Abstract
Thermal motions of microscopic probes limit the possibilities of experiments that are designed to resolve single-macromolecule dynamics in aqueous conditions. We investigate theoretical strategies for maximizing signal-to-noise ratios or resolution in typical situations, illustratin+g our discussion with examples from optical tweezers and atomic force microscopy experiments. A central result is that the viscous drag on a micromechanical probe is more important than the compliance of the probe. Within limits, increased stiffness of an AFM cantilever or of an optical trap does not increase resolution, and decreased stiffness does not provide the possibility of less invasive measurements.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Received: 15 August 1997 / Accepted: 11 September 1997
Rights and permissions
About this article
Cite this article
Gittes, F., Schmidt, C. Thermal noise limitations on micromechanical experiments. Eur Biophys J 27, 75–81 (1998). https://doi.org/10.1007/s002490050113