Abstract
We have developed a miniature γ-ray endoscopic probe consisting of dual BGO detector probes for tumor detection inside the body cavities. The dual detector system was coupled with random coincidence to decrease the distant background radiation and to improve its spatial resolution for tumor localization.Method: The performance of the probe was investigated with a point source and a water phantom. A solution of positron emitting18F isotope was used as the source. Clinical trials of the probe were done to localize tumors on the skin surface of four subjects carrying tumors close to the body surfaces, into whom67Ga-citrate and18F-FDG radiopharmaceuticals were injected.Results: Measurements indicated that the spatial resolution of the dual detector probes is around 1.5 times better than the single detector probe, and both single and dual detector endoscopic probe systems are capable of localizing a tumor on a large photon background.Conclusion: The endoscopic probe may be easier to insert inside body cavities due to the small crystal size and the flexible light guides. A single detector probe with higher sensitivity may be useful in searching for tumors over a wide intracavity area but a dual detector probe can be used for precise tumor localization. The detector probe may also be suitable for intraoperative observation.
Similar content being viewed by others
References
Hinkle GH, Abedel-Nabi H, Miller EA, et al. Radioimmunodetection for implanted tumors with gamma probe.NCI monographs 3: 83–87, 1987.
Hickernell TS, Barber HB, Barrett, Woolfenden JM. Dual detector probe for surgical tumor staging.J Nucl Med 29: 1101–1106, 1988.
Barber HB, Barrett HH, Woolfenden JM, Mayers KJ, Hickernel TS. Comparison ofin vivo probes and gamma cameras for detection of small, deep tumours.Phys Med Biol 34: 727–739, 1989.
Barber HB, Donahue DJ, Woolfenden JM, Silverstein ME. Miniature radiation detector for surgical tumor staging.Proc. 32nd Ann. Conf. on Engineering in Medicine and Biology 21: 142, 1979.
Raymond RR, Richard LW. A Fiber-Optically Coupled Positron-Sensitive Surgical Probe.J Nucl Med 35: 909–913, 1994.
Groch MW, Gottlieb S, Mallon SM, et al. A new dual probe system for the rapid beside assessment of left ventricular function.J Nucl Med 17: 930–936, 1976.
Saffer JR, Barrett HH, Barber HB, Woolfenden JM. Surgical probe design for a coincidence imaging system without a collimator. Information Processing in Medical Imaging, 12th International Conference, IPMI’ 91. Wye, UK, July 1991, Proceedings: 8–22, 1991.
Watabe H, Nakamura T, Takahashi H, Itoh M, Matsumoto M, Yamadera A. Development of a miniature gamma-ray endoscopic probe for tumor localization in nuclear medicine.IEEE Trans Nucl Sci 40: 88–94, 1993.
Briesmeister (Editor), MCNP—A General Monte Carlo Code for Neutron and Photon Transport, Version 3A. Los Alamos National Laboratory report, La-7396-M, Rev 2, 1986.
Wahl RL, Hutchins GD, Mudgett EE. Primary and metastatic breast carcinoma: initial clinical evaluation with PET with radiolabeled glucose analogue 2-[18F]-fluoro-2-deoxy-D-glucose.Radiology 179: 765–770, 1991.
Swinth KL, Ewins JH. Biomedical probe using a fiber-optic coupled scintillator.Med Phys 3: 109–121, 1976.
Woolfenden JM, Barber HB. Radiation detector probes for tumor localization using tumor-seeking radioactive traces.Am J Roentgenol 153: 35–39, 1989.
Martin DT, Hinckle GH, Tuttle S, et al. Intraoperative radioimmuno detection of colorectal tumor with a hand held radiation detector.Am J Surg 150: 672–675, 1985.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Deloar, H.M., Watabe, H., Hayashi, Y. et al. Performance study of a miniature gamma ray scintillationvivo probe for tumor localization. Ann Nucl Med 11, 173–181 (1997). https://doi.org/10.1007/BF03164830
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF03164830