Abstract.
Cementless acetabular components require good initial fixation to allow bony in-growth. The initial press-fit stability, important for designs that do not rely on supplemental fixation, was examined for three designs with different porous coating systems: beads, fiber mesh, and plasma spray. The ability to withstand tangential loads was determined (maximum rim loads: 122–1730 N). The plasma-sprayed acetabular cups withstood the greatest tangential load within the specified range of motion of 150 µm (P<0.01). Differences in surface preparation of titanium acetabular cups may significantly affect the initial stability of the implants in rim loading.
Résumé.
Les composants acétabulaires non-cimentés exigent une bonne fixation initiale pour permettre une repousse osseuse. La stabilité initiale après impaction, qui est importante pour les cupules qui n'ont pas de fixation complémentaire, a été examinée pour trois types de cupules avec un revêtement poreux différent : billes, filet et plasma vaporisé. La capacité à supporter des charges tangentielles a été déterminée (charges maximales du bord: 122–1730 N). Les cupules plasma – vaporisées ont supporté la plus grande charge tangentielle dans l'amplitude de mobilité de 150 µm (P <0.01). Les différences dans la préparation de la surface en titane peuvent affecter considérablement la stabilité initiale des implants soumis à des contraintes de bord.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Electronic Publication
Rights and permissions
About this article
Cite this article
Markel, D.C., Hora, N. & Grimm, M. Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systems. International Orthopaedics (SICOT) 26, 72–75 (2002). https://doi.org/10.1007/s00264-001-0314-4
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s00264-001-0314-4