Abstract
We describe two balloon-assisted techniques to facilitate the advancement of large, long sheaths through freshly placed stents or tortuous anatomy. The first technique uses a balloon inflated while partially housed within the sheath to help clear an edge, thus offering resistance. The second uses an angioplasty balloon inflated in a distal pulmonary artery branch to better anchor the wire while advancing the sheath through a tortuous path. The techniques are illustrated by the cases of four patients, including two undergoing Melody valve (Medtronic, Minneapolis, MN) implantation.
Similar content being viewed by others
References
Pac FA, Kibar E, Polat TB, Topaloglu S (2010) Pneumopericardium due to long sheath during transcatheter closure of PFO. Images Paediatr Cardiol 3:1–2
Recto MR, Ing FF, Grifka RG, Nihill MR, Mullins CE (2000) A technique to prevent newly implanted stent displacement during subsequent catheter and sheath manipulation. Catheter Cardivasc Interv 49:297–300
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Below is the link to the electronic supplementary material.
Movie 1 Flouroscopic run of 12F sheath advancing past the stent edge using the bumper-balloon technique with an OPTA Pro balloon (10 mm × 2 cm)
Rights and permissions
About this article
Cite this article
Prieto, L.R., Bellotti, C.A. Balloon-Assisted Techniques for Advancing Long Sheaths Through Difficult Anatomy. Pediatr Cardiol 34, 1125–1129 (2013). https://doi.org/10.1007/s00246-012-0617-9