Hydroxyapatite-coated RM cup in primary hip arthroplasty

  • Original Paper
  • Published:
International Orthopaedics Aims and scope Submit manuscript

Abstract.

We reviewed 113 patients with 127 primary total hip arthroplasties using a hydroxyapatite-coated RM (Robert Mathys) cup. Average patient age was 61 (30–70) years and mean follow-up was 9.4 (6–13) years. Four patients were lost to follow-up and nine patients died of unrelated causes during the course of the study. Three acetabular components were revised; two cups were radiologically loose and one was revised for marginal erosion. No radiolucency was noted in any of the three zones in the remaining cups. There were no cases of cup migration or dislocation. The life table method predicted a 10-year survival rate of 97.9% with revision of the acetabular component for any cause as the endpoint.

Résumé.

Nous avons examiné 113 malades avec 127 arthroplasties primaires de la hanche qui utilisent une cupule RM (Robert Mathys) recouverte d' hydroxyapatite. L'âge moyen de malades était 61 ans (30–70). Le temps de suivi moyen était 9.4 (6–13) années. Quatre malades ont été perdus de vue. Neuf malades sont morts de causes sans rapport pendant le cours de l'étude. Trois cupules ont été révisées. Deux des cupules révisées étaient radiologiquement descellées tachent; une troisième a été révisée pour érosion marginale. Aucun liseré radiologique n'a été noté dans les trois zones sur les cupules restantes. Il n'y avait pas de migration de cupule ni de luxation. La méthode actuarielle prédit un taux de survie à 10 années de 97.9% avec la révision du composant acetabulaire comme événement, quelqu'en soit la cause.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ali, S.M., Kumar, .A. Hydroxyapatite-coated RM cup in primary hip arthroplasty. International Orthopaedics (SICOT) 27, 90–93 (2003). https://doi.org/10.1007/s00264-002-0410-0

Download citation

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00264-002-0410-0

Keywords

Navigation