Introduction and hypothesis

We report a serious bleeding complication due to injury of the corona mortis following insertion of a transvaginal tape, TVT-Secur™ (Ethicon Women's Health, Sommerville, NJ, USA).

Methods

The TVT-Secur™ was inserted using the U-Approach: the tape was pushed along the periost behind the arc of the pubic bone at an angle of 45°.

Results

Surgery was unproblematic. Four hours after surgery, the patient showed shock signs; a CT scan revealed a 13.5 × 9.5-cm-sized retrosymphysic haematoma. Instant laparotomy with removal of the large haematoma, ligation of the bleeding corona mortis and administration of erythrocyte concentrates were necessary.

Conclusion

The corona mortis is an anastomosis between the arteria obturatoria and the external iliac vessels or its branches that passes over the superior pubic ramus. Its prevalence and anatomical relation to the pubic bone are important to bear in mind when applying new surgical techniques.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

References

  1. Larsson PG, Teleman P, Persson J (2010) A serious bleeding complication with injury of the corona mortis with the TVT-Secur procedure. Int Urogynecol J Pelvic Floor Dysfunct 21:1175–1177

    Article  PubMed  Google Scholar 

  2. Rusu MC, Cergan R, Motoc AG, Folescu R, Pop E (2010) Anatomical considerations on the corona mortis. Surg Radiol Anat 32:17–24

    Article  PubMed  Google Scholar 

  3. Hubka P, Svabik K, Martan A, Masata J (2010) A serious bleeding complication with injury of the corona mortis with the TVT-Secur procedure: two cases of contact of TVT-S with the corona mortis during cadaver study. Int Urogynecol J Pelvic Floor Dysfunct 21:1179–1180

    Article  PubMed  Google Scholar 

  4. Martan A, Svabik K, Masata J, Koleska T, El-Haddad R, Pavlikova M (2009) Correlation between changes in ultrasound measurements and clinical curative effect of tension-free vaginal tape-SECUR* procedure. Int Urogynecol J Pelvic Floor Dysfunct (in press)

Download references

Conflicts of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ursula Gobrecht.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Gobrecht, U., Kuhn, A. & Fellman, B. Injury of the corona mortis during vaginal tape insertion (TVT-Secur™ using the U-Approach). Int Urogynecol J 22, 443–445 (2011). https://doi.org/10.1007/s00192-010-1355-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00192-010-1355-2

Keywords

Navigation