions of the Declaration of Helsinki and has been reviewed and approved by the Institutional Review Board of The Third Hospital of Hebei Medical University.
Author information
Authors and Affiliations
Contributions
Y Z, W C and Q Z conceived the idea and designed the study. H W, W Y, and K D collected the relevant data. Y Z and K Z prepared the figures and tables. Y Z and C R performed the statistical analyses. All the authors interpreted the data and contributed to preparation of the manuscript. H W, W Y, and K D wrote the manuscript.
Corresponding authors
Ethics declarations
Ethics approval
This article does not contain any studies with human participants or animals performed by any of the authors.
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
The original version of this article was revised. The correct copyright should be "The Author(s) under exclusive licence to SICOT aisbl."
Rights and permissions
About this article
Cite this article
Wang, H., Yang, W., Ding, K. et al. Biomechanical study on the stability and strain conduction of intertrochanteric fracture fixed with proximal femoral nail antirotation versus triangular supporting intramedullary nail. International Orthopaedics (SICOT) 46, 341–350 (2022). https://doi.org/10.1007/s00264-021-05250-8
Received:
Accepted:
Published:
Issue Date: