Abstract
Introduction
It remains unclear if the increased biomechanical stability of the double-row repair is solely a result of the technique itself or due to the greater number of utilized materials. This study analyses the various rotator cuff (RC) repair interfaces and the influence of doubling the number of suture knots, stitch techniques and suture anchors on the primary stability of the RC reconstruction.
Materials and methods
Established cyclic testing protocols were used to test the primary stability of the knot interface with a Duncan loop and Roeder knot, the tendon-to-suture interface with a single-stitch and Mason-Allen-stitch and the anchor-to-bone interface with OPUS Magnum 2® and BioCorkscrew® anchors. After doubling the materials, the interfaces were tested again and compared to the yielded solitary values. The maximum failure loads (Fmax), clinical failure (CF), system displacement (SD) and modes of failure were recorded.
Results
Doubling the amount of knots, stitches and suture anchors improved almost all interface parameters. For example, Fmax, CF and SD of the Roeder knot could be significantly improved from 142.5 N to 256.7 N, from 82.5 N to 155.0 N and from 0.45 mm to 0.20 mm, respectively. The Mason-Allen stitch’s Fmax improved from 196.4 N to 324.9 N, and CF from 94.25 N to 139.8 N, with almost identical SD of 0.81 mm and 0.84 mm. Bio-Corkscrew® Fmax improved from 232.5 N to 317.5 N, CF from 182.5 N to 210 N, and SD from 0.85 mm to 0.64 mm.
Conclusion
Doubling the number of knots, stitches and suture anchors increases the primary stability of the various RC repair interfaces and may result in RC rerupture rate reduction. Furthermore, this study suggests that a suture anchor loaded with two or three sutures may yield the same primary stability as two or three suture anchors loaded with a single suture. This in vitro biomechanical study focuses solely on the rotator cuff interfaces at time zero; the biological healing process was therefore not analysed and requires further investigation.
Clinical relevance
Improved primary stability by increased number of suture materials may reduce rate of RC rerupture.
Similar content being viewed by others
References
Tashjian RZ, Henn RF, Kang L, Green A (2004) The effect of comorbidity on self-assessed function in patients with a chronic rotator cuff tear. J Bone Joint Surg Am 86-A(2):355–362
Teunis T, Lubberts B, Reilly BT, Ring D (2014) A systematic review and pooled analysis of the prevalence of rotator cuff disease with increasing age. J Shoulder Elb Surg 23(12):1913–1921. doi:10.1016/j.jse.2014.08.001
Cummins CA, Murrell GA (2003) Mode of failure for rotator cuff repair with suture anchors identified at revision surgery. J Shoulder Elb Surg 12(2):128–133
Boileau P, Brassart N, Watkinson DJ, Carles M, Hatzidakis AM, Krishnan SG (2005) Arthroscopic repair of full-thickness tears of the supraspinatus: does the tendon really heal? J Bone Joint Surg Am 87(6):1229–1240. doi:10.2106/JBJS.D.02035
Frank JB, ElAttrache NS, Dines JS, Blackburn A, Crues J, Tibone JE (2008) Repair site integrity after arthroscopic transosseous-equivalent suture-bridge rotator cuff repair. Am J Sports Med 36(8):1496–1503. doi:10.1177/0363546507313574
Kamath G, Galatz LM, Keener JD, Teefey S, Middleton W, Yamaguchi K (2009) Tendon integrity and functional outcome after arthroscopic repair of high-grade partial-thickness supraspinatus tears. J Bone Joint Surg Am 91(5):1055–1062. doi:10.2106/JBJS.G.00118
Sugaya H, Maeda K, Matsuki K, Moriishi J (2007) Repair integrity and functional outcome after arthroscopic double-row rotator cuff repair. A prospective outcome study. J Bone Joint Surg Am 89(5):953–960. doi:10.2106/JBJS.F.00512
Zumstein MA, Jost B, Hempel J, Hodler J, Gerber C (2008) The clinical and structural long-term results of open repair of massive tears of the rotator cuff. J Bone Joint Surg Am 90(11):2423–2431. doi:10.2106/JBJS.G.00677
Meier SW, Meier JD (2006) Rotator cuff repair: the effect of double-row fixation on three-dimensional repair site. J Shoulder Elb Surg 15(6):691–696
Brady PC, Arrigoni P, Burkhart SS (2006) Evaluation of residual rotator cuff defects after in vivo single- versus double-row rotator cuff repairs. Arthroscopy 22(10):1070–1075
Kim DH, Elattrache NS, Tibone JE, Jun BJ, DeLaMora SN, Kvitne RS, Lee TQ (2006) Biomechanical comparison of a single-row versus double-row suture anchor technique for rotator cuff repair. Am J Sports Med 34(3):407–414
Wall LB, Keener JD, Brophy RH (2009) Double-row vs single-row rotator cuff repair: a review of the biomechanical evidence. J Shoulder Elb Surg 18(6):933–941
Carbonel I, Martinez AA, Calvo A, Ripalda J, Herrera A (2012) Single-row versus double-row arthroscopic repair in the treatment of rotator cuff tears: a prospective randomized clinical study. Int Orthop 36(9):1877–1883. doi:10.1007/s00264-012-1559-9
Chen M, Xu W, Dong Q, Huang Q, Xie Z, Mao Y (2013) Outcomes of single-row versus double-row arthroscopic rotator cuff repair: a systematic review and meta-analysis of current evidence. Arthroscopy 29(8):1437–1449
Lapner PL, Sabri E, Rakhra K, McRae S, Leiter J, Bell K, Macdonald P (2012) A multicenter randomized controlled trial comparing single-row with double-row fixation in arthroscopic rotator cuff repair. J Bone Joint Surg Am 94(14):1249–1257
McCormick F, Gupta A, Bruce B, Harris J, Abrams G, Wilson H, Hussey K, Cole BJ (2014) Single-row, double-row, and transosseous equivalent techniques for isolated supraspinatus tendon tears with minimal atrophy: a retrospective comparative outcome and radiographic analysis at minimum 2-year followup. Int J Shoulder Surg 8(1):15–20. doi:10.4103/0973-6042.131850
Millett PJ, Warth RJ, Dornan GJ, Lee JT, Spiegl UJ (2014) Clinical and structural outcomes after arthroscopic single-row versus double-row rotator cuff repair: a systematic review and meta-analysis of level I randomized clinical trials. J Shoulder Elb Surg 23(4):586–597
Saridakis P, Jones G (2010) Outcomes of single-row and double-row arthroscopic rotator cuff repair: a systematic review. J Bone Joint Surg Am 92(3):732–742. doi:10.2106/JBJS.I.01295
Zhang Q, Ge H, Zhou J, Yuan C, Chen K, Cheng B (2013) Single-row or double-row fixation technique for full-thickness rotator cuff tears: a meta-analysis. PLoS One 8(7):e68515. doi:10.1371/journal.pone.0068515
Jost PW, Khair MM, Chen DX, Wright TM, Kelly AM, Rodeo SA (2012) Suture number determines strength of rotator cuff repair. J Bone Joint Surg Am 94(14):e100
Lo IK, Burkhart SS, Chan KC, Athanasiou K (2004) Arthroscopic knots: determining the optimal balance of loop security and knot security. Arthroscopy 20(5):489–502
Pietschmann MF, Sadoghi P, Hauser E, Scharpf A, Gulecyuz MF, Schroder C, Jansson V, Muller PE (2011) Influence of testing conditions on primary stability of arthroscopic knot tying for rotator cuff repair: slippery when wet? Arthroscopy 27(12):1628–1636
Baumgarten KM, Wright RW (2004) Arthroscopic knot tying: an instruction manual. Lippincott Raven, New York
Ma CB, MacGillivray JD, Clabeaux J, Lee S, Otis JC (2004) Biomechanical evaluation of arthroscopic rotator cuff stitches. J Bone Joint Surg Am 86-A(6):1211–1216
Pietschmann MF, Frohlich V, Ficklscherer A, Gulecyuz MF, Wegener B, Jansson V, Muller PE (2009) Suture anchor fixation strength in osteopenic versus non-osteopenic bone for rotator cuff repair. Arch Orthop Trauma Surg 129(3):373–379. doi:10.1007/s00402-008-0689-4
Pietschmann MF, Gulecyuz MF, Fieseler S, Hentschel M, Rossbach B, Jansson V, Muller PE (2010) Biomechanical stability of knotless suture anchors used in rotator cuff repair in healthy and osteopenic bone. Arthroscopy 26(8):1035–1044
Millett PJ, Warth RJ, Dornan GJ, Lee JT, Spiegl UJ (2014) Clinical and structural outcomes after arthroscopic single-row versus double-row rotator cuff repair: a systematic review and meta-analysis of level I randomized clinical trials. J Shoulder Elb Surg 23(4):586–597. doi:10.1016/j.jse.2013.10.006
Burks RT, Crim J, Brown N, Fink B, Greis PE (2009) A prospective randomized clinical trial comparing arthroscopic single- and double-row rotator cuff repair: magnetic resonance imaging and early clinical evaluation. Am J Sports Med 37(4):674–682
Charousset C, Grimberg J, Duranthon LD, Bellaiche L, Petrover D (2007) Can a double-row anchorage technique improve tendon healing in arthroscopic rotator cuff repair?: a prospective, nonrandomized, comparative study of double-row and single-row anchorage techniques with computed tomographic arthrography tendon healing assessment. Am J Sports Med 35(8):1247–1253. doi:10.1177/0363546507301661
Franceschi F, Ruzzini L, Longo UG, Martina FM, Zobel BB, Maffulli N, Denaro V (2007) Equivalent clinical results of arthroscopic single-row and double-row suture anchor repair for rotator cuff tears: a randomized controlled trial. Am J Sports Med 35(8):1254–1260. doi:10.1177/0363546507302218
Reardon DJ, Maffulli N (2007) Clinical evidence shows no difference between single- and double-row repair for rotator cuff tears. Arthroscopy 23(6):670–673. doi:10.1016/j.arthro.2007.01.031
Bisson L, Zivaljevic N, Sanders S, Pula D (2015) A cost analysis of single-row versus double-row and suture bridge rotator cuff repair methods. Knee Surg Sports Traumatol Arthrosc 23(2):487–493. doi:10.1007/s00167-012-2338-2
Genuario JW, Donegan RP, Hamman D, Bell JE, Boublik M, Schlegel T, Tosteson AN (2012) The cost-effectiveness of single-row compared with double-row arthroscopic rotator cuff repair. J Bone Joint Surg Am 94(15):1369–1377. doi:10.2106/JBJS.J.01876
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Gülecyüz, M., Bortolotti, H., Pietschmann, M. et al. Primary stability of rotator cuff repair: can more suture materials yield more strength?. International Orthopaedics (SICOT) 40, 989–997 (2016). https://doi.org/10.1007/s00264-015-3012-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-015-3012-3