Abstract
Purpose
Press-fit cementless radial head implant longevity relies on adequate bone ingrowth. Failed implant osseointegration remains a clinical concern and has been shown to lead to prosthetic failure. The purpose of this study was to test the hypothesis that implants with sufficient initial press-fit stability would be less likely to fail due to implant pull-out, as demonstrated by an increasing amount of energy required to remove the prosthesis from the canal.
Methods
Ten cadaveric radii were implanted with five sizes (6–10 mm in 1-mm increments) of grit-blasted, cementless radial head stems. A customised slap hammer was used to measure the energy required to remove each stem. Stem-bone micromotion was also measured.
Results
The suboptimally sized stem (Max − 1) (i.e. 1 mm undersized) required less energy (0.5 ± 0 J) to pull out than the optimally sized stem (Max) (1.7 ± 0.3 J) (p = 0.008). The optimally sized stem demonstrated greater initial stability (45 ± 7 μm) than the suboptimally sized stem (79 ± 12 μm) (p = 0.004).
Conclusions
This investigation demonstrates the importance of obtaining adequate press-fit stability for the prevention of radial head stem pull-out failure. These data add to the relatively scant knowledge in the literature regarding radial head biomechanics. The energy required to remove a prosthetic radial head ingrowth stem decreases in conjunction with diameter. The use of an inadequately sized stem increases the stem’s micromotion as well as the risk of prosthetic loosening due to pull-out.
Similar content being viewed by others
References
Grewal R, MacDermid JC, Faber KJ, Drosdowech DS, King GJ (2006) Comminuted radial head fractures treated with a modular metallic radial head arthroplasty. Study of outcomes. J Bone Joint Surg Am 88(10):2192–2200. doi:10.2106/JBJS.E.00962
Ashwood N, Bain GI, Unni R (2004) Management of Mason type-III radial head fractures with a titanium prosthesis, ligament repair, and early mobilization. J Bone Joint Surg Am 86-A(2):274–280
Moro JK, Werier J, MacDermid JC, Patterson SD, King GJ (2001) Arthroplasty with a metal radial head for unreconstructible fractures of the radial head. J Bone Joint Surg Am 83-A(8):1201–1211
Burkhart KJ, Mattyasovszky SG, Runkel M, Schwarz C, Küchle R, Hessmann MH, Rommens PM, Lars MP (2010) Mid- to long-term results after bipolar radial head arthroplasty. J Shoulder Elbow Surg 19(7):965–972. doi:10.1016/j.jse.2010.05.02
Ling RS (1986) Observations on the fixation of implants to the bony skeleton. Clin Orthop Relat Res 210:80–96
Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113
Moon JG, Berglund LJ, Domire Z, An KN, O’Driscoll SW (2009) Stem diameter and micromotion of press fit radial head prosthesis: a biomechanical study. J Shoulder Elbow Surg 18(5):785–790. doi:10.1016/j.jse.2009.02.014
Chanlalit C, Fitzsimmons JS, Shukla DR, An KN, O’Driscoll SW (2011) Micromotion of plasma spray versus grit-blasted radial head prosthetic stem surfaces. J Shoulder Elbow Surg 20(5):717–722. doi:10.1016/j.jse.2010.11.010
Ferreira LM, Stacpoole RA, Johnson JA, King GJ (2010) Cementless fixation of radial head implants is affected by implant stem geometry: an in vitro study. Clin Biomech (Bristol, Avon) 25(5):422–426. doi:10.1016/j.clinbiomech.2010.02.001
O’Driscoll SW, Herald JA (2012) Forearm pain associated with loose radial head prostheses. J Shoulder Elbow Surg 21:92–97. doi:10.1016/j.jse.2011.05.008
Chanlalit C, Shukla DR, Fitzsimmons JS, An KN, O’Driscoll SW (2011) Effect of hoop stress fracture on micromotion of textured ingrowth stems for radial head replacement. J Shoulder Elbow Surg 21(7):949–954. doi:10.1016/j.jse.2011.05.001
Chanlalit C, Fitzsimmons JS, Moon JG, Berglund LJ, An KN, O’Driscoll SW (2011) Radial head prosthesis micromotion characteristics: partial versus fully grit-blasted stems. J Shoulder Elbow Surg 20(1):27–32. doi:10.1016/j.jse.2010.05.030
Shukla DR, Fitzsimmons JS, An KN, O’Driscoll SW (2012) Effect of stem length on prosthetic radial head micromotion. J Shoulder Elbow Surg 21(11):1559–1564. doi:10.1016/j.jse.2011.05.009
Acknowledgements
This study was funded by the Mayo Foundation. The prosthetic components used in this study were provided by Acumed, LLC. This study was not supported by any outside funding or grants. The Mayo Clinic Institutional Review Board (IRB) that convened on 17 December 2010 approved the project, entitled “Prosthetic Radial Head Stability” (IRB protocol number 01–008186).
Conflict of interest
The authors would like to disclose that the senior author (SOD) and the Mayo Foundation receive royalties from commercial entities related to the subject of this article.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Shukla, D., Fitzsimmons, J., An, KN. et al. Prosthetic radial head stem pull-out as a mode of failure: a biomechanical study. International Orthopaedics (SICOT) 38, 89–93 (2014). https://doi.org/10.1007/s00264-013-2074-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00264-013-2074-3