Abstract
Introduction
Delayed ipsilateral intraparenchymal hemorrhage (IPH) has been reported following technically successful treatment of intracranial aneurysms using flow-diverting stents in up to 8.5 % of patients. We report a similar, though less frequent phenomenon in the setting of stent-assisted coil embolization.
Methods
Institutional review board approval was obtained. A retrospective analysis of a prospective neurointerventional procedure registry was performed to review all IPHs that occurred in aneurysm patients within 90 days of endovascular treatment performed between November 2002 and November 2014 at one institution. Age, sex, hypertension, dual antiplatelet therapy, and technical details of the procedure were recorded.
Results
A total of 1697 patients underwent endovascular treatment of an intracranial aneurysm without a flow diverter at our institution during the study period. Among these, 138 patients underwent stent-assisted coiling (8.1 %). Of these, three patients (2.2 %) suffered a delayed IPH within the vascular territory distal to the treated lesion (one woman, median age 60 years).
Conclusions
Recently described in the setting of flow diversion, delayed ipsilateral IPH is not limited to flow-diverting stents. Though less frequent, a potential for this complication may exist following any intracranial stenting procedure, possibly related to hemorrhagic conversion of microembolic phenomena in the setting of dual antiplatelet or anticoagulation therapy.
Similar content being viewed by others
References
Kadkhodayan Y, Rhodes N, Blackburn S et al (2013) Comparison of enterprise with neuroform stent-assisted coiling of intracranial aneurysms. AJR Am J Roentgenol 200:872–878. doi:10.2214/AJR.12.8954
King B, Vaziri S, Singla A et al (2014) Clinical and angiographic outcomes after stent-assisted coiling of cerebral aneurysms with enterprise and neuroform stents: a comparative analysis of the literature. J Neurointerventional Surg. doi:10.1136/neurintsurg-2014-011457
Kallmes DF, Hanel R, Lopes D et al (2015) International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study. AJNR Am J Neuroradiol 36:108–115. doi:10.3174/ajnr.A4111
Becske T, Kallmes DF, Saatci I et al (2013) Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology 267:858–868. doi:10.1148/radiol.13120099
Saatci I, Yavuz K, Ozer C et al (2012) Treatment of intracranial aneurysms using the pipeline flow-diverter embolization device: a single-center experience with long-term follow-up results. AJNR Am J Neuroradiol 33:1436–1446. doi:10.3174/ajnr.A3246
Szikora I, Berentei Z, Kulcsar Z et al (2010) Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device. AJNR Am J Neuroradiol 31:1139–1147. doi:10.3174/ajnr.A2023
Lylyk P, Miranda C, Ceratto R et al (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64:632–642. doi:10.1227/01.NEU.0000339109.98070.65, discussion 642–3; quiz N6
Cruz JP, Chow M, O’Kelly C et al (2012) Delayed ipsilateral parenchymal hemorrhage following flow diversion for the treatment of anterior circulation aneurysms. AJNR Am J Neuroradiol 33:603–608. doi:10.3174/ajnr.A3065
Delgado Almandoz JE, Crandall BM, Scholz JM et al (2013) Pre-procedure P2Y12 reaction units value predicts perioperative thromboembolic and hemorrhagic complications in patients with cerebral aneurysms treated with the pipeline embolization device. J Neurointerventional Surg 5(Suppl 3):iii3–iii10. doi:10.1136/neurintsurg-2012-010582
Delgado Almandoz JE, Crandall BM, Scholz JM et al (2014) Last-recorded P2Y12 reaction units value is strongly associated with thromboembolic and hemorrhagic complications occurring up to 6 months after treatment in patients with cerebral aneurysms treated with the pipeline embolization device. AJNR Am J Neuroradiol 35:128–135. doi:10.3174/ajnr.A3621
Wang Y, Wang Y, Zhao X et al (2013) Clopidogrel with aspirin in acute minor stroke or transient ischemic attack. N Engl J Med 369:11–19. doi:10.1056/NEJMoa1215340
Diener H-C, Bogousslavsky J, Brass LM et al (2004) Aspirin and clopidogrel compared with clopidogrel alone after recent ischaemic stroke or transient ischaemic attack in high-risk patients (MATCH): randomised, double-blind, placebo-controlled trial. Lancet 364:331–337. doi:10.1016/S0140-6736(04)16721-4
Delgado Almandoz JE, Kadkhodayan Y, Crandall BM et al (2014) Variability in initial response to standard clopidogrel therapy, delayed conversion to clopidogrel hyper-response, and associated thromboembolic and hemorrhagic complications in patients undergoing endovascular treatment of unruptured cerebral aneurysms. J Neurointerventional Surg 6:767–773. doi:10.1136/neurintsurg-2013-010976
Delgado Almandoz J, Kadkhodayan Y, Fease J et al (2014) O-005 frequency of post-procedural diffusion weighted imaging abnormalities and association with delayed ipsilateral intracerebral haemorrhages in patients undergoing endovascular treatment of cerebral aneurysms with the pipeline embolization device. J Neurointerventional Surg 6(Suppl 1):A2–A3. doi:10.1136/neurintsurg-2014-011343.5
Altay T, Kang HI, Woo HH et al (2011) Thromboembolic events associated with endovascular treatment of cerebral aneurysms. J Neurointerventional Surg 3:147–150. doi:10.1136/jnis.2010.003616
Cronqvist M, Wirestam R, Ramgren B et al (2005) Diffusion and perfusion MRI in patients with ruptured and unruptured intracranial aneurysms treated by endovascular coiling: complications, procedural results, MR findings and clinical outcome. Neuroradiology 47:855–873. doi:10.1007/s00234-005-1408-2
Rordorf G, Bellon RJ, Budzik RE et al (2001) Silent thromboembolic events associated with the treatment of unruptured cerebral aneurysms by use of Guglielmi detachable coils: prospective study applying diffusion-weighted imaging. AJNR Am J Neuroradiol 22:5–10
Seo D-H, Yoon S-M, Park H-R et al (2014) Thromboembolic event detected by diffusion weighted magnetic resonance imaging after coil embolization of cerebral aneurysms. J Cerebrovasc Endovasc Neurosurg 16:175–183. doi:10.7461/jcen.2014.16.3.175
Hu YC, Deshmukh VR, Albuquerque FC et al (2014) Histopathological assessment of fatal ipsilateral intraparenchymal hemorrhages after the treatment of supraclinoid aneurysms with the pipeline embolization device. J Neurosurg 120:365–374. doi:10.3171/2013.11.JNS131599
Brinjikji W, Lanzino G, Cloft HJ et al (2015) Risk factors for hemorrhagic complications following pipeline embolization device treatment of intracranial aneurysms: results from the international retrospective study of the pipeline embolization device. AJNR Am J Neuroradiol. doi:10.3174/ajnr.A4443
Kadkhodayan Y, Delgado Almandoz J, Scholz J et al (2014) O-021 induced hyper-response to clopidogrel after elective endovascular intracranial aneurysm treatment. J Neurointerventional Surg 6(Suppl 1):A11. doi:10.1136/neurintsurg-2014-011343.21
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
We declare that all human and animal studies have been approved by the Institutional Review Board and have therefore been performed in accordance with the ethical standards laid down in the 1964 Declaration of Helsinki and its later amendments. We declare that all patients gave informed consent prior to inclusion in this study.
Conflict of interest
YK consults for Medtronic and Microvention. JEDA consults for Accriva, Medtronic, Microvention, Penumbra and Sequent.
Rights and permissions
About this article
Cite this article
Kayan, Y., Almandoz, J.E.D., Fease, J.L. et al. Incidence of delayed ipsilateral intraparenchymal hemorrhage after stent-assisted coiling of intracranial aneurysms in a high-volume single center. Neuroradiology 58, 261–266 (2016). https://doi.org/10.1007/s00234-015-1624-3
Received:
Accepted:
Published:
Issue Date: