Abstract
Objective. Fluctuating equilibrium magnetic resonance (FEMR) is a rapid three-dimensional (3D) imaging sequence with high signal-to-noise ratio (SNR). FEMR may be useful for detecting cartilage defects in the knee. At 1.5 T, FEMR uses a TR with odd multiples of 2.2 ms for fat/water separation. With a TR of 6.6 ms, high-resolution 3D imaging of cartilage is possible.
Design and patients. The knees of 10 volunteers and two patients were imaged on a GE Signa 1.5 T scanner using an extremity coil. Scans were preceded by a shimming sequence optimizing linear terms. Subjects were imaged with FEMR, proton-density fast spin-echo (PD-FSE), T2-weighted fast spin-echo (T2-FSE), and 3D fat-suppressed spoiled-gradient-recalled echo (3D-SPGR).
Results. SNR and contrast-to-noise efficiency measurements for cartilage using FEMR were superior to those using PD-FSE, T2-FSE, and 3D-FS-SPGR. FSE images showed bright synovial fluid with limited cartilage detail. 3D-SPGR had comparable resolution to FEMR but suboptimal cartilage/fluid contrast and longer scan times (8 min versus 2 min). Cartilage surface detail, outlined by bright synovial fluid, was best seen on the FEMR images.
Discussion. FEMR obtains high-resolution 3D images of the entire knee in 2 min with excellent cartilage/fluid contrast. FEMR is sensitive to field inhomogeneity and requires shimming. Surface defects are outlined by bright synovial fluid, and cartilage has higher signal-to-noise efficiency compared with PD-FSE, T2-FSE, and 3D-SPGR techniques.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Electronic Publication
Rights and permissions
About this article
Cite this article
Vasnawala, S.S., Pauly, J.M., Nishimura, D.G. et al. MR imaging of knee cartilage with FEMR. Skeletal Radiol 31, 574–580 (2002). https://doi.org/10.1007/s00256-002-0562-4
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s00256-002-0562-4