Determination of fluorine in herbs and water samples by molecular absorption spectrometry after preconcentration on nano-TiO2 using ultrasound-assisted dispersive micro solid phase extraction

Open Access
Research Paper

Abstract

This work presents ultrasound-assisted dispersive micro solid phase extraction (USA DMSPE) for preconcentration of fluorine (F) in water and herb samples. TiO2 nanoparticles (NPs) were used as an adsorbent. The determination with slurry sampling was performed via molecular absorption of calcium monofluoride (CaF) at 606.440 nm using a high-resolution continuum source electrothermal absorption spectrometry (HR-CS ET MAS). Several factors influencing the efficiency of the preconcentration technique, such as the amount of TiO2, pH of sample solution, ultrasonication and centrifugation time and TiO2 slurry solution preparation before injection to HR-CS ET MAS, were investigated in detail. The conditions of detection step (wavelength, calcium amount, pyrolysis and molecule-forming temperatures) were also studied. After extraction, adsorbent with the analyte was mixed with 200 μL of H2O to prepare a slurry solution. The concentration limit of detection was 0.13 ng mL−1. The achieved preconcentration factor was 7. The relative standard deviations (RSDs, %) for F in real samples were 3–15%. The accuracy of this method was evaluated by analyses of certified reference materials after spiking: INCT-MPH-2 (Mixed Polish Herbs), INCT-SBF-4 (Soya Bean Flour), ERM-CAO11b (Hard Drinking Water) and TMDA-54.5 (Lake Ontario Water). The measured F contents in reference materials were in satisfactory agreement with the added amounts, and the recoveries were found to be 97–109%. Under the developed extraction conditions, the proposed method has been successfully applied for the determination of F in real water samples (lake, sea, tap water) and herbs.

Keywords

TiO2 Dispersive micro solid phase extraction Ultrasound Fluorine HR-CS ET MAS 

Introduction

Fluorine (F) in the environment, mainly as fluorides, represent about 0.06–0.09% of the earth’s crust [1]. Drinking water and food are defined as the primary sources of fluoride intake, but daily exposure depends heavily on the geographical area [2, 3]. In drinking water, it is considered as a beneficial component at levels of about 0.7 mg L−1. However, fluoride may be harmful if it exceeds 1.5 mg L−1. It is the limit set by the World Health Organization and is followed in many countries [1, 3, 4]. The most accessible form of fluoride and, hence, the most toxic is intake through drinking water [2].

In low concentrations, fluoride strengthens the tooth enamel [4]. Concentrations in the range of 1.5–4 mg L−1 result in dental fluorosis but higher values (4–10 mg L−1) may cause dental fluorosis to progress to skeletal fluorosis. Fluoride may displace hydroxide ions from hydroxyapatite, Ca5(PO4)3OH, the principal mineral constituent of teeth and bones, to form the harder and tougher fluorapatite, Ca5(PO4)3F. It makes the teeth and bones harder and more brittle [4]. It is also suggested that fluoride may combine with calcium forming fluorapatite during mineralisation resulting in a gradual reduction in the calcium content [5].

The determination of fluoride in water samples is usually performed using chromatographic methods, such as ion chromatography (IC) [6], gas chromatography (GC) [7] and high-performance liquid chromatography (HPLC) [8, 9] or electrochemical methods, such as ion selective electrodes (ISE) [10, 11], polarography [12] and voltammetry [13]. Spectroscopic methods, for example inductively coupled plasma-optical emission spectrometry (ICP-OES) or classical atomic absorption spectrometry (AAS), are not used for determination of F because of the very high ionisation potential (17.42 eV) and the resonance lines of the fluorine being below 100 nm. An alternative method for F determination is molecular absorption spectrometry (MAS) conduct with the use of high-resolution continuum source atomic absorption spectrometry (HR-CS AAS) [14, 15, 16]. This method is based on the formation of stable monofluorides of aluminium [17, 18, 19], gallium [20, 21, 22, 23, 24, 25], calcium [26, 27, 28] or strontium [29]. Diatomic molecules can absorb defined energy from a continually emitting spectral radiation source, creating molecular absorption spectra. The molecular absorption spectra correspond to the molecular transitions between the different molecule states. The molecular absorption spectrum has more lines than atomic absorption spectrum. The line width of various molecular absorption lines is roughly equal to the atomic absorption lines and can be resolved and used for analysis by HR-CS AAS.

In our previous studies, SPE in micro dispersive mode was presented using ZnO [30] as well as Ag [31] and TiO2 [32] nanoparticles for the isolation and preconcentration of Ge, Hg and Hg species, respectively. In this paper, the combination of ultrasound-assisted DMSPE (USA DMSPE) with nano-TiO2 as an adsorbent and high-resolution molecular absorption spectrometry with electrothermal vaporisation of the calcium monofluoride (HR-CS ET MAS) is presented for F determination.

To the best of our knowledge, a procedure combining these approaches has not been reported yet. Direct addition of TiO2 as an adsorbent to the sample solution provided rapid adsorption onto the material under short ultrasonication time. After extraction, the material was separated from the sample solution and injected as a slurry solution to HR-CS ET MAS. The accuracy of the method was demonstrated by analysis of spiked certified reference materials. Finally, the developed procedure was applied for extraction and determination of F in water samples and herb samples.

Materials and methods

Instrumentation

A ContrAA 700 HR-CS AAS (Analytik Jena, Germany) was used for MAS measurements. The spectrometer was equipped with a 300-W xenon short-arc lamp as a continuum radiation source for the wavelength range from 185 to 900 nm, a compact high-resolution double echelle monochromator and a charge-coupled device (CCD) array detector with a resolution of about 2 pm per pixel at 200 nm. For pyrolysis and molecule formation after extraction step, the graphite furnace technique (ET) and pyrolytically coated graphite tubes with integrated platform were employed. The operating parameters of the HR-CS ET MAS instrument during F determination after USA DMSPE are summarised in Table 1.
Table 1

Optimised experimental conditions for ultrasound-assisted dispersive micro solid phase extraction (USA DMSPE) with TiO2 as an adsorbent coupled to HR-CS ET MAS for F determination (parameters for preparation of real samples and reference materials are also presented)

Sample preparation

 Microwave-assisted digestion of solid samples

ca. 500 mg sample, 1 mL 30% H2O2, 5 mL 65% HNO3, 20 min, 300 W

 Direct analysis

Water samples

USA DMSPE with TiO2

 Sample volume (mL)

10

 Amount of TiO2 (mg)

20

 pH of sample solution

4.0

 Ultrasonication time (s)

10

 Centrifugation time (min)/rpm

3/4500

 Solution for TiO2 slurry/final volume (μL)

H2O/200

HR-CS ET MAS detection

 Wavelength (nm)

606.440

 Lamp current (A)

9

 Spectral range (pixel)

200

 Dispersion (pm pixel−1)

2

 Read time (s)

5

 Delay time (s)

0

 Measurement mode

Peak height

 Sample volume (μL)

20

 Molecule-forming element/amount (μg)

Ca/24

Furnace program steps

 Drying

80 °C, ramp 6 °C s−1, hold 20 s

 Drying

90 °C, ramp 3 °C s−1, hold 20 s

 Drying

120 °C, ramp 5 °C s−1, hold 10 s

 Pyrolysis

700 °C, ramp 300 °C s−1, hold 10 s

 Molecule formation

2450 °C, ramp 2500 °C s−1, hold 5 s

 Cleanout

2550 °C, ramp 500 °C s−1, hold 4 s

For SPE procedure, TiO2 nanoparticles were weighed using an M2P micro analytical balance (Sartorius, Gottingen, Germany) with a resolution of 1 μg (electronic weighing range up to 2 g). The pH values were measured with a pH meter (pH 211 Microprocessor, Hanna Instruments, Kehl, Germany) supplied with a glass-combined electrode. A Sonopuls HD 70 ultrasonic cell disruptor/homogeniser (70 W, 20 kHz, Bandelin, Germany) equipped with a 2-mm titanium microtip was used for dispersive extraction processes as well as for slurry preparation before MAS detection. Ultrasonic energy was fixed at 40 W.

A UniClever focused microwave sample preparation system (Plazmatronika, Wrocław, Poland) operating at 2450 MHz and 300 W maximum output was used for certified reference materials and real sample digestion before F determination. The high-pressure TFM-PTFE vessel capacity was 110 mL, and the maximum pressure and maximum temperature were 100 atm and 300 °C, respectively.

Reagents and solutions

Compressed argon of UHP 5.5 purity obtained from Air Products (Warsaw, Poland) was employed as a carrier gas without further purification. Standard solutions of fluoride were prepared from 1000 mg L−1 NaF standard solution traceable to SRM from NIST (Certipur, Merck). Ca(NO3)2·4H2O for analysis (EMSURE, Merck) was used as a calcium source and appropriate molecule-forming element. All working standard solutions were prepared daily to prevent any possible changes. The appropriate stock solution was diluted with high-purity water.

Titanium dioxide nanopowder TiO2 (≥ 99.5%, P25 Aeroxide, Degussa, Germany) with spherical particles (10–30 nm) was used as an adsorbent in dispersive micro solid phase extraction. Its average particle size was approximately 21 nm and specific surface according to BET 50 ± 15 m2 g−1 with an anatase/rutile ratio of about 80:20. The pH of the sample solutions was adjusted with 65% HNO3 and 30% NaOH (Suprapur, Merck). HNO3 (65%) and H2O2 (30%) (Suprapur, Merck) were used for digestion of the samples. High-purity water: deionised and doubly distilled water (quartz apparatus, Bi18, Heraeus, Germany) were used throughout the experiments.

Standard reference materials and real samples

The accuracy of the analytical procedure described in this work was verified using certified reference materials (CRMs). The following materials were chosen: INCT-MPH-2 (Mixed Polish Herbs) and INCT-SBF-4 (Soya Bean Flour) from the Institute of Nuclear Chemistry and Technology (Warsaw, Poland); ERM-CAO11b (Hard Drinking Water) supplied by LGC Limited (Teddington, UK); and TMDA-54.5 (Lake Ontario Water) supplied by Environment Canada (Burlington, Canada).

The certificates of the materials used do not contain certified values for the analyte. The materials were analysed before and after spiking with fluoride.

In the course of the study, the analyte was extracted and determined in three herbs samples: burdock (Arctium lappa), yarrow (Achillea millefolium) and horsetail (Equisetum). The samples were collected in the vicinity of Poznań (west-central Poland, on the Warta river), dried thoroughly, ground in an agate mortar and digested before further extraction procedure.

The water samples (lake, tap sample no.1 and well water) were collected in Nysa, a town in southwestern part of Poland on the Nysa Kłodzka River. The lake samples (nos. 1–4) were collected from different sampling points situated on the Nyskie Lake. The place of sampling was chosen not accidentally. The Neogene aquifer has been the main source of drinking water for Nysa and the surrounding area for years. However, due to the high fluoride content, the water supplies abstracting the Neogene aquifer water were closed. High fluoride content in water from the aquifer has caused fluorosis disease in the population of Nysa. Fluoride in water from this region originates from a long-time contact of the water with Precambrian and Palaeozoic igneous rocks composed of fluorine-rich minerals. For comparison, the samples of sea water (Baltic, Adriatic and Tyrrhenian) and tap water (sample no. 2) from the laboratory (Poznań University of Technology) were analysed.

Analytical procedures

Microwave-assisted digestion of solid samples before extraction step

Approximately 500 mg (herbs, soya flour) of powdered sample was placed in the TFM-PTFE vessel of the microwave digestion system and moistened by 1 mL of 30% H2O2. Then, 5 mL of 65% HNO3 was added. The samples were heated for 20 min at 300 W.

Then, the clear digested solution was transferred into 10 mL calibrated flask and diluted to volume with high-purity water. Before further analysis, this solution was appropriately diluted depending on the concentration level of the analyte. In all cases, a corresponding blank was also prepared according to the above microwave-assisted digestion procedure. After this procedure, it was possible to conduct the extraction procedure.

Ultrasound-assisted dispersive micro solid phase extraction

Ten millilitres of a sample solution was poured into a centrifuge tube. The pH of the sample was adjusted to 4.0. Then, 20 mg of TiO2 (as 200 μL of 10% m/v suspension) was added. The suspension (4 mL containing 400 mg of TiO2) was prepared before analysis with the use of ultrasound (2-mm titanium microtip, 40 W, 20 kHz, for 40 s). Subsequently, the sample solution with nanomaterial was ultrasonicated for 10 s. Homogenisation was achieved that promoted the interaction between F and the nanoparticles. After that, the mixture was centrifuged for 3 min at 4500 rpm. The analyte adsorbed on TiO2 settled on a bottom of the tube. A water phase was removed and solid phase was mixed with 200 μL of H2O and treated with ultrasound for 5 s (ultrasonic probe; 2-mm titanium microtip, 40 W, 20 kHz) to achieve a slurry solution. In order to determine F in the TiO2 slurry solution, 20 μL of the solution was injected for HR-CS ET MAS determination under the optimised conditions. Optimised experimental conditions for USA DMSPE with TiO2 as an adsorbent coupled to HR-CS ET MAS for fluorine determination are presented in Table 1.

Fluorine determination using HR-CS ET MAS

The graphite furnace temperature program used for the determination of F via their diatomic molecules CaF after USA DMSPE is given in Table 1. For analysis, 20 μL of the sample followed by 2 μL of a calcium nitrate (Ca(NO3)2·4H2O) solution at a concentration of 12 g L−1 was directly injected onto the platform of the graphite tube. Calibration was performed by the standard calibration technique using standards after ultrasound-assisted dispersive micro solid phase extraction.

Results and discussion

Optimisation of HR-CS ET MAS detection

Selection of wavelength

The formation of a competitive molecule other than the desirable molecule of the analyte is one of the potential sources of error in MAS. It reduces the concentration of the desirable molecule in the gas phase and affects the sensitivity. Due to the relatively high concentrations of calcium in the water and food samples, we decided to investigate the possibility of using CaF molecular absorption for analytical purposes. According to the literature data this diatomic molecule is stable (its bond dissociation energy is around 529 kJ mol−1) and the most intensive band head is found at 606.440 nm, which is part of the X2Σ+-A2Π electronic transition [33, 34]. The molecular absorption band is in the visible range of the spectrum where only very few atoms have absorption lines. It minimises the possible spectral interferences. However, there is the possible interference caused by Cl. A broadband of CaCl formed in the presence of Cl overlapped with the CaF line at 606.440 nm [35]. In our study, no significant interferences were observed. It can be due to the effective separation of the analyte from the matrix during preconcentration step. Additionally, calcium used as the molecule-forming reagent acts also as a chemical modifier, so that no other reagent had to be added.

Optimisation of calcium amount

There are many reports in the literature on the use of several elements (Al [17, 18, 19], Ga [20, 21, 22, 23, 24, 25], Ca [26, 27, 28] or Sr [29]) to formation of fluoride diatoms. In general, the reaction with Ga allows to achieve the lowest detection limits. However, this procedure requires the use of additional permanent modifiers of a graphite tube. Coating with zirconium, magnesium, palladium, ruthenium or tungsten is usually used [21, 22, 23, 24, 25]. It extends the time of analysis and increases the amount of reagents needed. At the same time, a choice of reagent should be connected with the composition of the sample matrix. Some authors have even investigated the use of Ca originally present in food samples to carry out this reaction [27]. Additionally, the works already published indicate that CaF does not require the use of additional modifiers of a graphite tube as it is with GaF [27]. This is why, for food and water samples, we decided to choose Ca as a molecule-forming element.

The efficiency of formation of CaF highly depends on the amount of calcium which should be high enough to provide constant and maximum sensitivity for F concentrations in all samples. Therefore, the influence of calcium amount on the analytical signal was investigated within the range of 2.4–96 μg. In this study, 2 μL of a calcium nitrate (Ca(NO3)2·4H2O) solution at a concentration of 12 g L−1 was used. The absorbance increased with an amount of calcium in the range of 2.4–24 μg (Fig. 1). For an amount of calcium equal to 24 μg, the absorbance reached its maximum and was found to have a plateau in the range 24–96 μg. Therefore, in this study, the quantity of 24 μg of calcium was chosen for further experiments. The molar ratio between calcium and fluorine ([Ca]:[F]) was found to be about 11 for 20 μL of standard solution at a concentration of 50 ng mL−1.
Fig. 1

The effect of Ca amount on the CaF molecular absorption. Conditions: standard solution of F (50 ng mL−1) analysed after USA DMSPE, pH = 3.5, sample volume 10 mL, ultrasonication time 10 s, centrifugation 4 min. The error bar is the standard deviation (SD, n = 3)

Optimisation of pyrolysis and molecule-forming temperature

The temperature program was optimised for a standard solution containing 50 ng mL−1 of fluoride after preconcentration on TiO2. Three drying steps were necessary in order to avoid spattering of the sample and to obtain a uniform solid deposit on the graphite surface. The effect of pyrolysis temperature on absorbance was studied within the range 600–900 °C (Fig. 2). No significant differences between analytical signals were observed. However, the absorbance obtained for 700 °C was relatively the highest and this value of the pyrolysis temperature was chosen for further experiments. After optimisation of pyrolysis conditions, the effect of molecule-forming temperature on analytical signals was studied within the range 2200–2650 °C (Fig. 2). Maximum absorbance was achieved at a temperature of 2450 °C. Complete vaporisation and CaF molecule formation were achieved in the graphite tube with minimal influence exerted by the matrix. The temperature program used for F determination in certified reference materials and real samples is shown in Table 1.
Fig. 2

Pyrolysis and molecule formation curves for CaF. Conditions: standard solution of F (50 ng mL−1) analysed after USA DMSPE, for pyrolysis runs Tmf 2400 °C, for molecule formation runs Tp 700 °C. The error bar is the standard deviation (SD, n = 3)

Some elements (e.g. vanadium, titanium, molybdenum) form stable carbides in the presence of graphite at high temperatures. Carbide formation can inhibit the decomposition process of the graphite tube and affect the analytical reproducibility and sensitivity or give a significant memory effect. In the course of the study, the pyrolytically coated graphite tube was used. It made the atomizer relatively unreactive chemically and reduced the tendency for carbides formation. Relative standard deviation (RSD) values recorded during the study were found to be satisfactory; therefore, the effect of reaction of C and TiO2 was considered to be neglected.

Optimisation of TiO2 USA DMSPE conditions

The parameters that affect the extraction efficiency, including the amount of TiO2, the sample pH, homogenisation (ultrasonication) and centrifugation time as well as a slurry preparation step before injection to HR-CS ET MAS were investigated. Initial conditions with the analyte concentration of 50 ng F mL−1 and sample volume of 10 mL were used. Regarding the amount of applied TiO2, 4 mL of a suspension containing 400 mg was prepared. Of the suspension added to the sample solution, 200 μL corresponded to 20 mg of TiO2 per sample. All the experiments were performed in triplicate (n = 3). Due to the construction of the centrifuge and vials, it was not possible to apply higher sample volume.

Particle amount

To achieve an optimum amount of TiO2 necessary for quantitative enrichment of the analyte, the introduced amount of nanoparticles was varied from 5 to 40 mg per sample (10 mL). It was found that effective adsorption of F (the highest analytical signal) was achieved with amounts of nanomaterial from 20 mg (Fig. 3). As observed, a significant decrease in the signal after exceeding the value of 20 mg of adsorbent is difficult to interpret. It can be caused by too high concentration of a suspension injecting to the furnace and therefore interferences in formation of diatomic molecules CaF. For further application, the amount of TiO2 equal to 20 mg was used (that corresponding to 200 μL of 10% TiO2 slurry solution added to 10 mL of the sample).
Fig. 3

The effect of amount of TiO2 on the determination of F with the use of USA DMSPE procedure. Conditions: standard solution of F (50 ng mL−1), pH = 3.5, sample volume 10 mL, ultrasonication time 10 s, centrifugation 4 min. The error bar is the standard deviation (SD, n = 3)

TiO2 nanoparticles with 80% of anatase were chosen due to the possible mechanism of interaction with the analyte and possibility of effective adsorption connected with high specific surface area. The adsorbent is known for its physical and chemical stability in acidic and alkaline solutions, low cost, nontoxicity and fast rate of adsorption and desorption [36, 37].

Chemical forms of the adsorbent and the analyte, as well as the nature of the interactions, depend strongly on the pH of the sample and are discussed in section “Sample pH”.

Sample pH

The pH value of the sample is one of the most important factors affecting the extraction efficiency. The effect of pH on the adsorption of F on nano-TiO2 was studied by varying the pH of the aqueous solutions in the range of 2–11. To modify the pH, appropriate amounts of NaOH or HNO3 were added to sample solution. The results of the pH optimisation are shown in Fig. 4. It was observed that when the pH was low (1.0–4.0), the adsorption of the analyte (absorbance values) increased. For pH = 4, the absorbance of the analyte reached its maximum value. However, with further increases in the pH (above 5.5), the absorbance decreased significantly for F. Therefore, in this study, a pH of 4.0 was chosen for further experiments.
Fig. 4

The effect of sample pH on the determination of F with the use of USA DMSPE procedure. Conditions: standard solution of F (50 ng mL−1), sample volume 10 mL, 20 mg of TiO2, ultrasonication time 10 s, centrifugation 4 min. The error bar is the standard deviation (SD, n = 3)

The adsorption mechanism of the analyte on nano-TiO2 depends on the type of the active sites on the adsorbent surface and the concentrations of produced sorbate species at the measured pH value [38]. The point of zero charge (PZC) of the anatase nanoparticles has been found at pH ca. 6.2 [36, 39, 40]. Below this value, the surface of anatase nanoparticles consisted of the mixture of positively charged TiOH2 + and neutral species of TiOH0. The neutral species of TiOH0 and the negatively charged TiO occur in the solution at pH above PZC (basic pH) [41, 42]. It was found that at pH 4.0, the surface of adsorbent is mostly covered with positively charged TiOH2 + [38].

Fluorine in an aqueous solution has different species, including F, HF and HF2 . HF is the dominant species at pH 1–3.18, while F is the dominant species at pH above 3.18. An increase in HF2 species is observed as total fluoride concentration increases. However, when the total dissolved fluoride is below 50 mg L−1, the maximum fraction of HF2 is less than 10% [43, 44]. As a result, there are only two soluble fluoride species considered, HF and F.

According to the theory, the adsorption of anions on titanium dioxide proceeds when the pH of the solution is lower than the point PZC of the oxide (ca. 6.2 of pH) [39, 45]. On the base of literature, the spontaneous adsorption of H+ on titania results in a positive charge on the anatase surface and thus the latter becomes active to the adsorption of fluoride [39, 46]. Additionally, pH < 5 is the most favourable for fluoride uptake due to the lower concentration of OH ions, which compete with fluoride for sorption sites with increasing concentration of OH in the solution and reduce the capacity for sorption of F. Speciation of the investigated anion is the second reason for lower fluoride sorption at pH < 3.18 when it exists as HF. Maximal sorption capacity is observed at pH 4, at which the predominant form of the anion is F. Thus, the adsorption of the analyte can be described as interaction of positively charged TiOH2 + with F [44].

Ultrasonication and centrifugation time

The positive effect of dispersion on the active area of a solid sorbent used in DMSPE procedure is well known [31, 32, 47]. The dispersion can be chemically achieved (usually using an organic solvent) or assisted by an ultrasound energy. The influence of the external energy source allows for effective reduction of the particles size of the sorbent. This results in a significant increase in the area-to-volume ratio [31, 32, 47].

To investigate the effect of ultrasonication on the extraction process, the dispersion was performed at 40 W ultrasonic power in continuous mode and was evaluated in the range of 5–30 s. An ultrasonic probe (2-mm diameter) was used and the study was conducted for 10 mL of sample solution containing adsorbent and the analyte. During the ultrasonication step, the solutions became milky by dispersing the nanomaterial in the volume of sample. In the first 10 s of extraction, the absorbance increased then after exceeding this time remained approximately constant. Therefore, for further study, an ultrasound-assisted dispersion step of 10 s was chosen.

After the extraction step, the sample was centrifuged to separate the liquid and solid phases.

To investigate the absorbance as a function of centrifugation time, the experiments were performed in the range of 1–6 min (at the rate of 4500 rpm). The volume of sedimented phase (as well as absorbance) increased with increasing time of centrifugation.

A complete settling of the solid phase to the bottom of the centrifuge tube was observed for 3 min. Above this time, the absorbance value was constant. Finally, during the extraction, the sample was centrifuged for 3 min. Thus, 3 min was selected as the final centrifugation time (at the rate of 4500 rpm).

TiO2 slurry solution preparation

In this research, the injection of slurry samples into the atomizer during HR-CS ET MAS analysis was performed. Therefore, after extraction step, the TiO2 slurry containing the analyte was prepared before the analysis. The volume of H2O was varied in the range of 150–1000 μL. For a volume of H2O, less than 150 μL obtained precision (expressed as RSD) was not satisfactory (above 15%). On the other hand, a decrease of analytical signals was observed for volumes above 250 μL. It was caused by too high dilution of the samples. Therefore, in this study, H2O volume of 200 μL was chosen as a solution for slurry preparation before analysis.

For slurry preparation, the water phase was removed after centrifugation. Then, the solid phase was mixed with 200 μL of H2O and ultrasonicated for 5 s (2-mm titanium microtip, 40 W, 20 kHz) to achieve a stable slurry solution which was injected to the detector (20 μL) prior to HR-CS ET MAS analysis.

Analytical figures of merit

The analytical parameters characterising the proposed USA DMSPE HR-CS ET MAS method were investigated under the optimised experimental conditions (Table 1). Seven replicate measurements of the total procedure for standard solution at a concentration of 50 ng mL−1 were carried out, and the RSD was calculated. The RSD achieved for F was 7%. The detection limit (LOD) achieved for 10 mL of real sample solution was 0.13 ng mL−1. This value was calculated as the concentration of the analyte yielding a signal equivalent to three times the standard deviation of the blank value (n = 5). Calibration was performed by the standard calibration technique using standards after USA DMSPE extraction. Received correlation coefficient (0.9961) was acceptable. The preconcentration factor achieved for F was 7. This factor was calculated using the ratio of the analyte concentration in the slurry solution containing solid phase with the analyte to the initial concentration of the analyte within the sample solution [32, 48].

Accuracy verification

The accuracy of the analytical procedure described in this work was verified using CRMs. The following materials were chosen: INCT-MPH-2 (Mixed Polish Herbs), INCT-SBF-4 (Soya Bean Flour), ERM-CAO11b (Hard Drinking Water) and TMDA-54.5 (Lake Ontario Water). The certificates of the materials used do not contain certified values for the analyte. The materials were analysed before and after spiking with fluoride (4.0 mg kg−1, calculated on the mass of solid material and 0.150 or 0.100 mg L−1 for water matrixes). After spiking, the solid CRMs were digested before further extraction procedure. The waters were spiking directly before extraction step. The recoveries were between 97 and 109% that is considered satisfactory for the low concentration levels of the analyte (Table 2).
Table 2

Determination of F in certified reference materials (CRMs) (before and after spiking) by USA DMSPE coupled to HR-CS ET MAS using the optimised parameters

CRM

Determined

Added

Value of t test

Significanced

Total F concentration (mg kg−1)

Added

Recovery (%)

RSD (%)

F concentration (mg kg−1)

INCT-MPH-2a

Mixed Polish Herbs

13.0 ± 1.4

11

INCT-MPH-2b

Mixed Polish Herbs

17.1 ± 1.2

4.1 ± 0.3

105

7

4.0

0.577

NS

INCT-SBF-4a

Soya Bean Flour

10.2 ± 0.8

8

INCT-SBF-4b

Soya Bean Flour

14.6 ± 0.9

4.4 ± 0.3

109

6

4.0

2.309

NS

ERM-CAO11ba

Hard Drinking Water

0.809c ± 0.097

12

ERM-CAO11bb

Hard Drinking Water

0.954c ± 0.057

0.145c ± 0.009

97

6

0.150c

0.962

NS

TMDA-54.5a

Lake Ontario Water

0.269c ± 0.030

11

TMDA-54.5b

Lake Ontario Water

0.375c ± 0.019

0.106c ± 0.005

107

5

0.100c

2.078

NS

Recovery and relative standard deviation (RSD) values also are shown. Obtained values: average value ± standard deviation (n = 3)

NS not significant

aAnalysed before spiking

bAnalysed after spiking

cMilligrams per litre

dSignificance of t test (n = 3) at 95% confidence level, t critical = 4.303

The certified reference materials (hard and lake waters) used in the research are characterised by a matrix containing relatively high concentrations (μg L−1) of the elements (e.g. Al, Ba, Cr, Cu, Fe, Pb, Sr, Zn), sometimes in the range of milligrams per litre (Ca, K, Mg, Na). The solid materials (herbs and soya matrices) contain also high values of Al, Ca, Cl, Cu, K, Mg, Mn, Na, Ni, Pb, Sr and Zn in the range of milligrams per kilograms or even percent. The obtained recoveries of the analyte, in the presence of these elements, proved that the interferences from foreign ions can be ignored. Due to the satisfactory results, the additional study in terms of matrix influence was not carried out.

Obtained results show that the proposed method can be applied for the preconcentration and determination of F in the samples with complex matrices. The precision expressed as RSD (%) was in the range of 5–7% for the certified reference materials after spiking.

Fluorine determination in real samples

In order to evaluate the usefulness of the proposed method for determination of F, contents in four water samples (lake, tap, well and sea waters) as well as in herb samples were analysed using previously optimised experimental conditions. First, the water samples were collected in 500-mL Schott Duran glass bottles. Then, they were acidified with HNO3 (0.5 mol L−1) and stored in the dark at 4 °C. A Cameo syringe filter with a polytetrafluoroethylene membrane and a pore size of about 0.22 mm (GE Water & Process Technologies, USA) was used for sample filtration. Before further extraction procedure, herb samples were digested. Digestion procedure was described in section “Microwave-assisted digestion of solid samples before extraction step”. In section “Analytical procedures”, the analytical procedure used for fluorine preconcentration and determination was described. The RSDs (%) achieved for real samples were 3–15% (Table 3).
Table 3

Determination of fluorine in real samples by USA DMSPE coupled to HR-CS ET MAS using the optimised parameters

Samplea

Fluorine concentration (mg L−1)

Lake water no. 1

0.118 ± 0.007

Lake water no. 2

0.110 ± 0.008

Lake water no. 3

0.132 ± 0.004

Lake water no. 4

0.122 ± 0.006

Tap water no. 1

0.156 ± 0.012

Well water

0.239 ± 0.017

Tap water no. 2

0.173 ± 0.019

Sea water (Baltic)

0.153 ± 0.023

Sea water (Adriatic)

0.319 ± 0.016

Sea water (Tyrrhenian)

0.146 ± 0.007

Burdock

4.0b ± 0.5

Yarrow

8.2b ± 0.5

Horsetail

3.4b ± 0.4

Obtained values: average value ± standard deviation (n = 3)

aDescription of the samples in section “Standard reference materials and real samples

bMilligrams per kilogram

As it was already mentioned, drinking water and food are the primary sources of fluoride intake but exposure mainly depends on the geographical origin of a product [2, 3]. According to the WHO limit, fluoride from water may be harmful if it exceeds 1.5 mg L−1 [1, 3, 4].

Gomez et al. [17] found the concentration of fluoride in drinking waters in the range 0.05–1.10 mg L−1 and in sea waters 1.30–2.50 mg L−1. Ozbek et al. [29] determined the fluoride concentration in bottled drinking waters in the range 0.12–0.25 and in the tap waters 0.97–0.04 mg L−1. In the course of the study, the established average concentrations of fluoride in analysed waters were 0.121 and 0.206 mg L−1, for lake and seas, respectively. Tap waters contained 0.156 and 0.173 mg L−1 and it was below the level announced by WHO (1.5 mg L−1) (Table 3). Samples collected from areas (Nysa) where soil contains the high fluoride content (lake water nos. 1–4, tap water no.1) were not characterised by higher concentrations of the analyte.

The levels of F in food items depend mainly on the fluoride contents in the soil and water used for irrigation and on the place of origin [2]. Vegetables and fruits contribute little to exposure as they contain low concentrations of fluoride (e.g. 0.1–0.4 mg kg−1). Cow’s milk typically contains also low levels of fluoride, e.g. 0.02–0.05 mg L−1. However, tea leaves may contain high levels of F (event up to 400 mg kg−1 dry weight) [1]. Unfortunately, there is a lack of scientific reports on the concentration of F in herbs intended for the preparation of infusions. In herb samples analysed by USA DMSPE coupled to HR-CS ET MAS, detected average concentration of fluorine was 5.2 mg kg−1 (Table 3).

Fluorine is less determined in food samples because of its relatively low concentrations and thus the requirement for low limits of detection. Therefore, it is difficult to compare the results (for samples as for CRMs before spiking) with literature. Additionally, there is no information about EU legislation setting the maximum concentrations of this element in food products.

Comparison with other methods

In analytical chemistry, for determination of fluoride, solid phase extraction (SPE) methods have been proposed with different adsorbents. For this purpose, SPE columns with charcoal [8], nanometre-size zirconia (ZrO2) [49] and mini-column with polymer-zirconium complex (Zr(IV) with a hydrophilic with an iminodiacetic acid group) [50] were investigated. As an effective adsorbent, magnetic iron oxide nanoparticles (MIONs) [51] were used as well as headspace solid phase microextraction (HS-SPME) with carboxen-polydimethylsiloxane fibre (CAR/PDMS)) [52, 53].

The LOD achieved for F (0.13 ng mL−1) is comparable with the LOD obtained using H2SO4-activated nanometre-size zirconia as a solid sorbent prior to determination of fluoride by ion chromatography [49]. The LOD obtained for F is also superior by about two orders of magnitude to the LODs achieved using magnetic iron oxide nanoparticles prior to spectrophotometric determination of the analyte [51] and HS-SPME with carboxen-polydimethylsiloxane fibre prior to determination by gas chromatography-tandem mass spectrometry [53]. The LOD reported by Wejnerowska et al. [52] is about 50 times worse than that achieved for F when using nano-TiO2 as a solid sorbent. The LOD obtained using SPE columns with charcoal for fluoride preconcentration prior to reversed-phase high-performance liquid chromatography (RP HPLC) determination is superior by a factor of 7 to the LOD presented in this study [8].

Additionally, the obtained detection limit was compared with those achieved without the enrichment step, only using MAS detection (Table 4). Unquestionably, the preconcentration procedure conducted by the USA DMSPE allows a significant improvement in the detection limit using MAS technique.
Table 4

Comparison of the proposed method with other reported in the literature

Detection technique

Form of fluoride

LOD (μg L−1)

Ref.

HR-CS F MAS

AlF

5500

16

ETAAS

AlF

20

17

Solid sampling ET MAS

AlF

0.17a

19

HR-CS ET MAS

SrF

36

29

HR-CS F MAS

GaF

1000

20

HR-CS ET MAS

GaF

0.9

21

HR-CS ET MAS

GaF

0.26

22

HR-CS ET MAS

GaF

0.26

23

HPLC-HR-CS ET MAS

GaF

4b

24

HR-CS ET MAS

GaF

0.58

25

HR-CS ET MAS

CaF

160

26

HR-CS ET MAS

CaF

26

27

Slurry sampling ET MAS

CaF

2.2a

28

HR-CS ET MAS

CaF

380

33

Solid sampling ET MAS

CaF

0.36a

34

Slurry sampling ET MAS

CaF

5a

35

USA DMSPE HR-CS ET MAS

CaF

0.13

This work

aMicrograms per gram

bPicogram

Conclusions

The combination of ultrasound-assisted dispersive micro solid phase extraction, with the use of TiO2 as an adsorbent and high-resolution molecular absorption spectrometry with electrothermal vaporisation of the calcium monofluoride for preconcentration and determination of F, was presented. The use of nano-TiO2 as a solid sorbent enabled to achieve low detection limit and relatively good preconcentration factor. USA DMSPE is effective and takes place in a relatively short time of about 4 min. The use of ultrasonication to make the solid phase well dispersed allowed to reduce the time of the whole procedure. The extraction procedure does not require the use of organic solvents and detection step can be performed with the use of calibration against aqueous standards (after USA DMSPE). Additionally, the procedure enables to extend the analytical capabilities of the HR-CS ET MAS technique. This is the first published combination of the enrichment step using USA DMSPE with fluorine determination by ET MAS. The procedure makes it possible to achieve lower limits of determination compared to other methods based only on ET MAS detection. Proposed methodology makes ET MAS more competitive compared to other published techniques for fluorine determination. Thus, it can be an alternative to them. The method was successfully applied to F determination in natural waters and herb samples without any significant matrix influence.

Notes

Acknowledgements

This work was supported by the 03/31/DSPB/0340 grant from the Polish Ministry of Science and Higher Education.

The authors acknowledge the guidance provided by Mr. Marek Rymarz during the sampling procedure.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflicts of interest.

References

  1. 1.
    Fawell J, Bailey K, Chilton J, Dahi E, Fewtrell L, Magara Y. Fluoride in drinking-water. World Health Organization (WHO) London: IWA Publishing; 2006.Google Scholar
  2. 2.
    Jagtap S, Yenkie MK, Labhsetwar N, Rayalu S. Fluoride in drinking water and defluoridation of water. Chem Rev. 2012;112:2454–66.CrossRefGoogle Scholar
  3. 3.
    Guidelines for drinking-water quality. World Health Organization. 2006. http://www.who.int/water_sanitation_health/dwq/gdwq0506.pdf. Accessed 05 Apr 2017.
  4. 4.
    Mohapatra M, Anand S, Mishra BK, Giles DE, Singh P. Review of fluoride removal from drinking water. J Environ Manag. 2009;91:67–77.CrossRefGoogle Scholar
  5. 5.
    Ayoob S, Gupta AK. Fluoride in drinking water: a review on the status and stress effects. Crit Rev Environ Sci Technol. 2006;36:433–87.CrossRefGoogle Scholar
  6. 6.
    Kapinus EN, Revelsky IA, Ulogov VO, Lyalikov YA. Simultaneous determination of fluoride, chloride, nitrite, bromide, nitrate, phosphate and sulfate in aqueous solutions at 10-9 to 10-8% level by ion chromatography. J Chromatogr B. 2004;800:321–3.CrossRefGoogle Scholar
  7. 7.
    Pagliano E, Meija J, Ding J, Sturgeon RE, D’Ulivo A, Mester Z. Novel ethyl-derivatization approach for the determination of fluoride by headspace gas chromatography/mass spectrometry. Anal Chem. 2012;85:877–81.CrossRefGoogle Scholar
  8. 8.
    Musijowski J, Szostek B, Koc M, Trojanowicz M. Determination of fluoride as fluorosilane derivative using reversed-phase HPLC with UV detection for determination of total organic fluorine. J Sep Sci. 2010;33:2636–44.CrossRefGoogle Scholar
  9. 9.
    Xu XR, Li HB, Gu JD, Paeng KJ. Determination of fluoride in water by reversed-phase high-performance liquid chromatography using F-La3+-alizarin complexone ternary complex. Chromatographia. 2004;59:745–7.CrossRefGoogle Scholar
  10. 10.
    Lupo M, Fina BL, Aguirre MC, Armendariz M, Rigalli A. Determination of water fluoride concentration and the influence of the geographic coordinate system and time. Water Air Soil Pollut. 2012;223:5221–5.CrossRefGoogle Scholar
  11. 11.
    El Jaoudi R, Mamouch F, Ait El Cadi M, Bousliman Y, Cherrah Y, Bouklouze A. Determination of fluoride in tap water in Morocco using a direct electrochemical method. Bull Environ Contam Toxicol. 2012;89:390–4.CrossRefGoogle Scholar
  12. 12.
    Guanghan L, Qiongling W, Xiaogang W, Tong Z, Xin Y. Polarographic determination of trace fluoride in foods. Food Chem. 1999;66:519–23.CrossRefGoogle Scholar
  13. 13.
    Mao Y, Chen Y, Chu L, Zhang X. Electrodeposited apatite coating for solid-phase microextraction and sensitive indirect voltammetric determination of fluoride ions. Talanta. 2013;115:500–5.CrossRefGoogle Scholar
  14. 14.
    Welz B, Becker-Ross H, Florek S, Heitmann U. High-resolution continuum source AAS. The better way to do atomic absorption spectrometry. Weinheim: Wiley-VCH Verlag; 2005.CrossRefGoogle Scholar
  15. 15.
    Welz B, Lepri FG, Araujo RGO, Ferreira SLC, Huang MD, Okruss M, et al. Determination of phosphorus, sulfur and the halogens using high-temperature molecular absorption spectrometry in flames and furnaces—a review. Anal Chim Acta. 2009;647:137–48.CrossRefGoogle Scholar
  16. 16.
    Ozbek N, Akman S. Method development for the determination of fluorine in toothpaste via molecular absorption of aluminium mono fluoride using a high-resolution continuum source nitrous oxide/acetylene flame atomic spectrophotometer. Talanta. 2012;94:246–50.CrossRefGoogle Scholar
  17. 17.
    Gomez Gomez M, Palacios Corvilio MA, Cámara RC. Determination of fluoride in drinking water and sea water by aluminium monofluoride molecular absorption spectrometry using an electrothermal graphite furnace. Analyst. 1988;113:1109–12.CrossRefGoogle Scholar
  18. 18.
    Gómez M, Rodríguez I, Cámara C, Palacios MA. Evaluation of biological sample mineralisation determination of fluorine by graphite furnace absorption spectrometry. Analyst. 1990;115:553–7.CrossRefGoogle Scholar
  19. 19.
    Flores ELM, Barin JS, Flores EMM, Dressler VL. A new approach for fluorine determination by solid sampling graphite furnace molecular absorption spectrometry. Spectrochim Acta B. 2007;62:918–23.CrossRefGoogle Scholar
  20. 20.
    Huang MD, Becker-Ross H, Florek S, Heitmann U, Okruss M. Determination of halogens via molecules in the air-acetylene flame using high-resolution continuum source absorption spectrometry: part I. Fluorine. Spectrochim Acta B. 2006;61:572–8.CrossRefGoogle Scholar
  21. 21.
    Heitmann U, Becker-Ross H, Florek S, Huang MD, Okruss M. Determination of non-metals via molecular absorption using high resolution continuum source absorption spectrometry and graphite furnace atomization. J Anal At Spectrom. 2006;21:314–1320.CrossRefGoogle Scholar
  22. 22.
    Gleisner H, Welz B, Einax JW. Optimization of fluorine determination via the molecular absorption of gallium mono-fluoride in a graphite furnace using a high-resolution continuum source spectrometer. Spectrochim Acta B. 2010;65:864–9.CrossRefGoogle Scholar
  23. 23.
    Gleisner H, Einax JW, Morés S, Welz B, Carasek E. A fast and accurate method for the determination of total and soluble fluorine in toothpaste using high-resolution graphite furnace molecular absorption spectrometry and its comparison with established techniques. J Pharm Biomed Anal. 2011;54:1040–6.CrossRefGoogle Scholar
  24. 24.
    Qin Z, McNee D, Gleisner H, Raab A, Kyeremeh K, Jaspars M, et al. Fluorine speciation analysis using reverse phase liquid chromatography coupled off-line to continuum source molecular absorption spectrometry (CS-MAS): identification and quantification of novel fluorinated organic compounds in environmental and biological samples. Anal Chem. 2012;84:6213–9.CrossRefGoogle Scholar
  25. 25.
    Würtenberger I, Gus R. A highly sensitive method for in vitro testing of fluorinated drug candidates using high-resolution continuum source molecular absorption spectrometry (HR-CS MAS). Anal Bioanal Chem. 2014;406:3431–42.CrossRefGoogle Scholar
  26. 26.
    Morés S, Monteiro GC, da Silva SF, Carasek E, Welz B. Determination of fluorine in tea using high-resolution molecular absorption spectrometry with electrothermal vaporization of the calcium mono-fluoride CaF. Talanta. 2011;85:2681–5.CrossRefGoogle Scholar
  27. 27.
    Ozbek N, Akman S. Determination of fluorine in milk samples via calcium-monofluoride by electrothermal molecular absorption spectrometry. Food Chem. 2013;138:650–4.CrossRefGoogle Scholar
  28. 28.
    Ozbek N, Akman S. Optimization and application of a slurry sampling method for the determination of total fluorine in flour using a high-resolution continuum source graphite furnace molecular absorption spectrometer. Food Anal Methods. 2016;9:2925–32.CrossRefGoogle Scholar
  29. 29.
    Ozbek N, Akman S. Method development for the determination of fluorine in water samples via the molecular absorption of strontium monofluoride formed in an electrothermal atomizer. Spectrochim Acta B. 2012;69:32–7.CrossRefGoogle Scholar
  30. 30.
    Stanisz E, Krawczyk-Coda M. ZnO nanoparticles as an adsorbent in ultrasound-assisted dispersive micro solid-phase extraction combined with high-resolution continuum source electrothermal atomic absorption spectrometry for determination of trace germanium in food samples. Microchem J. 2017;132:136–43.CrossRefGoogle Scholar
  31. 31.
    Krawczyk M, Stanisz E. Silver nanoparticles as a solid sorbent in ultrasound-assisted dispersive micro solid-phase extraction for the atomic absorption spectrometric determination of mercury in water samples. J Anal At Spectrom. 2015;30:2353–8.CrossRefGoogle Scholar
  32. 32.
    Krawczyk M, Stanisz E. Ultrasound-assisted dispersive micro solid-phase extraction with nano-TiO2 as adsorbent for the determination of mercury species. Talanta. 2016;161:384–91.CrossRefGoogle Scholar
  33. 33.
    Ozbek N, Akman S. Determination of fluorine in Turkish wines by molecular absorbance of CaF using a high resolution continuum source atomic absorption spectrometer. LWT Food Sci Technol. 2015;61:112–6.CrossRefGoogle Scholar
  34. 34.
    Borges AR, François LL, Welz B, Carasek E, Vale MGR. Determination of fluorine in plant materials via calcium mono-fluoride using high-resolution graphite furnace molecular absorption spectrometry with direct solid sample introduction. J Anal At Spectrom. 2014;29:1564–9.CrossRefGoogle Scholar
  35. 35.
    Huang MD, Becker-Ross H, Okruss M, Geisler S, Florek S, Richter S, et al. Direct determination of fluorine in niobium oxide using slurry sampling electrothermal high-resolution continuum source molecular absorption spectrometry. Spectrochim Acta B. 2014;94-95:34–8.CrossRefGoogle Scholar
  36. 36.
    Vu D, Li Z, Zhang H, Wang W, Wang Z, Xu X, et al. Adsorption of Cu(II) from aqueous solution by anatase mesoporous TiO2 nanofibers prepared via electrospinning. J Colloid Interface Sci. 2012;367:429–35.CrossRefGoogle Scholar
  37. 37.
    Jin Z, Gao H, Hu L. Removal of Pb(II) by nano-titanium oxide investigated by batch, XPS and model techniques. RSC Adv. 2015;5:88520–8.CrossRefGoogle Scholar
  38. 38.
    Kocabaş-Ataklı ZO, Yürüm Y. Synthesis and characterization of anatase nanoadsorbent and application in removal of lead, copper and arsenic from water. Chem Eng J. 2013;225:625–35.CrossRefGoogle Scholar
  39. 39.
    Zhang L, Ishi D, Shitou K, Morita Y, Isozaki A. Simultaneous multi-element analysis of total As, Se and Sb on titanium dioxide by slurry sampling-graphite furnace atomic absorption spectrometry. Talanta. 2005;68:336–42.CrossRefGoogle Scholar
  40. 40.
    Huang C, Jiang Z, Hu B. Mesoporous titanium dioxide as a novel solid-phase extraction material for flow injection micro-column preconcentration on-line coupled with ICP-OES determination of trace metals in environmental samples. Talanta. 2007;73:274–81.CrossRefGoogle Scholar
  41. 41.
    Dutta PK, Raya AK, Sharma VK, Millero FJ. Adsorption of arsenate and arsenite on titanium dioxide suspensions. J Colloid Interface Sci. 2004;278:270–5.CrossRefGoogle Scholar
  42. 42.
    Ragai J, Selim ST. Ion-exchange and surface properties of titania gels from Ti(III) solutions. J Colloid Interface Sci. 1987;115:139–46.CrossRefGoogle Scholar
  43. 43.
    Stumm W, Morgan JJ. Aquatic chemistry: chemical equilibria and rates in natural waters. New York: John Wiley & Sons; 2012.Google Scholar
  44. 44.
    Chubar NI, Samanidou VF, Kouts VS, Gallios GG, Kanibolotsky VA, Strelko VV, et al. Adsorption of fluoride, chloride, bromide, and bromate ions on a novel ion exchanger. J Colloid Interface Sci. 2005;291:67–74.CrossRefGoogle Scholar
  45. 45.
    Vassileva E, Proinova I, Hadjiivanov K. Solid-phase extraction of heavy metal ions on a high surface area titanium dioxide (anatase). Analyst. 1996;121:607–12.CrossRefGoogle Scholar
  46. 46.
    Vasileva E, Hadjiivanov K, Mandjukov P. Adsorption of Cr6+ oxo anions on pure and peroxide-modified TiO2 (anatase). Colloids Surf A Physicochem Eng Asp. 1994;90:9–15.CrossRefGoogle Scholar
  47. 47.
    Cruz-Vera M, Lucena R, Cárdenas S, Valcárcel M. Sample treatments based on dispersive (micro)extraction. Anal Methods. 2011;3:1719–28.CrossRefGoogle Scholar
  48. 48.
    Rydberg J. Solvent extraction principles and practice, revised and expanded. 2nd ed. Boca Raton: CRC Press; 2004.CrossRefGoogle Scholar
  49. 49.
    Hang YP, Liu JM. Determination of fluoride by an ion chromatography system using the preconcentration on nanometer-size zirconia. J Anal Chem. 2007;62:583–7.CrossRefGoogle Scholar
  50. 50.
    Yuchi A, Matsunaga K, Niwa T, Terao H, Wada H. Separation and preconcentration of fluoride at the ng ml−1 level with a polymer complex of zirconium(IV) followed by potentiometric determination in a flow system. Anal Chim Acta. 1999;388:201–8.CrossRefGoogle Scholar
  51. 51.
    Parham H, Rahbar N. Solid phase extraction-spectrophotometric determination of fluoride in water samples using magnetic iron oxide nanoparticles. Talanta. 2009;80:664–9.CrossRefGoogle Scholar
  52. 52.
    Wejnerowska G, Karczmarek A, Gaca J. Determination of fluoride in toothpaste using headspace solid-phase microextraction and gas chromatography-flame ionization detection. J Chromatogr A. 2007;1150:173–7.CrossRefGoogle Scholar
  53. 53.
    Kwon SM, Shin HS. Simple determination of fluoride in biological samples by headspace solid-phase microextraction and gas chromatography-tandem mass spectrometry. J Chromatogr A. 2015;1407:216–21.CrossRefGoogle Scholar

Copyright information

© The Author(s) 2017

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.