Introduction
Continuous monitoring of arterial blood saturation using pulse oximetry has become the standard of care in the ICU. With the proliferation of pulse oximeters, episodic hypoxemia is detected much more commonly than previously suspected. By alerting the clinician to the presence of hypoxemia, pulse oximeters can lead to a more rapid treatment of serious hypoxemia and possibly avoid serious complication. Moreover, pulse oximetry can reduce arterial blood gas analysis and potentially decrease health care costs [1].
Principles of pulse oximetry
Pulse oximeters determine oxygen (O2) saturation by measuring light absorption of arterial blood at two specific wavelengths, 660 nm (red) and 940 nm (infrared) [2]. The ratio of absorbencies at the wavelengths is then calibrated empirically against direct measurements of arterial blood oxygen saturation (SaO2), and the resulting calibration curve is used to generate the pulse oximeter’s estimate of arterial saturation (SpO2). In addition to the digital read-out of O2 saturation, most pulse oximeters display a plethysmographic waveform, which can help clinicians distinguish an artifactual signal from the true signal (Fig. 1).
Common pulsatile signals on a pulse oximeter. Top panel Normal signal showing the sharp waveform with a clear dicrotic notch. Second panel Pulsatile signal during low perfusion showing a typical sine wave. Third panel Pulsatile signal with superimposed noise artifact giving a jagged appearance. Lowest panel Pulsatile signal during motion artifact showing an erratic waveform. (From [1])
The accuracy of commercially available oximeters in critically ill patients has been validated in several studies [3]. Compared with the measurement standard (multi-wavelength CO oximeter), pulse oximeters have a mean difference (bias) of less than 1% and a standard deviation (precision) of less than 2% when SaO2 is 90% or above [4]. While pulse oximetry is accurate in reflecting one-point measurements of SaO2, it does not reliably predict changes in SaO2 [4]. Moreover, the accuracy of pulse oximeters deteriorates when SaO2 falls to 80% or less. In critically ill patients, poor agreement between the oximeter and a CO oximeter has been observed, with bias the ranging from −12% to 18%, and oximetry tends systematically to underestimate SaO2 when it is 80% or less.
Limitations of pulse oximetry
Oximeters have a number of limitations which may lead to inaccurate readings [1]; these are presented below.
Physiological limitations
Oxyhemoglobin dissociation curve
Pulse oximeters measure SaO2, which is physiologically related to arterial oxygen tension (PaO2) according to the oxyhemoglobin dissociation curve. Because the dissociation curve has a sigmoid shape, oximetry is relatively insensitive in detecting the development of hypoxemia in patients with high baseline levels of PaO2.
Limitation in the signal processing
Ambient light
Although pulse oximeters correct for ambient light, falsely low SpO2 readings have been reported with fluorescent and xenon arc surgical lamps. Wrapping the probe with an opaque shield can minimize this effect.
Low perfusion
Pulse oximetry depends on satisfactory arterial perfusion of the skin, and thus low cardiac output, vasoconstriction, or hypothermia can make it difficult for a sensor to distinguish the true signal from background noise. In cardiac surgery patients experiencing hypothermia and poor perfusion, only 2 of 20 oximeters (Criticare CSI 503, Datex Satlite) provide measurements within ±4% of the CO oximeter value.
Motion artifact
The occurrence of motion artifacts continues to be a significant source of error and false alarms. In 235 surgical patients managed in the ICU, 67% of pulse oximeter alarms were false [5]. An innovative technological approach, termed Masimo signal extraction technology, was introduced to extract the true signal from artifact due to noise and low perfusion. When tested in 50 postoperative patients, the pulse oximeter’s alarm frequency was decreased twofold with the new system vs. a conventional oximeter. When tested under conditions of low perfusion and motion, the ability to track changes in SpO2 and reduce nuisance alarms was improved with this technology [3].
Interference from substances
Dyshemoglobins
Pulse oximeters employ only two wavelengths of light and thus can distinguish only two substances, oxyhemoglobin and reduced hemoglobin. Accordingly, elevated carboxyhemoglobin and methemoglobin levels can cause inaccurate oximetry readings [1].
Intravenous dyes
Intravenous dyes such as methylene blue, indocyanine green, and indigo carmine can cause falsely low SpO2 readings, an effect that persists for up to 20 min.
Skin pigmentation and other pigments
Inaccurate oximetry readings have been observed in pigmented patients. In critically ill patients, a bias of more than 4% has been observed to occur more frequently in black (27%) than in white patients (11%). Nail polish, if blue, green, or black, causes inaccurate SpO2 readings; however, mounting the oximeter probe sideways alleviates the problem with nail polish. Acrylic nails do not interfere with readings.
Limited knowledge of technique
Many users have only a limited understanding of pulse oximetry. One survey revealed that 30% of physicians and 93% of nurses thought that the oximeter measured PaO2. A more recent audit demonstrated that less than 50% of nurses and physicians were able to identify that motion artifact, arrhythmias, and nail polish can affect the accuracy of pulse oximeter [6].
Clinical applications
Detection of hypoxemia
With the introduction of pulse oximetry hypoxemia (defined as an SpO2 value less than 90%) is detected more often in critically ill patients. Moreover, myocardial ischemia (defined as angina or ST segment depression) in postoperative patients is less common in patients monitored with pulse oximetry than those without oximetry [7].
Assessing pulmonary gas exchange
Pulse oximeters measure SaO2, which is physiologically related to PaO2. In critically ill patients receiving mechanical ventilation, changes in SpO2 may not accurately reflect changes in PaO2 and may in fact be in an opposite direction to the change in PO2. Decisions in therapy made on the basis of SpO2 alone can also differ from those based on PO2. Accordingly, caution is required when making decisions in critically ill patients based solely on pulse oximetry. While pulse oximetry is a suitable way of measuring arterial oxygenation, it does not assess ventilation. Indeed, measurements of SpO2 have been shown to be inaccurate in assessing abnormal pulmonary gas exchange, defined as an elevated alveolar-arterial O2 difference [1].
Titration of fractional inspired oxygen concentration
Pulse oximetry can assist with titration of fractional inspired oxygen concentration (FIO2) in ventilator-dependent patients, although the appropriate SpO2 target depends on a patient’s pigmentation. In white patients, an SpO2 target value of 92% predicts a satisfactory level of oxygenation, whereas black patients required an SpO2 target of 95%. In patients with severe acute respiratory distress syndrome, an SpO2 target of 88–90% is acceptable in order to minimize oxygen toxicity.
Blood pressure measurements
In pulse oximeters that display a pulsatile waveform, systolic blood pressure can be measured by noting the reappearance of the pulsatile waveform during cuff deflation or the waveform disappearance during slow cuff inflation (Fig. 1). In healthy volunteers, good agreement (i.e., bias <1.0 mmHg) was obtained when the average of oximetry based-systolic pressure estimates at the disappearance and reappearance of the waveform were compared with Korotokoff sound pressures and noninvasive equipment blood pressures.
Cardiopulmonary arrest
The usefulness of pulse oximetry as part of the first-line resuscitation equipment at the site of a cardiopulmonary arrest was assessed in 20 patients [8]. A signal in which the pulse rate on the oximeter was correlated with the electrocardiogram or chest compression rate was observed in the three patients who suffered only a respiratory arrest and in only 4 of 17 patients who suffered a cardiac arrest. The physicians judged the pulse oximeter was to be of definite benefit in the management of 7 of 20 patients, 5 of whom survived.
Screening test for cardiopulmonary disease
The potential usefulness of pulse oximetry as a screening tool for cardiopulmonary disease that could supplement or supplant respiratory rate as a “pulmonary vital sign” was investigated in patients managed in the emergency department [9]. An inverse but weak relationship (correlation coefficient −0.16) was observed between SpO2 and respiratory rate. Overall only one-third of patients with an SpO2 value below 90% would exhibit an increase in respiratory rate. While pulse oximetry could be used as a screening tool for cardiopulmonary disease, there are no data to suggest that decisions based on SpO2 improve outcome over decisions based on respiratory rate.
Screening for respiratory failure in asthma
Pulse oximetry has been evaluated as a means of screening for respiratory failure in patients with severe asthma [10]. Respiratory failure occurred in only 4% of the patients with an SaO2 value higher than 92%. The investigators concluded that an SpO2 higher than 92% in this setting suggests that respiratory failure is unlikely and therefore arterial blood gas measurements are unnecessary. Interestingly, this threshold value of 92% is the same target value that predicted reliably a satisfactory level of oxygenation during titration of FIO2 in ventilator-dependent patients.
Pulmonary embolus
In patients with documented pulmonary embolism the room air SpO2 level may be an important predictor of death; mortality was found in one study to be 2% in patients with pulse oximetry of 95% or higher vs. 20% with pulse oximetry less than 95% [11]. When the threshold value was prospectively evaluated in 119 patients, 10 of whom developed hospital complications, SpO2 less than 95% had a sensitivity of 90%, specificity of 64%, and overall diagnostic accuracy of 67%. Although the number of patients with complications were low, these data suggest that pulse oximetry may be useful in predicting outcome in patients with pulmonary embolus.
In summary, pulse oximetry is probably one of the most important advances in respiratory monitoring. The major challenge facing pulse oximetry is whether this technology can be incorporated effectively into diagnostic and management algorithms that improve the efficiency of clinical management in the ICU.
References
Jubran A (1998) Pulse oximetry. In: Tobin MJ (ed) Principles and practice of intensive care monitoring. McGraw-Hill, New York, pp 261–287
Wukitisch MW, Peterson MT, Tobler DR, Pologe JA (1988) Pulse oximetry: analysis of theory, technology, and practice. J Clin Monit 4:290–301
Emergency Care Research Institute (2003) Next-generation pulse oximetry. Health Devices 32:49–103
Van de Louw A, Cracco C, Cerf C, Harf A, Duvaldestin P, Lemaire F, Brochard L (2001) Accuracy of pulse oximetry in the intensive care unit. Intensive Care Med 27:1606–1613
Lutter NO, Urankar S, Kroeber S (2002) False alarm rates of three third-generation pulse oximeters in PACU, ICU and IABP patients. Anesth Analg 94:S69–S75
Howell M (2002) Pulse oximetry: an audit of nursing and medical staff understanding. Br J Nurs 11:91–197
Moller JT, Pedersen T, Rasmussen LS, Jensen PF, Pedersen BD, Ravlo O, Rasmussen NH, Espersen K, Johannessen NW, Cooper JB (1993) Randomized evaluation of pulse oximetry in 20,802 patients. I. Design, demography, pulse oximetry failure rate and overall complication rate. Anesthesiology 78:436–444
Spitall MJ (1993) Evaluation of pulse oximetry during cardiopulmonary resuscitation. Anaesthesia 48:701–703
Mower WR, Sachs C, Nicklin EL, Safa P, Baraff LJ (1996) A comparison of pulse oximetry and respiratory rate in patient screening. Respir Med 90:593–599
Carruthers DM, Harrison BDW (1995) Arterial blood gas analysis or oxygen saturation in the assessment of acute asthma. Thorax 50:186–188
Kline JA, Hernandez-Nino J, Newgard CD, Cowles DN, Jackson RE, Courtney DM (2003) Use of pulse oximetry to predict in-hospital complications in normotensive patients with pulmonary embolism. Am J Med 115:203–208
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Jubran, A. Pulse oximetry. Intensive Care Med 30, 2017–2020 (2004). https://doi.org/10.1007/s00134-004-2399-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00134-004-2399-x
Keywords
- Pulse Oximetry
- Pulse Oximeter
- Severe Acute Respiratory Distress Syndrome
- Oxyhemoglobin Dissociation Curve
- Plethysmographic Waveform