The Influence of Different IMRT Techniques on the Peripheral Dose

A Comparison between sMLM-IMRT and Helical Tomotherapy

Der Einfluss verschiedener IMRT-Technologien auf die periphere Dosis. Ein Vergleich zwischen sMLM-IMRT und Tomotherapie

Purpose:

To investigate how segmented multileaf modulation-(sMLM-)based intensity-modulated radiotherapy (IMRT) and dynamic helical tomotherapy (ToTh) affect the peripheral dose (PD) outside the treated region.

Material and Methods:

A cuboid Perspex phantom was scanned in a computed tomograph. Different artificial cases were contoured consisting of OARs surrounded by cylindrically shaped planning target volumes (PTVs) with different dimensions. Radiotherapy plans were generated with the sMLM system Konrad (Siemens) and with the ToTh planning system. The plans were optimized in such a way that the dose-volume histograms showed comparable results. The sMLM plans were applied with a linac Primus (Siemens OCS), the ToTh plans with the HiArt system (TomoTherapy); both with 6 MV. Measurements of PDs were performed along the longitudinal axis of the phantom outside the primary beam at different distances from the edge of the PTV (horizontal PD) and also at different depths at a fixed distance from the isocenter (vertical PD). Additional experiments to separate the scatter dose caused by the phantom were performed. This was realized by removing the part of the phantom lying in the primary beam, then applying the same plans like before.

Results:

All PD values were normalized to the median dose of the PTV. The PD values for the different PTVs decrease with decreasing PTV size. They also decrease with increasing distance from the isocenter. The horizontal values are in a range of 7% for the largest PTV (diameter = 15 cm) near the primary dose region to 0.2% for the smallest PTV (diameter = 5 cm) far from the primary dose region. The ToTh values are higher than the sMLM values by a maximal factor of 2 near the primary dose region. They become more similar with increasing distance from the edge of the PTV in longitudinal direction. The PD values are nearly equal at a distance of 25 cm from the edge of the PTV. The vertical PDs are higher for the ToTh at depths of > 1 cm but higher for sMLM close to the surface. By removing the scatter cube, the horizontal PD values at middle distances are reduced to one third of the PD values with scatter cube for ToTh (0.5%) and to one half for sMLM (0.8%). This means that without scatter cube the PD for ToTh is lower than that for sMLM. The measured PD values without scatter cube are in the same dimension as published data.

Conclusion:

The increasing PDs and their trend with increasing PTV size can be explained by Compton scattering of photons from the irradiated volume toward the off-axis measuring points. The further increase of the PD in case of ToTh relative to sMLM is not easy to explain. Different presumptions are possible. The larger field length (in longitudinal direction) of the ToTh plans (consisting of the “real” field length and the overlap) relative to the sMLM plans could be one reason for the higher PD values. The softer energy spectrum of the HiArt machine with more sideward Compton scattering contributions could be another reason.

Ziel:

Im Rahmen eines Vergleichs sollte der Einfluss von intensitätsmodulierter Radiotherapie (IMRT) mit „step-and-shoot“-Technik (sMLM) und von helikaler Tomotherapie (ToTh) auf die periphere Dosis (PD) außerhalb der primären Bestrahlungsregion untersucht werden.

Material und Methodik:

Ein Quaderphantom aus Plexiglas wurde im CT gescannt. Bestrahlungspläne für drei verschieden große, konstruierte Fälle (Risikoorgane umgeben von Hohlzylinder-Planungszielvolumina [PTVs]) erzeugt. Die Pläne wurden für sMLM mit Konrad (Siemens) und für ToTh mit dem HiArt-Planungssystem berechnet. Alle Pläne wurden so optimiert und normalisiert, dass die Dosis-Volumen-Histogramme vergleichbare Ergebnisse zeigten. Die sMLM-Pläne wurden mit einem Beschleuniger Primus (Siemens), die ToTh-Pläne mit einem HiArt-System appliziert, beide mit 6 MV. Dabei wurden PD-Messungen außerhalb des primären Strahlenfelds in verschiedenen Abständen vom PTV (horizontale PD) und in verschiedenen Tiefen bei festem Abstand vom PTV (vertikale PD) durchgeführt. Zur Separierung der Dosisanteile aus dem Phantom und aus dem Beschleunigerkopf bzw. der Strahlkollimiereinheit wurde ein Teil der Experimente ohne Phantom im Primärstrahl wiederholt.

Ergebnisse:

Alle PD-Werte wurden auf die mediane Dosis im PTV normalisiert. Die PD-Werte steigen mit abnehmender PTV-Größe. Sie sinken ebenfalls mit zunehmender Entfernung vom Isozentrum. Die horizontalen PD-Werte liegen im Bereich von 7% in Nähe der Primärregion für das größte PTV (Durchmesser = 15 cm) bis zu 0,2% für das kleinste PTV (Durchmesser = 5 cm) fern der Primärregion. Die PD-Werte sind bei ToTh nahe der Primärregion maximal um den Faktor 2 höher als die sMLM-Werte. Die PD-Werte bei sMLM und ToTh nähern sich mit wachsender longitudinaler Entfernung vom PTV an. Sie sind in 25 cm Entfernung vom PTV nahezu identisch. Die vertikalen PD-Werte sind in Tiefen > 1 cm für ToTh, nahe der Oberfläche jedoch für sMLM höher. Durch Entfernung des Streuquaders wurden die horizontalen PD-Werte in mittlerer Entfernung vom PTV verringert, beim ToTh-Plan auf ein Drittel (0,5%) und beim sMLM-Plan auf die Hälfte (0,8%). Damit sind die aus dem Kollimiersystem stammenden Dosisbeiträge vom HiArt-System geringer als vom „herkömmlichen“ Primus-Beschleunigerkopf mit Multileafkollimatoren. Die gemessenen PDs ohne Streuphantom liegen in der gleichen Größenordnung wie bereits publizierte Daten.

Schlussfolgerung:

Der Anstieg der PD-Werte und deren Zunahme mit der Größe des PTVs können mit der Compton-Streuung der Photonen vom primären Bestrahlungsvolumen in Richtung der Messpunkte erklärt werden. Die weitere Zunahme der PD bei ToTh im Verhältnis zu sMLM ist nicht einfach zu erklären. Die größere Feldlänge bei ToTh (in longitudinaler Richtung), bestehend aus „realer“ Feldlänge und Überlappung (bedingt durch Spiralapplikation), im Vergleich zu herkömmlicher IMRT kann ein Grund sein, das weichere Spektrum des HiArt-Systems mit daraus resultierender stärkerer Compton-Streuung zur Seite ein anderer.

This is a preview of subscription content, access via your institution.

References

  1. Abo Madyan Y, Polednik M, Rahn A, et al. Improving dose homogeneity in large breasts by IMRT. Efficacy and dosimetric accuracy of different techniques. Strahlenther Onkol 2008;184:86–92.

    PubMed  Article  Google Scholar 

  2. American Brachytherapy Society. Guidelines for accelerated partial breast irradiation (http://www.americanbrachytherapy.org/resources/healthapps.cfm).

  3. American Cancer Society (http://www.cancer.org/downloads/STT/2008CAFFfinalsecured.pdf).

  4. American Society of Breast Surgeons. Consensus statement for accelerated partial breast irradiation (http://www.breastsurgeons.org/officialstmts/officialstmt3.shtml).

  5. Gulybán A, Kovács P, Sebestyén Z, et al. Multisegmented tangential breast fields: a rational way to treat breast cancer. Strahlenther Onkol 2008;184:262–9.

    PubMed  Article  Google Scholar 

  6. Kaufmann M, Morrow M, von Minckwitz G, et al., the Biedenkopf Expert Panel Members. Local regional treatment of primary breast cancer: consensus recommendations from an international expert panel. Cancer 2009: in press.

  7. Lohr F, Heggemann F, Papavassiliu T, et al. Is cardiotoxicity still an issue after breast conserving surgery and could it be reduced by multifield IMRT? Strahlenther Onkol 2009;185:222–30.

    PubMed  Article  Google Scholar 

  8. Ludwig V, Schwab F, Guckenberger M, et al. Comparison of wedge versus segmented techniques in whole breast irradiation. Effects on dose exposure outside the treatment volume. Strahlenther Onkol 2008;184:307–12.

    PubMed  Article  Google Scholar 

  9. Mussari S, Sabino Della Sala W, Busana L, et al. Full dose intraoperative radiotherapy with electrons in breast cancer. First report on late toxicity and cosmetic results from a single institution experience. Strahlenther Onkol 2006;182:589–95.

    PubMed  Article  Google Scholar 

  10. Nagel G, Röhrig B, Hoyer H, et al. Bevölkerungsbezogene Studie über die Anwendung adjuvanter Strahlentherapie bei Patientinnen mit Mammakarzinom. Strahlenther Onkol 2002;178:589–96.

    PubMed  Article  Google Scholar 

  11. Nairz O, Deutschmann H, Kopp M, et al. A dosimetric comparison of IORT techniques in limited stage breast cancer. Strahlenther Onkol 2006;182:342–8.

    PubMed  Article  Google Scholar 

  12. NCCN guidelines (http://www.nccn.org/professionals/physician_gls/PDF/breast.pdf).

  13. Ott OJ, Lotter M, Fietkau R, et al. Accelerated partial-breast irradiation with interstitial implants. Analysis of factors affecting cosmetic outcome. Strahlenther Onkol 2009;185:170–6.

    PubMed  Article  Google Scholar 

  14. Ott OJ, Lotter M, Sauer R, et al. Accelerated partial-breast irradiation with interstitial implants. The clinical relevance of the calculation of skin doses. Strahlenther Onkol 2007;183:426–31.

    PubMed  Article  Google Scholar 

  15. Polednik M, Abo Madyan Y, Schneider F, et al. Evaluation of calculation algorithms implemented in different commercial planning systems on an anthropomorphic breast phantom using film dosimetry. Strahlenther Onkol 2007;183:667–72.

    PubMed  Article  Google Scholar 

  16. Polgar C, Fodor J, Major T, et al. Breast conserving treatment with partial or whole breast irradiation for low risk invasive breast carcinoma - 5 year results of a randomized trial. Int J Radiat Oncol Biol Phys 2007;69:694–702.

    PubMed  Google Scholar 

  17. Prosnitz LR, Horton JH, Wallner PE. Accelerated partial breast irradiation: caution and concern from an ASTRO task force. Int J Radiat Oncol Biol Phys 2009;74:981–4.

    PubMed  Google Scholar 

  18. Ribeiro GG, Magee B, Swindell R, et al. The Christie Hospital breast conservation trial: an update at 8 years from inception. Clin Oncol (R Coll Radiol) 1993;5:278–83.

    CAS  Google Scholar 

  19. Sanders ME, Scroggins T, Ampil FL, et al. Accelerated partial breast irradiation in early stage breast cancer. J Clin Oncol 2007;25:996–1002.

    PubMed  Article  Google Scholar 

  20. Sauer R, Budach W, Dunst J, et al. Leitlinie Radiotherapie des Mammakarzinoms. Strahlenther Onkol 2006;182:Suppl 1:1–28.

    Google Scholar 

  21. Sauer R, Sautter-Bihl ML, Budach W, et al. Accelerated partial breast irradiation. Consensus statement of 3 German oncology societies. Cancer 2007;110:1187–93.

    PubMed  Article  Google Scholar 

  22. Sauer R, Wenz F, Haase W, et al. Teilbrustbestrahlung nach brusterhaltenden Operationen wegen Mammakarzinom. Dtsch Ärztebl 2006;103:539–43.

    Google Scholar 

  23. Sauer R, Wenz F, Strnad V, et al. Teilbrustbestrahlung nach brusterhaltender Operation bei Brustkrebs. Senologie 2005;2:77–81.

    Google Scholar 

  24. Sauer R, Wenz F, Strnad V, et al. Teilbrustbestrahlung nach brusterhaltender Operation bei Brustkrebs. Strahlenther Onkol 2005;181:417–23.

    PubMed  Article  Google Scholar 

  25. Sautter-Bihl ML, Souchon R, Budach W, et al. DEGRO practical guidelines for radiotherapy of breast cancer II. Postmastectomy radiotherapy, irradiation of regional lymphatics, and treatment of locally advanced disease. Strahlenther Onkol 2008;184:347–53.

    PubMed  Article  Google Scholar 

  26. Smith BD, Arthur DW, Buchholz TA, et al. Accelerated partial breast irradiation consensus statement from the American Society for Radiation Oncology (ASTRO). Int J Radiat Oncol Biol Phys 2009;74:987–1001.

    PubMed  Google Scholar 

  27. Stranzl H, Zurl B. Postoperative irradiation of left sided breast cancer patients and cardiac toxicity. Does deep inspiration breath hold (DIBH) technique protect the heart? Strahlenther Onkol 2008;184:354–8.

    PubMed  Article  Google Scholar 

  28. Stranzl H, Zurl B, Langsenlehner T, et al. Wide tangential fields including the internal mammary lymph nodes in patients with left sided breast cancer. Influence of respiratory controlled radiotherapy (4D CT) on cardiac exposure. Strahlenther Onkol 2009;185:155–60.

    PubMed  Article  Google Scholar 

  29. Tiefenbacher U, Scheda A, Melchert F, et al. Radiotherapie bei brusterhaltender Operation. Onkologe 2003;9:1023–33.

    Article  Google Scholar 

  30. Vinh Hung V, Truong PT, Janni W, et al. The effect of adjuvant radiotherapy on mortality differs according to primary tumor location in women with node positive breast cancer. Strahlenther Onkol 2009;185:161–8.

    PubMed  Article  Google Scholar 

  31. Whelan T, Levine M. More evidence that locoregional radiation therapy improves survival: what should we do? J Natl Cancer Inst 2005;97:82–4.

    PubMed  Article  Google Scholar 

Download references

Author information

Affiliations

Authors

Corresponding author

Correspondence to Tilo Wiezorek.

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Wiezorek, T., Schwahofer, A. & Schubert, K. The Influence of Different IMRT Techniques on the Peripheral Dose. Strahlenther Onkol 185, 696–702 (2009). https://doi.org/10.1007/s00066-009-2005-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00066-009-2005-9

Key Words:

  • Peripheral dose
  • Intensity-modulated radiotherapy
  • IMRT
  • Tomotherapy

Schlüsselwörter:

  • Periphere Dosis
  • Intensitätsmodulierte Strahlentherapie
  • IMRT
  • Tomotherapie