Abstract
Insertable cardiac monitors (ICM) are subcutaneously implanted devices that monitor a patient’s heart rate and rhythm (Rossano in Pediatrics 112(3):e228, 2003). The diagnostic accuracy and safety of the Confirm RxTM (Abbott, Minneapolis, MN) ICM in pediatric patients is unknown. This is a single center, retrospective, IRB-approved review of patients ≤ 21 years implanted with Confirm RxTM ICMs from 2017 to 2020. Data collected included demographics, indications, presence of P-wave and R-wave amplitude at implantation and follow-up, number/appropriateness of transmissions pre and post implementation of SharpSenseTM technology, reprogramming to improve accuracy, time from implantation to arrhythmia detection, and complications. There were 29 patients (median age: 8 years, 59% females). P-waves were identified in all patients and average R-wave amplitude was 0.85 mV (0.26–1.03 mV). There was no significant difference in R-wave amplitude based on size (BSA ≥ 1.5 m2: 0.76 mV, < 1.5 m2: 0.91 mV) or congenital heart disease (+CHD: 0.86 mV, −CHD: 0.85 mV). Arrhythmias identified were the following: wide complex tachycardia (1), supraventricular tachycardia (4), bradycardia/sinus pause (3), and premature ventricular contraction (1). SharpSenseTM implementation decreased the false-positive rate in device-initiated transmissions (55.4% to 0%, p < 0.00001). Average time from implantation to arrhythmia detection was 2.63 months (range: 0–8.8). A single complication of cellulitis occurred. Confirm RxTM is appropriate for implant in pediatric patients regardless of age, BSA, or CHD. Implementation of SharpSense™ technology dramatically decreased the false-positive rate. Follow-up studies could utilize additional monitoring devices to provide analysis on potential events that the Confirm RxTM ICM missed.
Similar content being viewed by others
References
Rossano J, Bloemers B, Sreeram N, Balaji S, Shah MJ (2003) Efficacy of implantable loop recorders in establishing symptom-rhythm correlation in young patients with syncope and palpitations. Pediatrics 112(3):e228
Bisignani A, De Bonis S, Mancuso L, Ceravolo G, Bisignani G (2019) Implantable loop recorder in clinical practice. J Arrhythm 35(1):25–32
Giancaterino S, Lupercio F, Nishimura M, Hsu JC (2018) Current and future use of insertable cardiac monitors. JACC Clin Electrophysiol 4(11):1383–1396
Padmanabhan D, Kancharla K, El-Harasis MA et al (2019) Diagnostic and therapeutic value of implantable loop recorder: a tertiary care center experience. Pacing Clin Electrophysiol 42(1):38–45
Dwivedi A, Joza J, Malkani K et al (2018) Implantable Loop Recorder in Inherited Arrhythmia Diseases: A Critical Tool for Symptom Diagnosis and Advanced Risk Stratification. JACC Clin Electrophysiol 4(10):1372–1374
Bezzerides VJ, Walsh A, Martuscello M et al (2019) The Real-World Utility of the LINQ Implantable Loop Recorder in Pediatric and Adult Congenital Heart Patients. JACC Clin Electrophysiol 5(2):245–251
Lee R, Mittal S (2018) Utility and limitations of long-term monitoring of atrial fibrillation using an implantable loop recorder. Heart Rhythm 15(2):287–295
Sakhi R, Theuns D, Szili-Torok T, Yap SC (2019) Insertable cardiac monitors: current indications and devices. Expert Rev Med Devices 16(1):45–55
D’Souza R, Thomas E, Macicek S, Aziz P, Shivapour JK, Snyder C (2017) P-and R-wave amplitude sensed by reveal LINQt loop recorder in pediatric patients. J Innovat Cardiac Rhythm Manage. 8:2584
Drago F, Battipaglia I, Di Mambro C (2018) Neonatal and pediatric arrhythmias: clinical and electrocardiographic aspects. Card Electrophysiol Clin 10(2):397–412
Abbott. Confirm Rx™ ICM With SharpSense™ Technology. https://www.cardiovascular.abbott/us/en/hcp/products/cardiac-rhythm-management/confirm-rx-insertable-cardiac-monitor.html. Published 2020. Accessed.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Contributions
All authors are responsible for the contents and have read and approved the manuscript for submission.
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Yoon, J.G., Fares, M., Hoyt, W. et al. Diagnostic Accuracy and Safety of Confirm Rx™ Insertable Cardiac Monitor in Pediatric Patients. Pediatr Cardiol 42, 142–147 (2021). https://doi.org/10.1007/s00246-020-02463-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00246-020-02463-3