Customized implants for acetabular Paprosky III defects may be positioned with high accuracy in revision hip arthroplasty

  • Original Paper
  • Published:
International Orthopaedics Aims and scope Submit manuscript

Abstract

Purpose

In revision hip arthroplasty, custom-made implants are one option in patients with acetabular Paprosky III defects.

Methods

In a retrospective analysis, we identified 11 patients undergoing cup revision using a custom-made implant. The accuracy of the intended position of the implant was assessed on post-operative 3D CT and compared to the pre-operative 3D planning in terms of inclination, anteversion, and centre of rotation. In addition, the accuracy of post-operative plain radiographs for measuring implant position was evaluated in relation to the 3D CT standard.

Results

We found a mean deviation between the planned and the final position of the custom-made acetabular implant on 3D CT of 3.6° ± 2.8° for inclination and of − 1.2° ± 7.0° for anteversion, respectively. Restoration of center of rotation succeeded with an accuracy of 0.3 mm ± 3.9 mm in the mediolateral (x) direction, − 1.1 mm ± 3.8 mm in the anteroposterior (y) direction, and 0.4 mm ± 3.2 mm in the craniocaudal (z) direction. The accuracy of the post-operative plain radiographs in measuring the position of the custom-made implant in relation to 3D CT was 1.1° ± 1.7° for implant inclination, − 2.6° ± 1.3° for anteversion and 1.3 mm ± 3.5 mm in the x-direction, and − 0.9 mm ± 3.8 mm in the z-direction for centre of rotation.

Conclusion

Custom-made acetabular implants can be positioned with good accuracy in Paprosky III defects according to the pre-operative planning. Plain radiographs are adequate for assessing implant position in routine follow-up.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Learmonth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet 370:1508–1519. https://doi.org/10.1016/S0140-6736(07)60457-7

    Article  PubMed  Google Scholar 

  2. Hawker G, Wright J, Coyte P, Paul J, Dittus R, Croxford R, Katz B, Bombardier C, Heck D, Freund D (1998) Health-related quality of life after knee replacement. J Bone Joint Surg Am 80:163–173

    Article  CAS  Google Scholar 

  3. Ong KL, Mowat FS, Chan N, Lau E, Halpern MT, Kurtz SM (2006) Economic burden of revision hip and knee arthroplasty in Medicare enrollees. Clin Orthop Relat Res 446:22–28. https://doi.org/10.1097/01.blo.0000214439.95268.59

    Article  CAS  PubMed  Google Scholar 

  4. Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89:780–785. https://doi.org/10.2106/JBJS.F.00222

    Article  Google Scholar 

  5. Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplast 9:33–44

    Article  CAS  Google Scholar 

  6. Yu R, Hofstaetter JG, Sullivan T, Costi K, Howie DW, Solomon LB (2013) Validity and reliability of the Paprosky acetabular defect classification. Clin Orthop Relat Res 471:2259–2265. https://doi.org/10.1007/s11999-013-2844-7

    Article  PubMed  PubMed Central  Google Scholar 

  7. Baauw M, van Hooff ML, Spruit M (2016) Current construct options for revision of large acetabular defects: a systematic review. JBJS Rev 4. https://doi.org/10.2106/JBJS.RVW.15.00119

    Article  Google Scholar 

  8. Baauw M, van Hellemondt GG, van Hooff ML, Spruit M (2015) The accuracy of positioning of a custom-made implant within a large acetabular defect at revision arthroplasty of the hip. Bone Joint J 97-B:780–785. https://doi.org/10.1302/0301-620X.97B6.35129

    Article  CAS  PubMed  Google Scholar 

  9. Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components: a comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br 93:629–633. https://doi.org/10.1302/0301-620X.93B5.25893

    Article  CAS  PubMed  Google Scholar 

  10. Kalteis T, Handel M, Herold T, Perlick L, Paetzel C, Grifka J (2006) Position of the acetabular cup -- accuracy of radiographic calculation compared to CT-based measurement. Eur J Radiol 58:294–300. https://doi.org/10.1016/j.ejrad.2005.10.003

    Article  PubMed  Google Scholar 

  11. Bayraktar V, Weber M, von Kunow F, Zeman F, Craiovan B, Renkawitz T, Grifka J, Woerner M (2017) Accuracy of measuring acetabular cup position after total hip arthroplasty: comparison between a radiographic planning software and three-dimensional computed tomography. Int Orthop 41:731–738. https://doi.org/10.1007/s00264-016-3240-1

    Article  PubMed  Google Scholar 

  12. Choi HR, Anderson D, Foster S, Beal M, Lee JA, Barr C, Malchau H, McCarthy J, Kwon YM (2013) Acetabular cup positioning in revision total hip arthroplasty with Paprosky type III acetabular defects: Martell radiographic analysis. Int Orthop 37:1905–1910. https://doi.org/10.1007/s00264-013-2008-0

    Article  PubMed  PubMed Central  Google Scholar 

  13. Arauz P, Peng Y, MacAuliffe J, Kwon YM (2018) In-vivo 3-dimensional gait symmetry analysis in patients with bilateral total hip arthroplasty. J Biomech 77:131–137. https://doi.org/10.1016/j.jbiomech.2018.07.013

    Article  PubMed  Google Scholar 

  14. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60:217–220

    Article  CAS  Google Scholar 

  15. Audige L, Bhandari M, Kellam J (2004) How reliable are reliability studies of fracture classifications? A systematic review of their methodologies. Acta Orthop Scand 75:184–194. https://doi.org/10.1080/00016470412331294445

    Article  PubMed  Google Scholar 

  16. Woerner M, Sendtner E, Springorum R, Craiovan B, Worlicek M, Renkawitz T, Grifka J, Weber M (2016) Visual intraoperative estimation of cup and stem position is not reliable in minimally invasive hip arthroplasty. Acta Orthop:1–6. https://doi.org/10.3109/17453674.2015.1137182

    Article  Google Scholar 

  17. Barrack RL, Krempec JA, Clohisy JC, McDonald DJ, Ricci WM, Ruh EL, Nunley RM (2013) Accuracy of acetabular component position in hip arthroplasty. J Bone Joint Surg Am 95:1760–1768. https://doi.org/10.2106/JBJS.L.01704

    Article  PubMed  Google Scholar 

  18. Eilander W, Harris SJ, Henkus HE, Cobb JP, Hogervorst T (2013) Functional acetabular component position with supine total hip replacement. Bone Joint J 95-B:1326–1331. https://doi.org/10.1302/0301-620X.95B10.31446

    Article  CAS  PubMed  Google Scholar 

  19. Weber M, Craiovan B, Woerner ML, Schwarz T, Grifka J, Renkawitz TF (2017) Predictors of outcome after primary total joint replacement. J Arthroplast. https://doi.org/10.1016/j.arth.2017.08.044

    Article  Google Scholar 

  20. Weber M, Woerner M, Messmer B, Grifka J, Renkawitz T (2017) Navigation is equal to estimation by eye and palpation in preventing psoas impingement in THA. Clin Orthop Relat Res 475:196–203. https://doi.org/10.1007/s11999-016-5061-3

    Article  PubMed  Google Scholar 

  21. Weber M, Weber T, Woerner M, Craiovan B, Worlicek M, Winkler S, Grifka J, Renkawitz T (2015) The impact of standard combined anteversion definitions on gait and clinical outcome within one year after total hip arthroplasty. Int Orthop 39:2323–2333. https://doi.org/10.1007/s00264-015-2777-8

    Article  PubMed  Google Scholar 

  22. Mao Y, Xu C, Xu J, Li H, Liu F, Yu D, Zhu Z (2015) The use of customized cages in revision total hip arthroplasty for Paprosky type III acetabular bone defects. Int Orthop 39:2023–2030. https://doi.org/10.1007/s00264-015-2965-6

    Article  PubMed  Google Scholar 

  23. Flecher X, Appy B, Parratte S, Ollivier M, Argenson JN (2017) Use of porous tantalum components in Paprosky two and three acetabular revision. A minimum five-year follow-up of fifty one hips. Int Orthop 41:911–916. https://doi.org/10.1007/s00264-016-3312-2

    Article  PubMed  Google Scholar 

  24. Volpin A, Konan S, Biz C, Tansey RJ, Haddad FS (2018) Reconstruction of failed acetabular component in the presence of severe acetabular bone loss: a systematic review. Musculoskelet Surg. https://doi.org/10.1007/s12306-018-0539-7

    Article  Google Scholar 

  25. Weber M, Benditz A, Woerner M, Weber D, Grifka J, Renkawitz T (2017) Trainee surgeons affect operative time but not outcome in minimally invasive total hip arthroplasty. Sci Rep 7:6152. https://doi.org/10.1038/s41598-017-06530-3

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The support of Mr. S. Reuter during analysis of 3D-CT datasets is highly appreciated.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Markus Weber.

Ethics declarations

The investigation was approved by the local medical ethics committee (No.: 17-415-101).

Conflict of interest

TR has received research support by DePuy International, Otto Bock Foundation, Deutsche Arthose Hilfe. TR’s research group “patientindividual joint replacement” is supported by the German Ministry of Education and Research (BMBF, grant number 01EZ0915). JG got research support by MSD, Novartis, De Puy, Otto Bock Foundation. Further financial support is from De Puy, Orthotech, Ozo-zours, Fischer Fussfit, Urban & Kemmler. JW is an employee for AQ implants. All other authors declare no potential conflict of interest.

Additional information

This work was performed at Regensburg University Medical Center, Department of Orthopaedic Surgery, Bad Abbach, Germany.

Electronic supplementary material

ESM 1

(DOCX 20 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Weber, M., Witzmann, L., Wieding, J. et al. Customized implants for acetabular Paprosky III defects may be positioned with high accuracy in revision hip arthroplasty. International Orthopaedics (SICOT) 43, 2235–2243 (2019). https://doi.org/10.1007/s00264-018-4193-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00264-018-4193-3

Keywords

Navigation