Abstract
Purpose
In revision hip arthroplasty, custom-made implants are one option in patients with acetabular Paprosky III defects.
Methods
In a retrospective analysis, we identified 11 patients undergoing cup revision using a custom-made implant. The accuracy of the intended position of the implant was assessed on post-operative 3D CT and compared to the pre-operative 3D planning in terms of inclination, anteversion, and centre of rotation. In addition, the accuracy of post-operative plain radiographs for measuring implant position was evaluated in relation to the 3D CT standard.
Results
We found a mean deviation between the planned and the final position of the custom-made acetabular implant on 3D CT of 3.6° ± 2.8° for inclination and of − 1.2° ± 7.0° for anteversion, respectively. Restoration of center of rotation succeeded with an accuracy of 0.3 mm ± 3.9 mm in the mediolateral (x) direction, − 1.1 mm ± 3.8 mm in the anteroposterior (y) direction, and 0.4 mm ± 3.2 mm in the craniocaudal (z) direction. The accuracy of the post-operative plain radiographs in measuring the position of the custom-made implant in relation to 3D CT was 1.1° ± 1.7° for implant inclination, − 2.6° ± 1.3° for anteversion and 1.3 mm ± 3.5 mm in the x-direction, and − 0.9 mm ± 3.8 mm in the z-direction for centre of rotation.
Conclusion
Custom-made acetabular implants can be positioned with good accuracy in Paprosky III defects according to the pre-operative planning. Plain radiographs are adequate for assessing implant position in routine follow-up.
Similar content being viewed by others
References
Learmonth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet 370:1508–1519. https://doi.org/10.1016/S0140-6736(07)60457-7
Hawker G, Wright J, Coyte P, Paul J, Dittus R, Croxford R, Katz B, Bombardier C, Heck D, Freund D (1998) Health-related quality of life after knee replacement. J Bone Joint Surg Am 80:163–173
Ong KL, Mowat FS, Chan N, Lau E, Halpern MT, Kurtz SM (2006) Economic burden of revision hip and knee arthroplasty in Medicare enrollees. Clin Orthop Relat Res 446:22–28. https://doi.org/10.1097/01.blo.0000214439.95268.59
Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89:780–785. https://doi.org/10.2106/JBJS.F.00222
Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplast 9:33–44
Yu R, Hofstaetter JG, Sullivan T, Costi K, Howie DW, Solomon LB (2013) Validity and reliability of the Paprosky acetabular defect classification. Clin Orthop Relat Res 471:2259–2265. https://doi.org/10.1007/s11999-013-2844-7
Baauw M, van Hooff ML, Spruit M (2016) Current construct options for revision of large acetabular defects: a systematic review. JBJS Rev 4. https://doi.org/10.2106/JBJS.RVW.15.00119
Baauw M, van Hellemondt GG, van Hooff ML, Spruit M (2015) The accuracy of positioning of a custom-made implant within a large acetabular defect at revision arthroplasty of the hip. Bone Joint J 97-B:780–785. https://doi.org/10.1302/0301-620X.97B6.35129
Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components: a comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br 93:629–633. https://doi.org/10.1302/0301-620X.93B5.25893
Kalteis T, Handel M, Herold T, Perlick L, Paetzel C, Grifka J (2006) Position of the acetabular cup -- accuracy of radiographic calculation compared to CT-based measurement. Eur J Radiol 58:294–300. https://doi.org/10.1016/j.ejrad.2005.10.003
Bayraktar V, Weber M, von Kunow F, Zeman F, Craiovan B, Renkawitz T, Grifka J, Woerner M (2017) Accuracy of measuring acetabular cup position after total hip arthroplasty: comparison between a radiographic planning software and three-dimensional computed tomography. Int Orthop 41:731–738. https://doi.org/10.1007/s00264-016-3240-1
Choi HR, Anderson D, Foster S, Beal M, Lee JA, Barr C, Malchau H, McCarthy J, Kwon YM (2013) Acetabular cup positioning in revision total hip arthroplasty with Paprosky type III acetabular defects: Martell radiographic analysis. Int Orthop 37:1905–1910. https://doi.org/10.1007/s00264-013-2008-0
Arauz P, Peng Y, MacAuliffe J, Kwon YM (2018) In-vivo 3-dimensional gait symmetry analysis in patients with bilateral total hip arthroplasty. J Biomech 77:131–137. https://doi.org/10.1016/j.jbiomech.2018.07.013
Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60:217–220
Audige L, Bhandari M, Kellam J (2004) How reliable are reliability studies of fracture classifications? A systematic review of their methodologies. Acta Orthop Scand 75:184–194. https://doi.org/10.1080/00016470412331294445
Woerner M, Sendtner E, Springorum R, Craiovan B, Worlicek M, Renkawitz T, Grifka J, Weber M (2016) Visual intraoperative estimation of cup and stem position is not reliable in minimally invasive hip arthroplasty. Acta Orthop:1–6. https://doi.org/10.3109/17453674.2015.1137182
Barrack RL, Krempec JA, Clohisy JC, McDonald DJ, Ricci WM, Ruh EL, Nunley RM (2013) Accuracy of acetabular component position in hip arthroplasty. J Bone Joint Surg Am 95:1760–1768. https://doi.org/10.2106/JBJS.L.01704
Eilander W, Harris SJ, Henkus HE, Cobb JP, Hogervorst T (2013) Functional acetabular component position with supine total hip replacement. Bone Joint J 95-B:1326–1331. https://doi.org/10.1302/0301-620X.95B10.31446
Weber M, Craiovan B, Woerner ML, Schwarz T, Grifka J, Renkawitz TF (2017) Predictors of outcome after primary total joint replacement. J Arthroplast. https://doi.org/10.1016/j.arth.2017.08.044
Weber M, Woerner M, Messmer B, Grifka J, Renkawitz T (2017) Navigation is equal to estimation by eye and palpation in preventing psoas impingement in THA. Clin Orthop Relat Res 475:196–203. https://doi.org/10.1007/s11999-016-5061-3
Weber M, Weber T, Woerner M, Craiovan B, Worlicek M, Winkler S, Grifka J, Renkawitz T (2015) The impact of standard combined anteversion definitions on gait and clinical outcome within one year after total hip arthroplasty. Int Orthop 39:2323–2333. https://doi.org/10.1007/s00264-015-2777-8
Mao Y, Xu C, Xu J, Li H, Liu F, Yu D, Zhu Z (2015) The use of customized cages in revision total hip arthroplasty for Paprosky type III acetabular bone defects. Int Orthop 39:2023–2030. https://doi.org/10.1007/s00264-015-2965-6
Flecher X, Appy B, Parratte S, Ollivier M, Argenson JN (2017) Use of porous tantalum components in Paprosky two and three acetabular revision. A minimum five-year follow-up of fifty one hips. Int Orthop 41:911–916. https://doi.org/10.1007/s00264-016-3312-2
Volpin A, Konan S, Biz C, Tansey RJ, Haddad FS (2018) Reconstruction of failed acetabular component in the presence of severe acetabular bone loss: a systematic review. Musculoskelet Surg. https://doi.org/10.1007/s12306-018-0539-7
Weber M, Benditz A, Woerner M, Weber D, Grifka J, Renkawitz T (2017) Trainee surgeons affect operative time but not outcome in minimally invasive total hip arthroplasty. Sci Rep 7:6152. https://doi.org/10.1038/s41598-017-06530-3
Acknowledgements
The support of Mr. S. Reuter during analysis of 3D-CT datasets is highly appreciated.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
The investigation was approved by the local medical ethics committee (No.: 17-415-101).
Conflict of interest
TR has received research support by DePuy International, Otto Bock Foundation, Deutsche Arthose Hilfe. TR’s research group “patientindividual joint replacement” is supported by the German Ministry of Education and Research (BMBF, grant number 01EZ0915). JG got research support by MSD, Novartis, De Puy, Otto Bock Foundation. Further financial support is from De Puy, Orthotech, Ozo-zours, Fischer Fussfit, Urban & Kemmler. JW is an employee for AQ implants. All other authors declare no potential conflict of interest.
Additional information
This work was performed at Regensburg University Medical Center, Department of Orthopaedic Surgery, Bad Abbach, Germany.
Electronic supplementary material
ESM 1
(DOCX 20 kb)
Rights and permissions
About this article
Cite this article
Weber, M., Witzmann, L., Wieding, J. et al. Customized implants for acetabular Paprosky III defects may be positioned with high accuracy in revision hip arthroplasty. International Orthopaedics (SICOT) 43, 2235–2243 (2019). https://doi.org/10.1007/s00264-018-4193-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-018-4193-3