Abstract

While the pulmonary artery catheter (PAC) is still interesting in specific situations, there are many alternatives. A group of experts from different backgrounds discusses their respective interests and limitations of the various techniques and related measured variables. The goal of this review is to highlight the conditions in which the alternative devices will suffice and when they will not or when these alternative techniques can provide information not available with PAC. The panel concluded that it is useful to combine different techniques instead of relying on a single one and to adapt the “package” of interventions to the condition of the patient. As a first step, the clinical and biologic signs should be used to identify patients with impaired tissue perfusion. Whenever available, echocardiography should be performed as it provides a rapid and comprehensive hemodynamic evaluation. If the patient responds rapidly to therapy, either no additional monitoring or pulse wave analysis (allowing continuous monitoring in case potential degradation is anticipated) can be applied. If the patient does not rapidly respond to therapy or complex hemodynamic alterations are observed, pulse wave analysis coupled with TPTD is suggested.

Introduction

Human right heart catheterization was first performed in 1929 by Werner Forssmann and then developed by André Cournand and Dickinson W. Richards, the three authors receiving the Nobel Prize for it in 1956 [1]. It was in the 1970s, when Drs. Swan and Ganz added balloon flotation [2] and thermodilution [3] to the catheterization technique, that the pulmonary artery catheter (PAC) became popular. The PAC has been considered useful, useless, and even harmful [4,5,6]. A meta-analysis of PAC efficacy and safety in 5051 patients (13 RCTs) showed no evidence of harm or of a conferred overall benefit [7]. These studies mostly demonstrated that PAC is safe when properly used. Recent studies have suggested better outcomes in selected patients with heart failure and trauma when a PAC was part of a given strategy of care [8, 9]. Still, no monitoring device will improve patient outcomes unless coupled to a treatment that itself improves outcomes [10, 11].

As the PAC is a multifaceted hemodynamic monitor, its use is complex. When using all its capabilities, it offers measurement of pulmonary artery pressure, pulmonary artery occlusion pressure (PAOP), right atrial pressure (RAP), cardiac output (CO), and mixed-venous oxygen saturation (SvO2). Pulmonary artery capillary pressure can be estimated and up to ten additional variables can be calculated [12]. Using all of this at the bedside is a challenge to even the most experienced clinician. However, the unique data available from a PAC make it an attractive hemodynamic monitor in the care of patients with severe circulatory shock, particularly those with right ventricular (RV) dysfunction and/or acute respiratory failure, as recommended by the ESICM task force [13].

Due to concerns about the invasiveness of the PAC, less or even non-invasive techniques have become available [14]. Alternative techniques, including minimally invasive monitoring and echocardiography, can display many of the hemodynamic values and variables captured by the PAC. Some of these devices can offer information in addition to that available with a PAC.

It is important to know the intrinsic value and limitations of variables measured by the alternative devices, as well as the missing information, and to determine in which conditions these devices will or will not suffice or when these alternative techniques can provide information not available with the PAC (Table 1). This ambitious review/consensus article aims to address many of the issues involved in the use of these alternative hemodynamic monitoring devices.

Table 1 Comparisons of variables obtained by the pulmonary artery catheter and alternative methods

Clinical indices of tissue perfusion

In the early phase of circulatory shock, clinical examination offers valuable information. Tachycardia, delirium/confusion, anxious breathing, cold and clammy skin, and persistent purpura are among the prominent abnormalities and are used to characterize patients with shock and assess tissue perfusion [15].

Relatively simple subjective measures of peripheral perfusion exist in addition to more objective measures [16]. The subjective assessment of the skin temperature, temperature of the great toe, temperature difference between the great toe and central temperature/ambient temperature, and skin temperature difference between the distal arm and finger all have clinical relevance [17,18,19]. The major issues are to determine whether the reference should be ambient or central temperature and to which extent therapy should be influenced by temperature differences.

Measurements of skin perfusion using optical techniques also relate to outcome in critically ill patients. The perfusion index (PI) derived from the pulse oximetry signal and tissue hemoglobin saturation (StO2) measured by near infrared spectroscopy are related to the progression of organ failure, failure to decrease lactate levels, and increased morbidity and mortality [19,20,21,22]. However, both methods are influenced by decreased skin temperature [20, 22]. Nevertheless, these devices provide continuous monitoring, which is a real advantage.

Assessments of skin mottling and capillary refill time are two methods that stand out because of their immediate availability at no cost without requiring specific equipment. Mottling has been shown to be a strong predictor of early mortality in septic shock patients [23, 24]. Scoring of mottling allows for better quantification of the response to therapy. In addition, the discolored skin areas relate to decreased StO2 [25] and prolonged capillary refill time [26]. The latter has also been shown to be related to increased morbidity and mortality and decreased visceral organ perfusion in critically ill patients [26, 27].

There are important limitations to these indices. First, while both identify issues with peripheral perfusion, the nature of the cause of tissue hypoperfusion (septic vs. low-output shock) remains to be identified by other hemodynamic monitoring devices. Second, peripheral perfusion may fail to reflect more central tissue perfusion [28] and may be affected by local arteriopathy.

Finally, are these indices simple disease identifiers (trigger for therapy) or can they be used as a target of therapy (should we aim at improving them and, if yes, to what goal)? Indicators of abnormal peripheral perfusion can be improved by a titrated infusion of nitroglycerin [29]. In septic shock patients, targeting fluid resuscitation based on peripheral perfusion may be safe and associated with less fluid administration compared with therapy based on systemic hemodynamic variables [18].

Biologic indices of tissue perfusion

The measurement of central venous oxygen saturation (ScvO2) is often taken as a surrogate of SvO2, but it is not equivalent to it. ScvO2 is measured in blood drained from the upper part of the body but not blood drained from the lower part and coronary sinus. As splanchnic organs represent one-fourth to one-third of the cardiac output and as preferential desaturation occurs in the splanchnic area in shock states [30], O2 saturation in hepatic veins can markedly affect the agreement between SvO2 and ScvO2. In normal conditions, the SvO2 value is slightly higher than that of ScvO2 because of the contribution of the very high renal venous O2 saturation. In shock states, the SvO2 value is usually lower than that of ScvO2 because of the more significant decrease in venous saturation in the splanchnic and renal circulation. Hence, even though globally correlated [31], ScvO2 is less sensitive to decreased perfusion compared with SvO2. ScvO2 gives no indication of the origin of the hemodynamic alteration. In addition, it is frequently elevated in distributive shock because of microcirculatory alterations and/or cellular dysfunction. Despite these limitations, it is important to measure ScvO2 to help in the interpretation of the data gathered by the other hemodynamic devices. In sum, a low ScvO2 value reflects circulatory stress, whereas a normal ScvO2 value does not ensure circulatory sufficiency.

Blood lactate is a key marker of tissue hypoperfusion. In experimental conditions, there is a sharp rise in lactate once oxygen consumption becomes limited by oxygen transport, a condition that defines shock [13]. As such, it is one of the key measurements to obtain in the management of shock [15, 32]. There are nevertheless two potential issues in the interpretation of lactate measurements. First, lactate is not only of hypoxic origin, but can also result from the activation of glycolytic pathways under the influence of inflammation or catecholamines [33]. Second, while lactate levels rise sharply in case of hypoxia, the decrease in lactate values may take time as its clearance may also be affected. Nevertheless, looking at lactate levels and their kinetics is recommended in the management of shock states [32].

Measurements of the gradient between venous and arterial PCO2 (PvaCO2) can also be informative in the evaluation of the adequacy of tissue perfusion. In normal conditions, this gradient is ≤ 6 mmHg, so that a gradient > 6 mmHg suggests an inadequate tissue perfusion. Increased PvaCO2 may be the consequence of inadequate cardiac output (then ScvO2 levels should also be low) or microcirculatory dysfunction (then ScvO2 levels would usually be high) [34, 35].

It is important to realize that ScvO2, lactate and PvaCO2 values indicate that there is a circulatory problem, but none of these indicate the origin of the problem. Combination of these biologic measures with other hemodynamic variables is hence recommended [14].

Central venous pressure: uses and misuses

The central venous pressure (CVP) is measured at the tip of a central venous catheter. Its normal value ranges between 0 and 3 mmHg. It is an accepted surrogate of RAP.

CVP can be used for several purposes:

  • To optimize tissue perfusion pressure: As CVP is the back pressure for venous return, an elevated CVP results in an elevated tissue capillary pressure—causing “hydrostatic” tissue edema—and may compromise organ perfusion by decreasing organ perfusion pressure. In practice, to select the optimal mean arterial pressure (i.e., the upstream pressure for organ blood flow), CVP should be taken into account when it is high and neglected when it is low.

  • To guide fluid resuscitation: CVP is also considered representative of the RV filling pressure and hence an estimate of RV preload. However, both assumptions may not be correct in all conditions. First, CVP represents the RV filling pressure only if the RV surrounding pressure is neither too elevated nor too low. The RV surrounding pressure is either pericardial or intrathoracic pressure, whichever is higher [36]. To minimize increased intrathoracic pressure effects, CVP must be measured at end-expiration. However, in cases of (extrinsic or intrinsic) positive end-expiratory pressure, the measured CVP overestimates the transmural CVP. Transmural CVP is difficult to determine in practice. Second, the relationship of RV filling pressure and RV preload is complex: Under normal conditions, the RV distending pressure increases little during filling as the RV dilates. In cases of RV overdistension or hypertrophy, the opposite phenomenon occurs so that increases in CVP overestimate increases in RV volume. Consequently, CVP and its changes are not markers of RV preload. Importantly, even if we assume CVP to reflect RV preload, it is a rather poor marker of preload responsiveness. Indeed, due to inter-individual differences in the RV preload-stroke volume relationships (Frank-Starling mechanism), assessing preload is not assessing preload responsiveness. Therefore, a marker of preload cannot reliably predict fluid responsiveness, even though extreme values can be useful to predict the response to fluids [37]. Hence, recent guidelines suggest the use of dynamic markers of fluid responsiveness for fluid resuscitation in septic patients [13, 32]. Nevertheless, dynamic evaluation of CVP can be used to evaluate the patient’s tolerance of fluid administration [37].

  • To evaluate right ventricular dysfunction: As reported above, the relationship between the CVP and right ventricular volume is curvilinear. Hence, a significant right ventricular dysfunction is associated with an increased CVP.

In summary, CVP is important: (1) to estimate the backpressure for venous return and (2) to identify situations of high RV filling pressures (transmural CVP should be considered) and hence of RV dysfunction (to be confirmed by echocardiography). The CVP is a poor predictor of fluid responsiveness, but a dynamic evaluation of CVP can be used to assess the patient’s tolerance to fluid.

Cardiac output measurement by pulse wave analysis

The estimation of stroke volume (SV) and CO based on the mathematical analysis of the arterial blood pressure waveform is called pulse wave analysis, pulse pressure analysis, or pulse contour analysis. Left ventricular SV is one of the primary variables defining arterial pulse pressure, so extrapolating SV from pulse pressure seems reasonable.

Numerous algorithms have been developed to assess CO using pulse wave analysis [14, 38]. The shape of the waveform depends on various factors including ventricular ejection, aortic impedance, arterial compliance, and vasomotor tone that all influence the velocity of the pressure wave in large arteries and the degree of pressure wave reflection in peripheral arteries [14].

The arterial blood pressure waveform used for pulse wave analysis can be recorded invasively with an arterial catheter or noninvasively with probes placed on the patient’s finger (vascular unloading technology, also called the volume clamp method or simply the finger cuff technology) or over the radial artery (automated radial artery applanation tonometry). The measurement performance of pulse wave analysis depends on the impeccable quality of the arterial blood pressure waveform, which is often impaired in severely vasoconstricted patients.

Depending on the system used, the pulse wave analysis-derived values of SV or CO can be “uncalibrated” or “calibrated” to external values (e.g., obtained with dilution techniques). Calibration of the CO value to an external reference CO value increases the accuracy and precision of pulse wave analysis-derived CO measurements and is recommended in patients with deceased vasomotor tone such as septic patients or patients with liver failure [14]. However, these require frequent re-calibration, especially when vascular tone changes [39]. Uncalibrated systems estimate values of CO solely by analyzing the characteristics of the arterial pressure waveform and using patient-specific anthropometric, biometric, and demographic data. Uncalibrated systems differ in the way they assess the extent to which changes in the shape of the arterial pressure waveform are caused by changes in SV and/or changes in vascular tone. In most cases, the accuracy and precision of uncalibrated devices are moderate to poor [40, 41], but changes induced by fluids can reliably be tracked. Some systems use dedicated algorithms to also take into account vascular tone. While initial versions were not reliable in patients with low vascular tone [42], recent developments have made these more reliable even in these patients [43, 44].

Whatever the system used, marked and/or rapid changes in peripheral vascular resistance may affect the reliability of CO measurement from pulse wave analysis [39, 45]. In calibrated methods, recalibration helps to regain accuracy [39]. Uncalibrated methods assessing vascular tone require time to recover accuracy, and reliability is lost until the new equilibrium is reached.

The main advantage of pulse wave analysis is that it provides a continuous estimation of real-time CO allowing immediate detection of changes in CO in response to time, treatment, or maneuvers. As the value of SV may change with each single beat, pulse wave analysis also provides dynamic cardiac preload variables [respiratory variations in pulse pressure (PPV) and SV or response in cardiac output during a passive leg-raising test]. These indices outperform filling pressures in the prediction of fluid responsiveness and are now recommended for the management of shock and sepsis [13, 32].

The main disadvantage of most methods (except devices calibrated by transpulmonary thermodilution—see below) is that minimal information is provided beyond cardiac output measurements and dynamic cardiac preload variables, making their use less relevant in complex patients.

Transpulmonary thermodilution

Transpulmonary thermodilution (TPTD) is used to calibrate some pulse contour/pulse wave analysis devices. Transpulmonary thermodilution is as reliable as right-sided thermodilution for the measurement of CO.

In addition to CO, TPTD provides the following variables: the global end-diastolic volume (GEDV), which assesses preload volumetrically; the extravascular lung water (EVLW), which assesses the degree of pulmonary edema [46]; and the pulmonary vascular permeability index (PVPI), which may help to differentiate between cardiogenic and permeability pulmonary edema [47]. The respective advantages and disadvantages of the different variables are presented in Table 2. Of note, EVLW and PAOP do not provide identical information. EVLW provides information on the amount of fluid accumulated in the lung, while PAOP (or, even better, true pulmonary artery capillary pressure, which unfortunately is difficult to measure in routine practice) represents the instantaneous hydrostatic driving force.

Table 2 Advantages and drawbacks of transpulmonary thermodilution devices

As filling pressures with PAC, measurement of cardiac volumes is mostly interesting for the evaluation of cardiac function. Like any other static measurement, cardiac volumes do not predict the response to fluids. In addition, it is impossible to differentiate between right- and left-side dilation when faced with an increased end-diastolic volume. Hence, echocardiography is required to confirm the diagnosis when alterations in cardiac volumes are noticed.

Interestingly, a recent trial looking at the interpretation of simultaneously obtained TPTD and critical care echocardiography data in 127 patients with septic shock found good agreement with the concordance of therapeutic decisions in 78% of the patients [48]. While this does not imply that similar findings would be observed in other types of shock, this illustrates the clinical interest of the measured variables.

The combination of all the measured variables makes TPTD a valuable, less-invasive alternative to the PAC, especially in complex situations, such as hemodynamic instability associated with acute respiratory distress syndrome (ARDS), when each of the possible interventions may carry significant harm (‘therapeutic conflict’). In particular, fluid management based on fluid responsiveness variables has a better rationale than those based on filling pressures. The simultaneously measured EVLW and PVPI may identify patients at risk of pulmonary edema during fluid administration.

The additional information provided by TPTD has been shown to significantly alter physicians’ therapeutic decisions [49]. Together with the PAC, the use of TPTD has been recommended in the management of patients in severe shock and complex situations [13, 14]. While volumetric indices and intravascular pressures are physiologically related, the information on cardiovascular function provided by TPTD and PAC may sometimes diverge because of the curvilinear aspect of pressure/volume curves and errors in measurements. In most cases, the two techniques result in similar diagnostic classification [50], but divergence can be observed in some patients [50, 51].

Although considered to be less invasive than the PAC, TPTD still necessitates the insertion of a central venous catheter and the cannulation of a large artery (usually the femoral artery). The most commonly reported complications were small local hematomas (4.5%) and infections (2%) [52].

Direct head-to-head comparisons of outcomes with the two techniques are very limited. In a cohort of patients monitored with either TPTD or PAC, patients with TPTD had a greater positive fluid balance and fewer ventilator-free days compared with those with PAC [53]. There was no difference after adjustment for potential confounding factors. In a small randomized trial, no difference in survival was observed; nevertheless, the duration of mechanical ventilation was shorter with the PAC in the subgroup of patients with impaired cardiac function but not in the patients with ARDS [54].

In summary, compared with the PAC, TPTD provides an equally accurate CO, better assessment of fluid status, and the unique measurement of the EVLW, which can be helpful in cases of ARDS. Its main disadvantage is its poor ability to distinguish between right and left ventricular performance.

The role of echocardiography and lung ultrasounds

Echocardiography is being increasingly used in critically ill patients [55]. It is nowadays more frequently used than the PAC in patients with sepsis or congestive heart failure [55]. Echocardiography is usually considered an alternative to PAC for measuring and monitoring CO, but this is questionable. While the PAC may give continuous access to CO, echocardiography only provides an intermittent measurement. A recent systematic review of the literature reported that PAC and echocardiography were not interchangeable for evaluating CO and that, from an evidence-based point of view, echo studies were methodologically limited [56]. However, the disagreement mostly came when old studies were included, and new studies with better methodology for determining CO found better agreement [57]. Importantly, trends in CO, which are much more important in clinical practice than absolute CO values, are accurately detected by echocardiography [56].

For measuring CO, echocardiography has some advantages over PAC. First, echocardiography is less invasive, even with the transesophageal approach [58]. Second, echocardiography can be used in some situations in which PAC is limited, such as massive tricuspid regurgitation or very low flow states. Third, in contrast to PAC, echocardiography provides a beat-to-beat measurement of SV, which allows the assessment of fluid responsiveness through dynamic indices. Finally, and more importantly, echocardiography in the ICU provides much more information besides CO.

The concept of critical care echocardiography (CCE) nicely reflects the ubiquitous utility of echocardiography in the management of critically ill patients [59]. By its ability to directly visualize the heart function, echocardiography is unique in defining the mechanism of shock [15]. In contrast to the PAC, it provides a direct evaluation of left and RV function and of valvular structure. In patients with circulatory failure, a quick assessment of heart function by CCE is recommended, after having eliminated obvious hypovolemia [11, 13, 14]. One of the most important advantages of CCE is the understanding of heart-lung interactions in mechanically ventilated patients [60]. By assessing the right cardiac function, CCE provides important hemodynamic information to guide hemodynamic and respiratory therapy [61]. Among all its potential applications, CCE also allows evaluating the need for fluids through several dynamic indices, which is not possible with the PAC [62]. Finally, CCE is a unique tool that detects left ventricular diastolic dysfunction, which is strongly associated with weaning failure [63] and mortality in septic shock [64].

One of the most important limitations of echocardiography is its intermittent nature. Continuous miniaturized TEE has been developed [65], but these techniques only offer limited information beyond fluid responsiveness and visual evaluation of the ventricles as quantitative Doppler assessment is not feasible. Accordingly, measurements of filling pressures, pulmonary artery pressure, and cardiac output are not feasible with these devices.

Intensivists can also use the ultrasound probe for lung ultrasonography. The latter accurately detects pulmonary edema, condensation, pneumothorax, or pleural effusion [66]. Combining these approaches of critical care ultrasonography may be useful in situations with both respiratory and circulatory failures, such as pulmonary embolism, pulmonary edema (cardiogenic or non cardiogenic), sepsis-related pneumonia, or weaning failure [67].

Evaluating the microcirculation

Microvascular abnormalities are commonly observed in critically ill patients [68,69,70]. These are characterized by heterogeneity in perfusion, which results in hypoxic areas coexisting with over-perfused areas. Their severity and duration are associated with organ dysfunction and mortality [69, 71, 72]. Various mechanisms may be implicated [73].

Conventional hemodynamic tools often fail to detect it. Different hand-held microscopes are used for direct observation of microcirculation, mostly applied to the sublingual area [69,70,71, 74, 75]. The advantage of these devices is that they are non-invasive and easy and fast to use. Exploration of the microcirculation is intermittent but can be frequently repeated without inconvenience to the patient. Importantly, evaluation of the microcirculation requires acquisition of good quality images and a strict protocol for image analysis [76, 77].

Videomicroscopes cannot be used in all circumstances. They require technical skills and patient cooperation; they are also difficult to apply in dyspneic patients. Hence, alternative ways to routinely evaluate microcirculatory alterations are desired. Clinical signs of peripheral tissue hypoperfusion may fail to detect central alterations in microvascular perfusion [28]. Blood lactate levels are often increased in patients with microvascular alterations, but the slow changes in lactate kinetics make it a very indirect marker of microvascular dysfunction. Due to local shunting, SvO2 is not a good reflect of microvascular alterations. Interestingly, an increase in PvaCO2 is a good indicator of microvascular perfusion abnormalities [34], especially when the SvO2 value is normal, and changes in PvaCO2 reflect changes in microvascular perfusion more than changes in CO [34].

An important question is whether microcirculation targeted therapy is superior to “classical” hemodynamic monitoring once satisfactory systemic hemodynamic targets have been reached. At this stage, it seems premature to address this question. While there is no doubt that a better understanding of the pathophysiologic processes is desired, the major actual limitation is that we lack therapies specifically acting at the microcirculatory level. Fluids given in the early stages [78, 79] and dobutamine [80] may somewhat improve the microcirculation, but their effects are quite variable. Vasodilatory agents may improve the microvascular perfusion, but they lack selectivity and increase flow in already perfused areas as well. Modulation of endothelial nitric oxide synthase with various agents (including vitamin C) appears promising. Hence, it is important to understand the microcirculatory changes that occur with various therapies to plan resuscitation targets for the future.

Specificities in limited resource settings

Resource-limited settings include not only parts of the developing world where lack of healthcare funds lead to limited access to medical resources, but also include areas with limited access to monitoring within the hospital in developed countries, such as the emergency department and wards. The different hemodynamic resources available in these areas are presented in Table 3.

Table 3 Alternatives to the pulmonary artery catheter with special emphasis on techniques available in low-resource settings

Acute circulatory failure can be assessed using the three clinical windows for tissue perfusion [13, 15]: altered mentation, cold and mottled skin and oliguria, and with tachycardia and hypotension. The computation of the “shock index”, (ratio of heart rate to systolic blood pressure; normal range, 0.5–0.7) can also be useful [81].

The evaluation of cardiovascular function is much more difficult. When available, echocardiography and ultrasound are the techniques of choice. Though requiring provider skills, these give rapid access to the volume status, cardiac function, and assessment of lung edema. Echographic assessment is relatively cheap, though involving some initial expenditure.

The decision to give fluids can be triggered by clinical signs of organ hypoperfusion and elevated lactate levels in the absence of raised jugular venous pressure. When a central venous catheter is in place, the CVP, ScvO2, and PvaCO2 values help to monitor the physiologic process and guide therapy. However, as reported above, CVP has its limitations, and the other indices indicate inadequate perfusion but not fluid responsiveness.

As both excessive and insufficient fluid administration can worsen outcome, it is important to identify fluid responders before giving fluids. The effect of inadequate fluid administration is probably even more detrimental in resource-limited areas. In African children with septic shock, mortality increased when fluids were given until signs of intolerance occurred [82]. Several strategies can be used to predict response to fluids. PPV nicely predicts fluid responsiveness and does not require specific equipment if an arterial catheter is used [83]. These can be automatically calculated by most bedside monitors. However, it is unreliable during low tidal volume ventilation [84], which is now routinely used in all critically ill patients. The ‘tidal volume challenge’ is a novel test that may help to circumvent this limitation [85]. This involves transiently increasing tidal volume from 6 ml/kg PBW to 8 ml/kg PBW for 1 min and observing the change in PPV (ΔPPV6–8). Plethysmographic indices of fluid responsiveness are non-invasive and quickly obtained [86]; however, the signal quality is critically dependent on the peripheral perfusion [87]. Of note, SV variations and derived variables are also present in acute cor pulmonale [88]. A passive leg-raising test can be used and has the advantage of avoiding these limitations, but is also somewhat cumbersome. It requires evaluating changes in cardiac output in relation to changes in arterial pulse pressure or, better, to changes in end-tidal CO2 in intubated patients [89], or echocardiography or carotid/brachial Doppler, especially in emergency departments/wards, which usually have access to linear ultrasound probes.

Finally, studies are needed to evaluate whether bundles, protocols, and potentially telemedicine, would cost-effectively improve outcomes in limited-resource settings.

Putting it all together

As described in this article, there are many alternatives to the PAC. Each has its own interesting features and limitations. Instead of relying on a single technique, it is useful to combine different techniques and, even more importantly, to adapt the “package” of interventions to the patient's condition, taking into consideration the actual hemodynamic state and the likelihood of deterioration and comorbidities. As a first step, the clinical and biologic signs should be used to identify patients with impaired tissue perfusion. Whenever available, echocardiography should be performed as it provides a rapid and comprehensive hemodynamic evaluation. If the patient rapidly responds to therapy, either no additional monitoring or pulse wave analysis (allowing continuous monitoring in case potential degradation is anticipated) can be applied. If the patient does not rapidly respond to therapy or if complex hemodynamic alterations are observed, pulse wave analysis coupled with TPTD is suggested, supplemented, of course, with regular assessment of clinical and biologic indices of perfusion and coupled, if needed, with repeated echocardiographic evaluation. When macrocirculation seems within reasonable goals and the metabolic indices remain altered, it seems reasonable to suspect microvascular alterations. While therapeutic options remain limited at this stage, the indication is that further attempting to increase CO is probably unnecessary.

References

  1. The Nobel Prize in Physiology or Medicine 1956. 2014. Nobelprize.org <http://www.nobelprize.org/nobel_prizes/medicine/laureates/1956/>. Accessed 8-4-2018 (Online Source)

  2. Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D (1970) Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med 283:447–451

    Article  CAS  PubMed  Google Scholar 

  3. Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ (1971) A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396

    Article  CAS  PubMed  Google Scholar 

  4. Shoemaker WC, Appel PL, Kram HB, Waxman K, Lee T-S (1988) Prospective trial of supranormal values of survivors as therapeutic goals in high-risk surgical patients. Chest 94:1176–1186

    Article  CAS  PubMed  Google Scholar 

  5. Connors AF, Speroff T, Dawson NV, Thomas C, Harrell FE, Wagner D et al (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA 276:889–897

    Article  PubMed  Google Scholar 

  6. Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D et al (2003) Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA 290:2713–2720

    Article  CAS  PubMed  Google Scholar 

  7. Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G et al (2005) Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA 294:1664–1670

    Article  CAS  PubMed  Google Scholar 

  8. Sotomi Y, Sato N, Kajimoto K, Sakata Y, Mizuno M, Minami Y et al (2014) Impact of pulmonary artery catheter on outcome in patients with acute heart failure syndromes with hypotension or receiving inotropes: from the ATTEND Registry. Int J Cardiol 172:165–172

    Article  PubMed  Google Scholar 

  9. Friese RS, Shafi S, Gentilello LM (2006) Pulmonary artery catheter use is associated with reduced mortality in severely injured patients: a national trauma data bank analysis of 53,312 patients. Crit Care Med 34:1597–1601

    Article  PubMed  Google Scholar 

  10. Pinsky MR, Vincent JL (2005) Let us use the pulmonary artery catheter correctly and only when we need it. Crit Care Med 33:1119–1122

    Article  PubMed  Google Scholar 

  11. Vincent JL, Rhodes A, Perel A, Martin GS, Della RG, Vallet B et al (2011) Clinical review: update on hemodynamic monitoring—a consensus of 16. Crit Care 15:229

    Article  PubMed  PubMed Central  Google Scholar 

  12. De Backer D, Fagnoul D, Herpain A (2013) The role of invasive techniques in cardiopulmonary evaluation. Curr Opin Crit Care 19:228–233

    Article  PubMed  Google Scholar 

  13. Cecconi M, De Backer D, Antonelli M, Beale RJ, Bakker J, Hofer C et al (2014) Consensus on circulatory shock and hemodynamic monitoring. Task Force of the European Society of Intensive Care Medicine. Intensive Care Med 40:1795–1815

    Article  PubMed  PubMed Central  Google Scholar 

  14. Teboul JL, Saugel B, Cecconi M, De Backer D, Hofer CK, Monnet X et al (2016) Less invasive hemodynamic monitoring in critically ill patients. Intensive Care Med 42:1350–1359

    Article  PubMed  Google Scholar 

  15. Vincent JL, De Backer D (2013) Circulatory shock. N Engl J Med 369:1726–1734

    Article  CAS  PubMed  Google Scholar 

  16. Ait-Oufella H, Bakker J (2016) Understanding clinical signs of poor tissue perfusion during septic shock. Intensive Care Med 42:2070–2072

    Article  PubMed  Google Scholar 

  17. Joly HR, Weil MH (1969) Temperature of the great toe as an indication of the severity of shock. Circulation 39:131–138

    Article  CAS  PubMed  Google Scholar 

  18. van Genderen ME, Engels N, van der Valk RJ, Lima A, Klijn E, Bakker J et al (2015) Early peripheral perfusion-guided fluid therapy in patients with septic shock. Am J Respir Crit Care Med 191:477–480

    Article  PubMed  Google Scholar 

  19. Lima A, Jansen TC, van Bommel J, Ince C, Bakker J (2009) The prognostic value of the subjective assessment of peripheral perfusion in critically ill patients. Crit Care Med 37:934–938

    Article  PubMed  Google Scholar 

  20. Lima A, van Bommel J, Sikorska K, van Genderen M, Klijn E, Lesaffre E et al (2011) The relation of near-infrared spectroscopy with changes in peripheral circulation in critically ill patients. Crit Care Med 39:1649–1654

    Article  PubMed  Google Scholar 

  21. Oskay A, Eray O, Dinc SE, Aydin AG, Eken C (2015) Prognosis of critically ill patients in the ED and value of perfusion index measurement: a cross-sectional study. Am J Emerg Med 33:1042–1044

    Article  PubMed  Google Scholar 

  22. He HW, Liu DW, Long Y, Wang XT (2013) The peripheral perfusion index and transcutaneous oxygen challenge test are predictive of mortality in septic patients after resuscitation. Crit Care 17:R116

    Article  PubMed  PubMed Central  Google Scholar 

  23. Ait-Oufella H, Lemoinne S, Boelle PY, Galbois A, Baudel JL, Lemant J et al (2011) Mottling score predicts survival in septic shock. Intensive Care Med 37:801–807

    Article  CAS  PubMed  Google Scholar 

  24. Coudroy R, Jamet A, Frat JP, Veinstein A, Chatellier D, Goudet V et al (2015) Incidence and impact of skin mottling over the knee and its duration on outcome in critically ill patients. Intensive Care Med 41:452–459

    Article  PubMed  Google Scholar 

  25. Ait-Oufella H, Joffre J, Boelle PY, Galbois A, Bourcier S, Baudel JL et al (2012) Knee area tissue oxygen saturation is predictive of 14-day mortality in septic shock. Intensive Care Med 38:976–983

    Article  CAS  PubMed  Google Scholar 

  26. Ait-Oufella H, Bige N, Boelle PY, Pichereau C, Alves M, Bertinchamp R et al (2014) Capillary refill time exploration during septic shock. Intensive Care Med 40:958–964

    Article  CAS  PubMed  Google Scholar 

  27. Brunauer A, Kokofer A, Bataar O, Gradwohl-Matis I, Dankl D, Bakker J et al (2016) Changes in peripheral perfusion relate to visceral organ perfusion in early septic shock: a pilot study. J Crit Care 35:105–109

    Article  PubMed  Google Scholar 

  28. Boerma EC, Kuiper MA, Kingma WP, Egbers PH, Gerritsen RT, Ince C (2008) Disparity between skin perfusion and sublingual microcirculatory alterations in severe sepsis and septic shock: a prospective observational study. Intensive Care Med 34:1294–1298

    Article  PubMed  PubMed Central  Google Scholar 

  29. Lima A, van Genderen ME, van Bommel J, Klijn E, Jansem T, Bakker J (2014) Nitroglycerin reverts clinical manifestations of poor peripheral perfusion in patients with circulatory shock. Crit Care 18:R126

    Article  PubMed  PubMed Central  Google Scholar 

  30. De Backer D, Creteur J, Noordally O, Smail N, Gulbis B, Vincent JL (1998) Does hepato-splanchnic VO2/DO2 dependency exist in critically ill septic patients? Am J Respir Crit Care Med 157:1219–1225

    Article  PubMed  Google Scholar 

  31. Reinhart K, Kuhn HJ, Hartog C, Bredle DL (2004) Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill. Intensive Care Med 30:1572–1578

    Article  PubMed  Google Scholar 

  32. Rhodes A, Evans LE, Alhazzani W, Levy MM, Antonelli M, Ferrer R et al (2017) Surviving sepsis campaign: international guidelines for management of sepsis and septic shock: 2016. Intensive Care Med 43:304–377

    Article  PubMed  Google Scholar 

  33. Levy B, Gibot S, Franck P, Cravoisy A, Bollaert PE (2005) Relation between muscle Na + K + ATPase activity and raised lactate concentrations in septic shock: a prospective study. Lancet 365:871–875

    Article  CAS  PubMed  Google Scholar 

  34. Ospina-Tascon GA, Umana M, Bermudez WF, Bautista-Rincon DF, Valencia JD, Madrinan HJ et al (2016) Can venous-to-arterial carbon dioxide differences reflect microcirculatory alterations in patients with septic shock? Intensive Care Med 42:211–221

    Article  CAS  PubMed  Google Scholar 

  35. Perner A, Gordon AC, De Backer D, Dimopoulos G, Russell JA, Lipman J et al (2016) Sepsis: frontiers in diagnosis, resuscitation and antibiotic therapy. Intensive Care Med 42:1958–1969

    Article  CAS  PubMed  Google Scholar 

  36. Pinsky M, Vincent JL, De Smet JM (1991) Estimating left ventricular filling pressure during positive end-expiratory pressure in humans. Am Rev Respir Dis 143:25–31

    Article  CAS  PubMed  Google Scholar 

  37. De Backer D, Vincent JL (2018) Should we measure the central venous pressure to guide fluid management? Ten answers to 10 questions. Crit Care 22:43

    Article  PubMed  PubMed Central  Google Scholar 

  38. Hadian M, Kim HK, Severyn DA, Pinsky MR (2010) Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters. Crit Care 14:R212

    Article  PubMed  PubMed Central  Google Scholar 

  39. Hamzaoui O, Monnet X, Richard C, Osman D, Chemla D, Teboul JL (2008) Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period. Crit Care Med 36:434–440

    Article  PubMed  Google Scholar 

  40. Joosten A, Desebbe O, Suehiro K, Murphy LS, Essiet M, Alexander B et al (2017) Accuracy and precision of non-invasive cardiac output monitoring devices in perioperative medicine: a systematic review and meta-analysisdagger. Br J Anaesth 118:298–310

    Article  CAS  PubMed  Google Scholar 

  41. Peyton PJ, Chong SW (2010) Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology 113:1220–1235

    Article  PubMed  Google Scholar 

  42. Sakka SG, Kozieras J, Thuemer O, van Hout N (2007) Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth 99:337–342

    Article  CAS  PubMed  Google Scholar 

  43. Scolletta S, Franchi F, Romagnoli S, Carla R, Donati A, Fabbri LP et al (2016) Comparison between Doppler-Echocardiography and uncalibrated pulse contour method for cardiac output measurement: a multicenter observational study. Crit Care Med 44:1370–1379

    Article  PubMed  Google Scholar 

  44. De Backer D, Marx G, Tan A, Junker C, Van NM, Huter L et al (2011) Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients. Intensive Care Med 37:233–240

    Article  PubMed  Google Scholar 

  45. Meng L, Phuong TN, Alexander BS, Laning K, Chen G, Kain ZN et al (2011) The impact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FLOTRAC and esophageal Doppler cardiac output measurements. Anesth Analg 113:751–757

    Article  CAS  PubMed  Google Scholar 

  46. Monnet X, Teboul JL (2017) Transpulmonary thermodilution: advantages and limits. Crit Care 21:147

    Article  PubMed  PubMed Central  Google Scholar 

  47. Monnet X, Anguel N, Osman D, Hamzaoui O, Richard C, Teboul JL (2007) Assessing pulmonary permeability by transpulmonary thermodilution allows differentiation of hydrostatic pulmonary edema from ALI/ARDS. Intensive Care Med 33:448–453

    Article  PubMed  Google Scholar 

  48. Vignon P, Begot E, Mari A, Silva S, Chimot L, Delour P et al (2018) Hemodynamic assessment of patients with septic shock using transpulmonary thermodilution and critical care echocardiography: a comparative study. Chest 153:55–64

    Article  PubMed  Google Scholar 

  49. Perel A, Saugel B, Teboul JL, Malbrain ML, Belda FJ, Fernandez-Mondejar E et al (2016) The effects of advanced monitoring on hemodynamic management in critically ill patients: a pre and post questionnaire study. J Clin Monit Comput 30:511–518

    Article  PubMed  Google Scholar 

  50. Ritter S, Rudiger A, Maggiorini M (2009) Transpulmonary thermodilution-derived cardiac function index identifies cardiac dysfunction in acute heart failure and septic patients: an observational study. Crit Care 13:R133

    Article  PubMed  PubMed Central  Google Scholar 

  51. Hilty MP, Franzen DP, Wyss C, Biaggi P, Maggiorini M (2017) Validation of transpulmonary thermodilution variables in hemodynamically stable patients with heart diseases. Ann Intensive Care 7:86

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  52. Belda FJ, Aguilar G, Teboul JL, Pestana D, Redondo FJ, Malbrain M et al (2011) Complications related to less-invasive haemodynamic monitoring. Br J Anaesth 106:482–486

    Article  CAS  PubMed  Google Scholar 

  53. Uchino S, Bellomo R, Morimatsu H, Sugihara M, French C, Stephens D et al (2006) Pulmonary artery catheter versus pulse contour analysis: a prospective epidemiological study. Crit Care 10:R174

    Article  PubMed  PubMed Central  Google Scholar 

  54. Trof RJ, Beishuizen A, Cornet AD, de Wit RJ, Girbes AR, Groeneveld AB (2012) Volume-limited versus pressure-limited hemodynamic management in septic and nonseptic shock. Crit Care Med 40:1177–1185

    Article  PubMed  Google Scholar 

  55. Papolos A, Narula J, Bavishi C, Chaudhry FA, Sengupta PP (2016) U.S. Hospital use of echocardiography: insights from the Nationwide Inpatient Sample. J Am Coll Cardiol 67:502–511

    Article  PubMed  Google Scholar 

  56. Wetterslev M, Moller-Sorensen H, Johansen RR, Perner A (2016) Systematic review of cardiac output measurements by echocardiography vs. thermodilution: the techniques are not interchangeable. Intensive Care Med 42:1223–1233

    Article  PubMed  Google Scholar 

  57. Mercado P, Maizel J, Beyls C, Titeca-Beauport D, Joris M, Kontar L et al (2017) Transthoracic echocardiography: an accurate and precise method for estimating cardiac output in the critically ill patient. Crit Care 21:136

    Article  PubMed  PubMed Central  Google Scholar 

  58. Huttemann E, Schelenz C, Kara F, Chatzinikolaou K, Reinhart K (2004) The use and safety of transoesophageal echocardiography in the general ICU—a minireview. Acta Anaesthesiol Scand 48:827–836

    Article  CAS  PubMed  Google Scholar 

  59. Mayo PH, Beaulieu Y, Doelken P, Feller-Kopman D, Harrod C, Kaplan A et al (2009) American College of Chest Physicians/La Societe de Reanimation de Langue Francaise statement on competence in critical care ultrasonography. Chest 135:1050–1060

    Article  PubMed  Google Scholar 

  60. Vieillard-Baron A, Chergui K, Augarde R, Prin S, Page B, Beauchet A et al (2003) Cyclic changes in arterial pulse during respiratory support revisited by Doppler echocardiography. Am J Respir Crit Care Med 168:671–676

    Article  PubMed  Google Scholar 

  61. Vieillard-Baron A, Matthay M, Teboul JL, Bein T, Schultz M, Magder S et al (2016) Experts’ opinion on management of hemodynamics in ARDS patients: focus on the effects of mechanical ventilation. Intensive Care Med 42:739–749

    Article  CAS  PubMed  Google Scholar 

  62. Vignon P, Repesse X, Begot E, Leger J, Jacob C, Bouferrache K et al (2016) Comparison of echocardiographic indices used to predict fluid responsiveness in ventilated patients. Am J Respir Crit Care Med 195:1022–1032

    Article  Google Scholar 

  63. Papanikolaou J, Makris D, Saranteas T, Karakitsos D, Zintzaras E, Karabinis A et al (2011) New insights into weaning from mechanical ventilation: left ventricular diastolic dysfunction is a key player. Intensive Care Med 37:1976–1985

    Article  PubMed  Google Scholar 

  64. Sanfilippo F, Corredor C, Arcadipane A, Landesberg G, Vieillard-Baron A, Cecconi M et al (2017) Tissue Doppler assessment of diastolic function and relationship with mortality in critically ill septic patients: a systematic review and meta-analysis. Br J Anaesth 119:583–594

    Article  CAS  PubMed  Google Scholar 

  65. Vieillard-Baron A, Slama M, Mayo P, Charron C, Amiel JB, Esterez C et al (2013) A pilot study on safety and clinical utility of a single-use 72-hour indwelling transesophageal echocardiography probe. Intensive Care Med 39:629–635

    Article  PubMed  Google Scholar 

  66. Volpicelli G, Elbarbary M, Blaivas M, Lichtenstein DA, Mathis G, Kirkpatrick AW et al (2012) International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med 38:577–591

    Article  PubMed  Google Scholar 

  67. Mayo P, Volpicelli G, Lerolle N, Schreiber A, Doelken P, Vieillard-Baron A (2016) Ultrasonography evaluation during the weaning process: the heart, the diaphragm, the pleura and the lung. Intensive Care Med 42:1107–1117

    Article  CAS  PubMed  Google Scholar 

  68. De Backer D, Creteur J, Preiser JC, Dubois MJ, Vincent JL (2002) Microvascular blood flow is altered in patients with sepsis. Am J Respir Crit Care Med 166:98–104

    Article  PubMed  Google Scholar 

  69. den Uil CA, Lagrand WK, van der Martin E, Jewbali LS, Cheng JM, Spronk PE et al (2010) Impaired microcirculation predicts poor outcome of patients with acute myocardial infarction complicated by cardiogenic shock. Eur Heart J 31:3032–3039

    Article  Google Scholar 

  70. Edul VS, Enrico C, Laviolle B, Vazquez AR, Ince C, Dubin A (2012) Quantitative assessment of the microcirculation in healthy volunteers and in patients with septic shock. Crit Care Med 40:1443–1448

    Article  PubMed  Google Scholar 

  71. De Backer D, Donadello K, Sakr Y, Ospina-Tascon GA, Salgado DR, Scolletta S et al (2013) Microcirculatory alterations in patients with severe sepsis: impact of time of assessment and relationship with outcome. Crit Care Med 41:791–799

    Article  CAS  PubMed  Google Scholar 

  72. Sakr Y, Dubois MJ, De Backer D, Creteur J, Vincent JL (2004) Persistant microvasculatory alterations are associated with organ failure and death in patients with septic shock. Crit Care Med 32:1825–1831

    Article  PubMed  Google Scholar 

  73. De Backer D, Donadello K, Taccone FS, Ospina-Tascon G, Salgado D, Vincent JL (2011) Microcirculatory alterations: potential mechanisms and implications for therapy. Ann Intensive Care 1:27

    Article  PubMed  PubMed Central  Google Scholar 

  74. Cohen ES, Law WR, Easington CR, Cruz KQ, Nardulli BA, Balk RA et al (2002) Adenosine deaminase inhibition attenuates microvascular dysfunction and improves survival in sepsis. Am J Respir Crit Care Med JID - 9421642 166:16–20

    Article  Google Scholar 

  75. Tachon G, Harrois A, Tanaka S, Kato H, Huet O, Pottecher J et al (2014) Microcirculatory alterations in traumatic hemorrhagic shock. Crit Care Med 42:1433–1441

    Article  PubMed  Google Scholar 

  76. De Backer D, Hollenberg S, Boerma C, Goedhart P, Buchele G, Ospina-Tascon G et al (2007) How to evaluate the microcirculation: report of a round table conference. Crit Care 11:R101

    Article  PubMed  PubMed Central  Google Scholar 

  77. Ince C, Boerma EC, Cecconi M, De Backer D, Shapiro NI, Duranteau J et al (2018) Second consensus on the assessment of sublingual microcirculation in critically ill patients: results from a task force of the European Society of Intensive Care Medicine. Intensive Care Med 44:281–299

    Article  PubMed  Google Scholar 

  78. Ospina-Tascon G, Neves AP, Occhipinti G, Donadello K, Buchele G, Simion D et al (2010) Effects of fluids on microvascular perfusion in patients with severe sepsis. Intensive Care Med 36:949–955

    Article  PubMed  Google Scholar 

  79. Pottecher J, Deruddre S, Teboul JL, Georger J, Laplace C, Benhamou D et al (2010) Both passive leg raising and intravascular volume expansion improve sublingual microcirculatory perfusion in severe sepsis and septic shock patients. Intensive Care Med 36:1867–1874

    Article  PubMed  Google Scholar 

  80. De Backer D, Creteur J, Dubois MJ, Sakr Y, Koch M, Verdant C et al (2006) The effects of dobutamine on microcirculatory alterations in patients with septic shock are independent of its systemic effects. Crit Care Med 34:403–408

    Article  PubMed  Google Scholar 

  81. Rady MY, Nightingale P, Little RA, Edwards JD (1992) Shock index: a re-evaluation in acute circulatory failure. Resuscitation 23:227–234

    Article  CAS  PubMed  Google Scholar 

  82. Andrews B, Muchemwa L, Kelly P, Lakhi S, Heimburger DC, Bernard GR (2014) Simplified severe sepsis protocol: a randomized controlled trial of modified early goal-directed therapy in Zambia. Crit Care Med 42:2315–2324

    Article  PubMed  PubMed Central  Google Scholar 

  83. Michard F, Boussat S, Chemla D, Anguel N, Mercat A, Lecarpentier Y et al (2000) Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med 162:134–138

    Article  CAS  PubMed  Google Scholar 

  84. De Backer D, Heenen S, Piagnerelli M (2005) koch M, Vincent JL: pulse pressure variations to predict fluid responsiveness: influence of tidal volume. Intensive Care Med 31:517–523

    Article  PubMed  Google Scholar 

  85. Myatra SN, Prabu NR, Divatia JV, Monnet X, Kulkarni AP, Teboul JL (2017) The changes in pulse pressure variation or stroke volume variation after a “tidal volume challenge” reliably predict fluid responsiveness during low tidal volume ventilation. Crit Care Med 45:415–421

    Article  PubMed  Google Scholar 

  86. Chu H, Wang Y, Sun Y, Wang G (2016) Accuracy of pleth variability index to predict fluid responsiveness in mechanically ventilated patients: a systematic review and meta-analysis. J Clin Monit Comput 30:265–274

    Article  PubMed  Google Scholar 

  87. Monnet X, Guerin L, Jozwiak M, Bataille A, Julien F, Richard C et al (2012) Pleth variability index is a weak predictor of fluid responsiveness in patients receiving norepinephrine. Br J Anaesth 110:207–213

    Article  PubMed  Google Scholar 

  88. Mahjoub Y, Pila C, Friggeri A, Zogheib E, Lobjoie E, Tinturier F et al (2009) Assessing fluid responsiveness in critically ill patients: false-positive pulse pressure variation is detected by Doppler echocardiographic evaluation of the right ventricle. Crit Care Med 37:2570–2575

    Article  PubMed  Google Scholar 

  89. Monnet X, Bataille A, Magalhaes E, Barrois J, Le Corre M, Gosset C et al (2012) End-tidal carbon dioxide is better than arterial pressure for predicting volume responsiveness by the passive leg raising test. Intensive Care Med 39:93–100

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Daniel De Backer.

Ethics declarations

Conflicts of interest

Daniel De Backer: Consultant to and given material for studies by Edwards Lifesciences. Jan Bakker: No conflict of interest. Maurizio Cecconi: No conflict of interest. Ludhmila Hajjar: No conflict of interest. Da-Wei Liu: No conflict of interest. Suzana Margareth Lobo: No conflict of interest. Xavier Monnet: Consultant to Pulsion/GETINGE, Munich, Germany. Andrea Morelli: No conflict of interest. Sheila Nainan Myatra: No conflict of interest. Azriel Perel: Consultant to Pulsion/GETINGE, Munich, Germany, and to Masimo Inc., Irvine, CA, USA. Michael R. Pinsky: Consultant to Edwards Lifesciences, Cheetah Medical, LiDCO Ltd., honoraria for lectures from Masimo Inc., Edwards LifeSciences, Cheetah Medical, and stock options with LiDCO Ltd., Cheetah Medical. Bernd Saugel: Collaboration with Pulsion Medical Systems SE (Feldkirchen, Germany) as a member of the medical advisory board and received honoraria for lectures and refunds of travel expenses from Pulsion Medical Systems SE. BS received institutional research grants, unrestricted research grants, and refunds of travel expenses from Tensys Medical Inc. (San Diego, CA, USA). BS received honoraria for lectures and refunds of travel expenses from CNSystems Medizintechnik AG (Graz, Austria). BS received research support from Edwards Lifesciences (Irvine, CA, USA). Jean-Louis Teboul: Consultant to Pulsion/GETINGE, Munich, Germany. Antoine Vieillard-Baron: Received a grant from GSK for conducting clinical research and is a member of its scientific advisory board. Jean Louis Vincent: No conflict of interest.

Rights and permissions

Reprints and Permissions

About this article

Verify currency and authenticity via CrossMark

Cite this article

De Backer, D., Bakker, J., Cecconi, M. et al. Alternatives to the Swan–Ganz catheter. Intensive Care Med 44, 730–741 (2018). https://doi.org/10.1007/s00134-018-5187-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00134-018-5187-8

Keywords

  • Hemodynamic monitoring
  • Cardiac output
  • Tissue perfusion
  • Cardiac failure