High failure rate of a new pressfit cup in mid-term follow-up

  • Original Paper
  • Published:
International Orthopaedics Aims and scope Submit manuscript

Abstract

Background

Unacceptably high rates of loosening of a novel cementless cup system (Mathys, SeleXys TH+) have been reported. In this study, our mid-term-results with the SeleXys TH+ cup are compared with an established pressfit cup (Ceraver, Cerafit-R).

Methods

Between 2008 and 2011, a total of 407 implantations, 19 of them bilaterally, were conducted on 388 patients with the implants in question. A total of 280 cups were available for follow-up. Of these, 160 SeleXys TH+ cups and 120 Cerafit-R cups had been used.

Results

The mean follow-up period was 48.6 months (min. 28, max. 75). A total of 18 cases of loosening were found in the SeleXys TH+ group. Of these, one was of a traumatic nature and one was of septic origin. This left 16 cases of aseptic, atraumatic loosening (10 % loosening rate). In the Cerafit-R group, only two cases (1.7 %) of loosening were seen over the same period. Both were of traumatic origin. No cases of aseptic, atraumatic loosening were seen in the follow-up period. Further, significantly more cases of loosening were observed in women than in men with the SeleXys cup.

Conclusions

The SeleXys TH+ cup shows an unacceptably high failure rate after six years. The time of revision is not limited to the first months after implantation, which means that close clinical and radiological controls will continue to be necessary in future in patients treated with SeleXys TH+ cups.

Level of Evidenve: III (retrospective cohort study)

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Hamadouche M, Boutin P, Daussange J et al (2002) Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study. J Bone Joint Surg Am 84-A:69–77

    PubMed  Google Scholar 

  2. Ihle M, Mai S, Pfluger D, Siebert W (2008) The results of the titanium-coated RM acetabular component at 20 years: a long-term follow-up of an uncemented primary total hip replacement. J Bone Joint Surg (Br) 90:1284–1290

    Article  CAS  Google Scholar 

  3. Petsatodis GE, Papadopoulos PP, Papavasiliou KA et al (2010) Primary cementless total hip arthroplasty with an alumina ceramic-on-ceramic bearing: results after a minimum of twenty years of follow-up. J Bone Joint Surg Am 92:639–644

    Article  PubMed  Google Scholar 

  4. Molloy D, Jack C, Esposito C, Walter WL (2012) A mid-term analysis suggests ceramic on ceramic hip arthroplasty is durable with minimal wear and low risk of squeak. HSS J 8:291–294

    Article  PubMed Central  PubMed  Google Scholar 

  5. Tozun IR, Ozden VE, Dikmen G, Beksac B (2014) Mid-term result of ceramic bearings in total hip arthroplasty. Int Orthop 38:2027–2031

    Article  PubMed  Google Scholar 

  6. Synder M, Drobniewski M, Sibiński M (2012) Long-term results of cementless hip arthroplasty with ceramic-on-ceramic articulation. Int Orthop 36:2225–2229

    Article  PubMed Central  PubMed  Google Scholar 

  7. Chevillotte C, Trousdale RT, An K-N et al (2012) Retrieval analysis of squeaking ceramic implants: are there related specific features? Orthop Traumatol Surg Res 98:281–287

    Article  CAS  PubMed  Google Scholar 

  8. Hannouche D, Zaoui A, Zadegan F et al (2011) Thirty years of experience with alumina-on-alumina bearings in total hip arthroplasty. Int Orthop 35:207–213

    Article  PubMed Central  PubMed  Google Scholar 

  9. Goebel P, Kluess D, Wieding J et al (2013) The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis. J Orthop Sci 18:264–270

    Article  PubMed  Google Scholar 

  10. Haverkamp D, Westerbos S, Campo MM et al (2013) Early loosening of a press-fit cup with ceramic-on-ceramic articulation: our early results. Arch Orthop Trauma Surg 133:1757–1762

    Article  PubMed  Google Scholar 

  11. Ilchmann T, Zwicky L, Gersbach S, Clauss M (2014) Poor outcome of a spherical pressfit cup with a modern ceramic liner: a prospective cohort study of 181 cups. Hip Int 24:333–339

    Article  PubMed  Google Scholar 

  12. Bauer R, Kerschbaumer F, Poisel S, Oberthaler W (1979) The transgluteal approach to the hip joint. Arch Orthop Trauma Surg 95:47–49

    Article  CAS  PubMed  Google Scholar 

  13. Boyer P, Huten D, Loriaut P et al (2010) Is alumina-on-alumina ceramic bearings total hip replacement the right choice in patients younger than 50 years of age? A 7- to 15-year follow-up study. Orthop Traumatol Surg Res 96:616–622

    Article  CAS  PubMed  Google Scholar 

  14. Nowak M, Nowak TE, Schmidt R et al (2011) Prospective study of a cementless total hip arthroplasty with a collum femoris preserving stem and a trabeculae oriented pressfit cup: minimun 6-year follow-up. Arch Orthop Trauma Surg 131:549–555

    Article  PubMed  Google Scholar 

  15. Schroeder K, Moehlenbruch A, Zimmermann-Stenzel M, Parsch D (2010) Five-year survival rate of the Allofit titanium press-fit cup. Orthopade 39:87–91

    Article  CAS  PubMed  Google Scholar 

  16. Streit MR, Weiss S, Andreas F et al (2014) 10-year results of the uncemented Allofit press-fit cup in young patients. Acta Orthop 85:368–374

    Article  PubMed Central  PubMed  Google Scholar 

  17. Hothan A, Huber G, Weiss C et al (2011) Deformation characteristics and eigenfrequencies of press-fit acetabular cups. Clin Biomech (Bristol, Avon) 26:46–51

    Article  Google Scholar 

  18. Maistrelli GL, Fornasier V, Binnington A et al (1991) Effect of stem modulus in a total hip arthroplasty model. J Bone Joint Surg (Br) 73:43–46

    CAS  Google Scholar 

  19. Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 108–113

  20. Le Cann S, Galland A, Rosa B et al (2014) Does surface roughness influence the primary stability of acetabular cups? A numerical and experimental biomechanical evaluation. Med Eng Phys 36:1185–1190

    Article  PubMed  Google Scholar 

  21. Schreiner U, Simnacher M, Scheller G, Scharf H-P (2007) The influence of different surface treatments on the primary stability of cementless acetabular cups: an in vitro study. Biomed Tech (Berl) 52:243–247

    Article  Google Scholar 

  22. Lass R, Kolb A, Skrbensky G et al (2014) A cementless hip system with a new surface for osseous integration. Int Orthop 38:703–709

    Article  PubMed Central  PubMed  Google Scholar 

  23. Chung YY, Kim HD, Kim KS (2002) Bone ingrowth on a smooth-surfaced hydroxyapatite-coated acetabular cup. Int Orthop 26:283–286

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  24. Nordin BE, MacGregor J, Smith DA (1966) The incidence of osteoporosis in normal women: its relation to age and the menopause. Q J Med 35:25–38

    CAS  PubMed  Google Scholar 

  25. Adler E, Stuchin SA, Kummer FJ (1992) Stability of press-fit acetabular cups. J Arthroplasty 7:295–301

    Article  CAS  PubMed  Google Scholar 

  26. Ries MD, Harbaugh M, Shea J, Lambert R (1997) Effect of cementless acetabular cup geometry on strain distribution and press-fit stability. J Arthroplasty 12:207–212

    Article  CAS  PubMed  Google Scholar 

  27. Toossi N, Adeli B, Timperley AJ et al (2013) Acetabular components in total hip arthroplasty: is there evidence that cementless fixation is better? J Bone Joint Surg Am 95:168–174

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Georg Matziolis.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Brodt, S., Matziolis, G., Windisch, C. et al. High failure rate of a new pressfit cup in mid-term follow-up. International Orthopaedics (SICOT) 39, 1813–1817 (2015). https://doi.org/10.1007/s00264-015-2872-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00264-015-2872-x

Keywords

Navigation