Dose reduction fluoroscopy in pediatrics

  • Original Article
  • Published:
Pediatric Radiology Aims and scope Submit manuscript

Abstract

Background. It is essential that we find ways to reduce radiation exposure to children and maintain image quality.

Objectives. We compared radiation dose, image quality, and spatial resolution when continuous and pulse fluoroscopy with a full and half dose are applied to a phantom. The film-screen technique was compared to fluoroscopy with the digitized spot technique (fluoro grab image) in procedures such as voiding cystourethrogram (VCUG).

Materials and methods. Using a 15.1-cm Plexiglas phantom, we obtained dosimetry in milligrays (mGy), spatial resolution in number of line pairs per millimeter (lp/mm), and threshold contrast resolution in number of visible holes. To measure total radiation dose, we calculated the average elapsed fluoroscopy time for VCUG to be approximately 3 min and estimated the average number of exposures as 10. Dosimetry was obtained for full dose and half dose continuous, for 15 pulses per second (pps), 7.5 pps, and 3.75 pps. These were also calculated with normal, magnification 1, and magnification 2 factors.

Results. Results of the two most relevant parameters are shown: continuous full-dose fluoroscopy, 3 min, 10 photo spots, total dose of 28.7 mGy with 2 lp/mm of resolution and a threshold contrast of 2.2%, versus 3.75 pps half-dose fluoroscopy, 3 min, grab images, total dose of 3.7 mGy with 1.9 lp/mm of resolution and a threshold contrast of 2.3%.

Conclusion. With minimal loss of resolution there is significant dose reduction (87%) when using 3.75 pps with digitized imaging.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lederman, H.M., Khademian, Z.P., Felice, M. et al. Dose reduction fluoroscopy in pediatrics. Ped Radiol 32, 844–848 (2002). https://doi.org/10.1007/s00247-002-0696-5

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00247-002-0696-5

Navigation