Patient-specific total knee arthroplasty: the importance of planning by the surgeon

  • Knee
  • Published:
Knee Surgery, Sports Traumatology, Arthroscopy Aims and scope

Abstract

Purpose

The purpose of the study was to evaluate the accuracy of the planning of the patient-specific pin guides in total knee arthroplasty (TKA). This planning was performed primarily by a technician of the company and offered to the surgeon. All parameters of the implantation can either be modified or accepted by the surgeon. The hypothesis was that the plan needs preoperative intervention by the surgeon.

Methods

A prospective study in 50 patients was carried out. All patients received the same posterior-stabilised implant with patient-specific instrumentation. All surgical parameters (coronal, sagittal, rotational alignment, femoral and tibial resection levels and implant sizes) were checked by the orthopaedic surgeon and changed if necessary.

Results

Preoperatively, the femoral size was changed in 8 patients (16 %), the femoral flexion in 23 patients (46 %), the femoral shift in 34 patients (68 %), the tibial size in 24 patients (48 %) and the tibial rotation in all patients. The epicondylar axis was accepted in 47 patients (94 %) in the technician plan. Mean planning time was 8 ± 4 min. Intraoperatively, the femoral anterior–posterior size was in 50 patients (100 %) the same as in the surgeon and in 42 patients (84 %) the same as in the technician plan (p = 0.003). The tibial component implanted was in 42 patients (84 %) the same as in the surgeon and in 19 patients (38 %) the same as in the technician plan (p < 0.0001). A femoral distal recut was necessary in 31 patients (62 %) and a change of the tibial proximal cut in 17 patients (34 %) during surgery. Intraoperatively, no changes of the femoral and tibial alignment, the femoral anterior–posterior size, the femoral flexion, the femoral shift, the femoral and tibial rotation were necessary. Postoperatively, the coronal mechanical overall axis was within ±3° in 47 patients (94 %) with a maximum deviation of 5.6°.

Conclusions

Significant changes of the technician plan were necessary to get an accurate preoperative plan. Intraoperative changes were significant less compared to the surgeon than to the technician plan. No major changes (alignment, femoral anterior–posterior size and rotation) of the surgeon plan were necessary. Surgeons using patient-specific pin guides in TKA may verify the default plan provided by the technician. A blind reply on the technician plan may be not recommended.

Level of evidence

Therapeutic study, Level III.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Akagi M, Oh M, Nonaka T et al (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 420:213–219

    Article  PubMed  Google Scholar 

  2. Bali K, Walker P, Bruce W (2012) Custom-fit total knee arthroplasty: our initial experience in 32 knees. J Arthroplasty 27:1149–1154

    Article  PubMed  Google Scholar 

  3. Bathis H, Perlick L, Tingart M et al (2004) Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique. J Bone Joint Surg Br 86:682–687

    Article  PubMed  CAS  Google Scholar 

  4. Berger RA, Crossett LS, Jacobs JJ et al (1998) Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res 356:144–153

    Article  PubMed  Google Scholar 

  5. Chin PL, Yang KY, Yeo SJ et al (2005) Randomized control trial comparing radiographic total knee arthroplasty implant placement using computer navigation versus conventional technique. J Arthroplasty 20:618–626

    Article  PubMed  Google Scholar 

  6. Cobb JP, Dixon H, Dandachli W et al (2008) The anatomical tibial axis: reliable rotational orientation in knee replacement. J Bone Joint Surg Br 90:1032–1038

    Article  PubMed  CAS  Google Scholar 

  7. Dutton AQ, Yeo SJ, Yang KY et al (2008) Computer-assisted minimally invasive total knee arthroplasty compared with standard total knee arthroplasty. A prospective, randomized study. J Bone Joint Surg Am 90:2–9

    Article  PubMed  Google Scholar 

  8. Heyse TJ, Tibesku CO (2012) Improved femoral component rotation in TKA using patient-specific instrumentation. Knee. doi:10.1016/s09680160(12)00198-6

  9. Howell SM, Kuznik K, Hull ML et al (2008) Results of an initial experience with custom-fit positioning total knee arthroplasty in a series of 48 patients. Orthopedics 31:857–863

    Article  PubMed  Google Scholar 

  10. Incavo SJ, Coughlin KM, Pappas C et al (2003) Anatomic rotational relationships of the proximal tibia, distal femur, and patella: implications for rotational alignment in total knee arthroplasty. J Arthroplasty 18:643–648

    Article  PubMed  Google Scholar 

  11. Issa K, Rifai A, McGrath MS et al (2013) Reliability of templating with patient-specific instrumentation in total knee arthroplasty. J Knee Surg. doi:10.1055/s-0033-1343615

  12. Ng VY, Declaire JH, Berend KR et al (2011) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107

    Article  Google Scholar 

  13. Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplasty 27:153–155

    Article  PubMed  Google Scholar 

  14. Paley D, Tetsworth K (1992) Mechanical axis deviation of the lower limbs. Preoperative planning of multiapical frontal plane angular and bowing deformities of the femur and tibia. Clin Orthop Relat Res 280:65–71

    PubMed  Google Scholar 

  15. Pietsch M, Djahani O, Zweiger C et al (2012) Custom-fit minimally invasive total knee arthroplasty: effect on blood loss and early clinical outcomes. Knee Surg Sports Traumatol Arthrosc. doi:10.1007/s00167-012-2284-z

  16. Pietsch M, Hofmann S (2012) Early revision for isolated internal malrotation of the femoral component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 20:1057–1063

    Article  PubMed  Google Scholar 

  17. Spencer BA, Mont MA, McGrath MS et al (2009) Initial experience with custom-fit total knee replacement: intra-operative events and long-leg coronal alignment. Int Orthop 33:1571–1575

    Article  PubMed  Google Scholar 

  18. Stronach BM, Pelt CE, Erickson J et al (2013) Patient-specific total knee arthroplasty required frequent surgeon-directed changes. Clin Orthop Relat Res 471:169–174

    Article  PubMed  Google Scholar 

  19. Victor J, Dujardin J, Vandenneucker H et al (2013) Patient-specific guides do not improve accuracy in total knee arthroplasty: a prospective randomized controlled trial. Clin Orthop Relat Res. doi:10.1007/s11999-013-2997-4

  20. Whiteside LA, Arima J (1995) The anteroposterior axis for femoral rotational alignment in valgus total knee arthroplasty. Clin Orthop Relat Res 321:168–172

    PubMed  Google Scholar 

  21. Yoo JH, Chang CB, Shin KS et al (2008) Anatomical references to assess the posterior tibial slope in total knee arthroplasty: a comparison of 5 anatomical axes. J Arthroplasty 23:586–592

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. Pietsch.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pietsch, M., Djahani, O., Hochegger, M. et al. Patient-specific total knee arthroplasty: the importance of planning by the surgeon. Knee Surg Sports Traumatol Arthrosc 21, 2220–2226 (2013). https://doi.org/10.1007/s00167-013-2624-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00167-013-2624-7

Keywords

Navigation