Ultrasound findings in asymptomatic patients with modular metal on metal total hip arthroplasty

  • Scientific Article
  • Published:
Skeletal Radiology Aims and scope Submit manuscript

Abstract

Objective

The use of metal-on-metal and modular total hip arthroplasty is associated with potentially serious local and systemic complications. The primary aim of this study was to identify the prevalence of a pseudotumor in asymptomatic patients with a particular metal-on-metal hip prosthesis after a minimum follow-up of 5 years using ultrasound evaluation. A secondary purpose was to identify associations between the presence of pseudotumor and serum metal ion levels following implantation.

Methods

We prospectively evaluated data collected from 36 asymptomatic patients who underwent implantation of a Profemur Z metal-on-metal total hip arthroplasty from January 2004 to January 2010. Serum metal ion levels were collected in 2012 and 2015. Hip ultrasounds were performed in 2015.

Results

Pseudotumors were found in 7/36 patients (19.4%). The average pseudotumor size measured 38.2 cm3 (range 7.35 cm3–130.81 cm3). Elevated metal ion levels were found in all patients at all time points. No statistical correlation was found between the presence of pseudotumor and patient age, age of the implant, component design, and any of the serum metal ion levels or ratios.

Conclusions

One in every five asymptomatic patients with metal-on-metal implants was found to have a periarticular pseudotumor. There was no dose-dependent relationship found between elevated serum metal ion levels and the development of a pseudotumor. Our findings suggest that in patients with known elevated metal ion levels, continued monitoring of ion levels may not be a reliable predictor of pseudotumor formation, and ultrasound surveillance can and should be routinely used to document the presence and progression of pseudotumor.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17:649–61.

    Article  PubMed  Google Scholar 

  2. Harris WH. Wear and periprosthetic osteolysis: the problem. Clin Orthop Relat Res. 2001;66–70.

  3. Antoniou J, Zukor DJ, Mwale F, Minarik W, Petit A, Huk OL. Metal ion levels in the blood of patients after hip resurfacing: a comparison between twenty-eight and thirty-six-millimeter-head metal-on-metal prostheses. J Bone Joint Surg Am. 2008;90:142–8.

    Article  PubMed  Google Scholar 

  4. Brodner W, Bitzan P, Meisinger V, Kaider A, Gottsauner-Wolf F, Kotz R. Serum cobalt levels after metal-on-metal total hip arthroplasty. J Bone Joint Surg Am. 2003;85-A:2168–73.

    Article  PubMed  Google Scholar 

  5. Daniel J, Ziaee H, Pradhan C, McMinn DJ. Systemic metal exposure in large- and small-diameter metal-on-metal total hip replacements. Orthopedics. 2008; 31:Suppl 2.

  6. Dobbs HS, Minski MJ. Metal ion release after total hip replacement. Biomaterials. 1980;1:193–8.

    Article  CAS  PubMed  Google Scholar 

  7. Grubl A, Marker M, Brodner W, et al. Long-term follow-up of metal-on-metal total hip replacement. J Orthop Res. 2007;25:841–8.

    Article  PubMed  Google Scholar 

  8. Grubl A, Weissinger M, Brodner W, et al. Serum aluminium and cobalt levels after ceramic-on-ceramic and metal-on-metal total hip replacement. J Bone Joint Surg Br. 2006;88:1003–5.

    Article  CAS  PubMed  Google Scholar 

  9. Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008;90:1143–51.

    Article  CAS  PubMed  Google Scholar 

  10. Sauve P, Mountney J, Khan T, De Beer J, Higgins B, Grover M. Metal ion levels after metal-on-metal Ring total hip replacement: a 30-year follow-up study. J Bone Joint Surg Br. 2007;89:586–90.

    Article  CAS  PubMed  Google Scholar 

  11. Vendittoli PA, Mottard S, Roy AG, Dupont C, Lavigne M. Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip. J Bone Joint Surg Br. 2007;89:441–8.

    Article  PubMed  Google Scholar 

  12. Chang EY, McAnally JL, Van Horne JR, et al. Relationship of plasma metal ions and clinical and imaging findings in patients with ASR XL metal-on-metal total hip replacements. J Bone Joint Surg Am. 2013;95:2015–20.

    Article  PubMed  Google Scholar 

  13. Daniel J, Holland J, Quigley L, Sprague S, Bhandari M. Pseudotumors associated with total hip arthroplasty. J Bone Joint Surg Am. 2012;94:86–93.

    Article  PubMed  Google Scholar 

  14. van der Weegen W, Sijbesma T, Hoekstra HJ, Brakel K, Pilot P, Nelissen RG. Treatment of pseudotumors after metal-on-metal hip resurfacing based on magnetic resonance imaging, metal ion levels and symptoms. J Arthroplasty. 2014;29:416–21.

    Article  PubMed  Google Scholar 

  15. Tower SS. Arthroprosthetic cobaltism: neurological and cardiac manifestations in two patients with metal-on-metal arthroplasty: a case report. J Bone Joint Surg Am. 2010;92:2847–51.

    Article  PubMed  Google Scholar 

  16. Zywiel MG, Brandt JM, Overgaard CB, Cheung AC, Turgeon TR, Syed KA. Fatal cardiomyopathy after revision total hip replacement for fracture of a ceramic liner. Bone Joint J. 2013;95-B:31–7.

    Article  CAS  PubMed  Google Scholar 

  17. Wimmer MA, Fischer A, Buscher R, et al. Wear mechanisms in metal-on-metal bearings: the importance of tribochemical reaction layers. J Orthop Res. 2010;28:436–43.

    PubMed  Google Scholar 

  18. Hart AJ, Satchithananda K, Liddle AD, et al. Pseudotumors in association with well-functioning metal-on-metal hip prostheses: a case-control study using three-dimensional computed tomography and magnetic resonance imaging. J Bone Joint Surg Am. 2012;94:317–25.

    Article  PubMed  Google Scholar 

  19. Grammatopoulos G, Pandit H, Kamali A, et al. The correlation of wear with histological features after failed hip resurfacing arthroplasty. J Bone Joint Surg Am. 2013;95, e81.

    Article  PubMed  Google Scholar 

  20. Pandit H, Glyn-Jones S, McLardy-Smith P, et al. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–51.

    Article  CAS  PubMed  Google Scholar 

  21. Kwon YM, Glyn-Jones S, Simpson DJ, et al. Analysis of wear of retrieved metal-on-metal hip resurfacing implants revised due to pseudotumours. J Bone Joint Surg Br. 2010;92:356–61.

    Article  PubMed  Google Scholar 

  22. Langton DJ, Jameson SS, Joyce TJ, et al. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93:1011–6.

    Article  CAS  PubMed  Google Scholar 

  23. Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.

    Article  CAS  PubMed  Google Scholar 

  24. Lavigne M, Belzile EL, Roy A, Morin F, Amzica T, Vendittoli PA. Comparison of whole-blood metal ion levels in four types of metal-on-metal large-diameter femoral head total hip arthroplasty: the potential influence of the adapter sleeve. J Bone Joint Surg Am. 2011;93:128–36.

    Article  PubMed  Google Scholar 

  25. Williams DH, Greidanus NV, Masri BA, Duncan CP, Garbuz DS. Prevalence of pseudotumor in asymptomatic patients after metal-on-metal hip arthroplasty. J Bone Joint Surg Am. 2011;93:2164–71.

    Article  PubMed  Google Scholar 

  26. Grammatopoulos G, Pandit H, Kwon YM, et al. Hip resurfacings revised for inflammatory pseudotumour have a poor outcome. J Bone Joint Surg Br. 2009;91:1019–24.

    Article  CAS  PubMed  Google Scholar 

  27. Munro JT, Masri BA, Duncan CP, Garbuz DS. High complication rate after revision of large-head metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2014;472:523–8.

    Article  PubMed  Google Scholar 

  28. CMS.gov. CMS Physician Fee Schedule Online Database. In, 2016.

  29. Garbuz DS, Hargreaves BA, Duncan CP, Masri BA, Wilson DR, Forster BB. The John Charnley Award: Diagnostic accuracy of MRI versus ultrasound for detecting pseudotumors in asymptomatic metal-on-metal THA. Clin Orthop Relat Res. 2014;472:417–23.

    Article  PubMed  Google Scholar 

  30. Nam D, Barrack RL, Potter HG. What are the advantages and disadvantages of imaging modalities to diagnose wear-related corrosion problems? Clin Orthop Relat Res. 2014;472:3665–73.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Dimitreu DDN, Antoci V, Liow M, Tsai T, Li G, Rubash H, et al. Is ultrasound as useful as MARS MRI in the Longitudinal surveillance of Metal-on-Metal hip patients? Orlando: Poster session presented at: American Academy of Orthopaedic Surgeons 2016 Annual Meeting; 2016.

    Google Scholar 

  32. Cooper HJ, Urban RM, Wixson RL, Meneghini RM, Jacobs JJ. Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Joint Surg Am. 2013;95:865–72.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Kop AM, Keogh C, Swarts E. Proximal component modularity in THA—at what cost? An implant retrieval study. Clin Orthop Relat Res. 2012;470:1885–94.

    Article  CAS  PubMed  Google Scholar 

  34. Jacobs JJ, Cooper HJ, Urban RM, Wixson RL, Della Valle CJ. What do we know about taper corrosion in total hip arthroplasty? J Arthroplasty. 2014;29:668–9.

    Article  CAS  PubMed  Google Scholar 

  35. Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res. 2010;468:318–25.

    Article  PubMed  Google Scholar 

  36. Muraoka K, Naito M, Nakamura Y, Hagio T, Takano K. Usefulness of ultrasonography for detection of pseudotumors after metal-on-metal total hip arthroplasty. J Arthroplasty. 2015;30:879–84.

    Article  PubMed  Google Scholar 

  37. van Holsbeeck MT, Eyler WR, Sherman LS, et al. Detection of infection in loosened hip prostheses: efficacy of sonography. AJR Am J Roentgenol. 1994;163:381–4.

    Article  PubMed  Google Scholar 

  38. Bishop N, Witt F, Pourzal R, et al. Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings. J Orthop Res. 2013;31:1116–22.

    Article  PubMed  Google Scholar 

  39. Gill IP, Webb J, Sloan K, Beaver RJ. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation. J Bone Joint Surg Br. 2012;94:895–900.

    Article  CAS  PubMed  Google Scholar 

  40. Matthies AK, Skinner JA, Osmani H, Henckel J, Hart AJ. Pseudotumors are common in well-positioned low-wearing metal-on-metal hips. Clin Orthop Relat Res. 2012;470:1895–906.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Dr. Lisa Walker for her assistance in the completion of this project. The authors would also like to thank Meredith Mahan for her assistance with the statistical analysis.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Nicholas B. Frisch.

Ethics declarations

Conflicts of interest

None.

Support

Internal.

Institutional review board status

Approved.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Frisch, N.B., Wessell, N.M., Taliaferro, K. et al. Ultrasound findings in asymptomatic patients with modular metal on metal total hip arthroplasty. Skeletal Radiol 46, 641–649 (2017). https://doi.org/10.1007/s00256-017-2592-y

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00256-017-2592-y

Keywords

Navigation