Biomechanics of Olive Wire Positioning and Tensioning Characteristics

  • Original Article
  • Published:
European Journal of Trauma Aims and scope Submit manuscript

Abstract

Objectives:

The effect of the olive wire on the stiffness of external fixation was studied previously, but there is limited data on the effect of the olive wire positioning and tensioning characteristics on the stiffness of external fixation. This study evaluated the influence of olive wire positioning (medially, laterally, posteriorly, and anteriorly) and tensioning characteristics (tensioning one end or both ends) on the stiffness of ring external fixation.

Methods:

A fiberglass composite tibia fixed into an idealized ring external frame was tested. All mechanical testing was performed with a servo-hydraulic test frame (MTS Bionix 858, Minneapolis, MN, USA). Load–deformation behavior was compared among different wire design, olive wire positioning and tensioning characteristics under central axial compression, medial compression-bending, posterior compression-bending, posteromedial compression-bending and torsion. The differences between tension values on opposite ends of the wire during different modalities of olive wire tensioning were also investigated.

Results:

Olive wires tensioned on both ends provided significantly greater bending stiffness compared to smooth wires and olive wires tensioned only on the end opposite to the olive (p < 0.05). There was a significant difference between the tension values on opposite ends of the olive wire during tensioning (p < 0.05).

Conclusions:

Tensioning both ends of the olive wires and positioning the olives on the side of bending significantly increases the bending stiffness of ring external fixation. This work provides a rationale for clinical decision-making about the use of tensioned olive wires as opposed to smooth wires in ring external fixation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Craig S. Roberts MD.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Antoci, V., Voor, M.J., Antoci, V. et al. Biomechanics of Olive Wire Positioning and Tensioning Characteristics. Eur J Trauma 32, 456–463 (2006). https://doi.org/10.1007/s00068-006-5091-8

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00068-006-5091-8

Key Words

Navigation