Evaluation of a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography of scaphoid fixation screws

  • Lukas Filli
  • Magda Marcon
  • Bernhard Scholz
  • Maurizio Calcagni
  • Tim Finkenstädt
  • Gustav Andreisek
  • Roman Guggenberger
Scientific Article

Abstract

Objectives

The aim of this study was to evaluate a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography (FDCT) of scaphoid fixation screws. FDCT has gained interest in imaging small anatomic structures of the appendicular skeleton. Angiographic C-arm systems with flat detectors allow fluoroscopy and FDCT imaging in a one-stop procedure emphasizing their role as an ideal intraoperative imaging tool. However, FDCT imaging can be significantly impaired by artefacts induced by fixation screws.

Materials and methods

Following ethical board approval, commercially available scaphoid fixation screws were inserted into six cadaveric specimens in order to fix artificially induced scaphoid fractures. FDCT images corrected with the algorithm were compared to uncorrected images both quantitatively and qualitatively by two independent radiologists in terms of artefacts, screw contour, fracture line visibility, bone visibility, and soft tissue definition. Normal distribution of variables was evaluated using the Kolmogorov–Smirnov test. In case of normal distribution, quantitative variables were compared using paired Student's t tests. The Wilcoxon signed-rank test was used for quantitative variables without normal distribution and all qualitative variables. A p value of < 0.05 was considered to indicate statistically significant differences.

Results

Metal artefacts were significantly reduced by the correction algorithm (p < 0.001), and the fracture line was more clearly defined (p < 0.01). The inter-observer reliability was “almost perfect” (intra-class correlation coefficient 0.85, p < 0.001).

Conclusions

The prototype correction algorithm in FDCT for metal artefacts induced by scaphoid fixation screws may facilitate intra- and postoperative follow-up imaging.

Key Points

Flat detector computed tomography (FDCT) is a helpful imaging tool for scaphoid fixation.

The correction algorithm significantly reduces artefacts in FDCT induced by scaphoid fixation screws.

This may facilitate intra- and postoperative follow-up imaging.

Keywords

Tomography scanners, X-ray computed Scaphoid bone Fracture fixation, internal Artefacts Algorithms 

Abbreviations

FDCT

Flat detector computed tomography

HU

Hounsfield units

ICC

Intra-class correlation coefficient

kV

Kilovolts

mAs

Milliampere-seconds

MDCT

Multi-detector computed tomography

ROI

Region of interest

Notes

Disclosures

One author (BS) is an employee of the vendor of the prototype algorithm, but had no influence on the measurements and results of this study. The other authors had full control of the data and declare no conflict of interest.

References

  1. 1.
    Sendher R, Ladd AL. The scaphoid. Orthop Clin North Am. 2013;44(1):107–20. doi: 10.1016/j.ocl.2012.09.003.PubMedCrossRefGoogle Scholar
  2. 2.
    Oduwole KO, Cichy B, Dillon JP, Wilson J, O’Beirne J. Acutrak versus Herbert screw fixation for scaphoid non-union and delayed union. J Orthop Surg (Hong Kong). 2012;20(1):61–5.Google Scholar
  3. 3.
    Gutow AP. Percutaneous fixation of scaphoid fractures. J Am Acad Orthop Surg. 2007;15(8):474–85.PubMedGoogle Scholar
  4. 4.
    Temple CL, Ross DC, Bennett JD, Garvin GJ, King GJ, Faber KJ. Comparison of sagittal computed tomography and plain film radiography in a scaphoid fracture model. J Hand Surg. 2005;30(3):534–42. doi: 10.1016/j.jhsa.2005.01.001.CrossRefGoogle Scholar
  5. 5.
    Lozano-Calderon S, Blazar P, Zurakowski D, Lee SG, Ring D. Diagnosis of scaphoid fracture displacement with radiography and computed tomography. J Bone Joint Surg Am Volume. 2006;88(12):2695–703. doi: 10.2106/JBJS.E.01211.CrossRefGoogle Scholar
  6. 6.
    Smith M, Bain GI, Turner PC, Watts AC. Review of imaging of scaphoid fractures. ANZ J Surg. 2010;80(1–2):82–90. doi: 10.1111/j.1445-2197.2009.05204.x.PubMedCrossRefGoogle Scholar
  7. 7.
    Kalender WA, Kyriakou Y. Flat-detector computed tomography (FD-CT). Eur Radiol. 2007;17(11):2767–79. doi: 10.1007/s00330-007-0651-9.PubMedCrossRefGoogle Scholar
  8. 8.
    Kyriakou Y, Richter G, Dorfler A, Kalender WA. Neuroradiologic applications with routine C-arm flat panel detector CT: evaluation of patient dose measurements. AJNR Am J Neuroradiol. 2008;29(10):1930–6. doi: 10.3174/ajnr.A1237.PubMedCrossRefGoogle Scholar
  9. 9.
    Gupta R, Cheung AC, Bartling SH, et al. Flat-panel volume CT: fundamental principles, technology, and applications. Radiographics. 2008;28(7):2009–22. doi: 10.1148/rg.287085004.PubMedCrossRefGoogle Scholar
  10. 10.
    Prell D, Kyriakou Y, Beister M, Kalender WA. A novel forward projection-based metal artifact reduction method for flat-detector computed tomography. Phys Med Biol. 2009;54(21):6575–91. doi: 10.1088/0031-9155/54/21/009.PubMedCrossRefGoogle Scholar
  11. 11.
    Liu PT, Pavlicek WP, Peter MB, Spangehl MJ, Roberts CC, Paden RG. Metal artifact reduction image reconstruction algorithm for CT of implanted metal orthopedic devices: a work in progress. Skeletal Radiol. 2009;38(8):797–802. doi: 10.1007/s00256-008-0630-5.PubMedCrossRefGoogle Scholar
  12. 12.
    Mahnken AH, Raupach R, Wildberger JE, et al. A new algorithm for metal artifact reduction in computed tomography: in vitro and in vivo evaluation after total hip replacement. Investig Radiol. 2003;38(12):769–75. doi: 10.1097/01.rli.0000086495.96457.54.CrossRefGoogle Scholar
  13. 13.
    Guggenberger R, Winklhofer S, Osterhoff G, et al. Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. Eur Radiol. 2012;22(11):2357–64. doi: 10.1007/s00330-012-2501-7.PubMedCrossRefGoogle Scholar
  14. 14.
    Wang G, Frei T, Vannier MW. Fast iterative algorithm for metal artifact reduction in X-ray CT. Acad Radiol. 2000;7(8):607–14.PubMedCrossRefGoogle Scholar
  15. 15.
    Bal M, Spies L. Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering. Med Phys. 2006;33(8):2852–9.PubMedCrossRefGoogle Scholar
  16. 16.
    Lemmens C, Faul D, Nuyts J. Suppression of metal artifacts in CT using a reconstruction procedure that combines MAP and projection completion. IEEE Trans Med Imaging. 2009;28(2):250–60. doi: 10.1109/TMI.2008.929103.PubMedCrossRefGoogle Scholar
  17. 17.
    Meyer E, Raupach R, Lell M, Schmidt B, Kachelriess M. Frequency split metal artifact reduction (FSMAR) in computed tomography. Med Phys. 2012;39(4):1904–16. doi: 10.1118/1.3691902.PubMedCrossRefGoogle Scholar
  18. 18.
    Wang G, Snyder DL, O’Sullivan JA, Vannier MW. Iterative deblurring for CT metal artifact reduction. IEEE Trans Med Imaging. 1996;15(5):657–64. doi: 10.1109/42.538943.PubMedCrossRefGoogle Scholar
  19. 19.
    Zhang X, Wang J, Xing L. Metal artifact reduction in x-ray computed tomography (CT) by constrained optimization. Med Phys. 2011;38(2):701–11.PubMedCrossRefPubMedCentralGoogle Scholar
  20. 20.
    Prell D, Kalender WA, Kyriakou Y. Development, implementation and evaluation of a dedicated metal artefact reduction method for interventional flat-detector CT. Brit J Radiol. 2010;83(996):1052–62. doi: 10.1259/bjr/19113084.PubMedCrossRefPubMedCentralGoogle Scholar
  21. 21.
    Buhk JH, Groth M, Sehner S, Fiehler J, Schmidt NO, Grzyska U. Application of a Novel Metal Artifact Correction Algorithm in Flat-Panel CT After Coil Embolization of Brain Aneurysms: Intraindividual Comparison. RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2013;185(9):824–9. doi: 10.1055/s-0033-1350117.PubMedCrossRefGoogle Scholar
  22. 22.
    Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.PubMedCrossRefGoogle Scholar
  23. 23.
    Suzuki S, Furui S, Yamaguchi I, et al. Entrance surface dose during three-dimensional imaging with a flat-panel detector angiography system. J Vasc Interv Radiol : JVIR. 2008;19(9):1361–5. doi: 10.1016/j.jvir.2008.05.020.PubMedCrossRefGoogle Scholar
  24. 24.
    Chida K, Inaba Y, Saito H, et al. Radiation dose of interventional radiology system using a flat-panel detector. AJR Am J Roentgenol. 2009;193(6):1680–5. doi: 10.2214/AJR.09.2747.PubMedCrossRefGoogle Scholar
  25. 25.
    Guggenberger R, Morsbach F, Alkadhi H, et al. C-arm flat-panel CT arthrography of the wrist and elbow: first experiences in human cadavers. Skeletal Radiol. 2013;42(3):419–29. doi: 10.1007/s00256-012-1501-7.PubMedCrossRefGoogle Scholar
  26. 26.
    Guggenberger R, Winklhofer S, Spiczak JV, Andreisek G, Alkadhi H. In vitro high-resolution flat-panel computed tomographic arthrography for artificial cartilage defect detection: comparison with multidetector computed tomography. Investig Radiol. 2013;48(8):614–21. doi: 10.1097/RLI.0b013e318289fa78.CrossRefGoogle Scholar

Copyright information

© ISS 2014

Authors and Affiliations

  • Lukas Filli
    • 1
  • Magda Marcon
    • 1
    • 2
  • Bernhard Scholz
    • 3
  • Maurizio Calcagni
    • 4
  • Tim Finkenstädt
    • 1
  • Gustav Andreisek
    • 1
  • Roman Guggenberger
    • 1
  1. 1.Department of Diagnostic and Interventional RadiologyUniversity Hospital of ZurichZurichSwitzerland
  2. 2.Institute of Diagnostic Radiology, Department of Medical and Biological SciencesUniversity of UdineUdineItaly
  3. 3.Siemens AG, Healthcare SectorImaging & Therapy DivisionForchheimGermany
  4. 4.Division of Plastic Surgery and Hand SurgeryUniversity Hospital of ZurichZurichSwitzerland

Personalised recommendations