Influence of the lens effect in a sample with large dielectric constant in a loop-gap resonator on the EPR signal intensity at 700 MHz

  • Published:
Applied Magnetic Resonance Aims and scope Submit manuscript

Abstract

The influence of the lens effect on the electron paramagnetic resonance (EPR) signal intensity was investigated in a loop-gap resonator (LGR) with an inner diameter of 41 mm. TheQ- value and EPR signal intensity were measured when the phantoms containing 3-carbamoyl-2,2,5,5-tetramethyl-pyrrolidin-l-yloxy dissolved in sodium chloride aqueous solutions were put in the LGR. TheQ- value and signal intensity reduced with increasing concentrations of sodium chloride in the phantom, indicating that the imaginary part of the dielectric constant is larger in the phantom with the higher concentration of sodium chloride. However, relationships betweenQ-values of the resonator and EPR signal intensities were not proportional and signal intensities were relatively higher compared with theQ-values. These findings suggest that the signal reduction due to lowQ is slightly compensated by the lens effect in the sample with the large real part of the dielectric constant. In the distribution of the signal intensities of a pinpoint sample made of diphenylpicrylhydrazyl in the agar medium containing sodium chloride in the LGR, it was found that the signal intensity decreased according to the distance from the center and the difference in the signal intensity within 10 mm from the center was about 20%, indicating the inhomogeneity of the alternating magnetic field at the center and marginal region in the sample with the large dielectric constant caused by the lens effect.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Froncisz W., Hyde J.S.: J. Magn. Reson.47, 515–521 (1982)

    Google Scholar 

  2. Ono M., Ogata T., Hsieh K., Suzuki M., Yoshida E., Kamada H.: Chem. Lett.1986, 491–494.

  3. Hirata H., Ono M.: Rev. Sci. Instrum.67, 1–5 (1996)

    Article  Google Scholar 

  4. Ishida S., Matsumoto S., Yokoyama H., Mori N., Kumashiro H., Tsuchihashi N., Ogata T., Yamada M., Ono M., Kitajima H., Yoshida E.: Magn. Reson. Imag.10, 21–27 (1992)

    Article  Google Scholar 

  5. Yokoyama H., Ogata T., Tsuchihashi N., Hiramatsu M., Mori N.: Magn. Reson. Imag.14, 559–563 (1996)

    Article  Google Scholar 

  6. Yokoyama H., Fujii S., Yoshimura T., Ohya-Nishiguchi H., Kamada H.: Magn. Reson. Imag.15, 249–253 (1997)

    Article  Google Scholar 

  7. Yokoyama H., Itoh O., Ogata T., Ohya-Nishiguchi H., Kamada H.: Magn. Reson. Imag.15, 1079–1084 (1997)

    Article  Google Scholar 

  8. Yokoyama H., Lin Y., Itoh O., Ueda Y., Nakajima A., Ogata T., Sato T., Ohya-Nishiguchi H., Kamada H.: Free Radic. Biol. Med.27, 442–449 (1999)

    Article  Google Scholar 

  9. Ono M., Hirata H.: International Symposium on in vivo ESR Spectroscopy. In conjunction with 5th Conference on in vivo ESR, p. 5–6. Fukuoka, Japan 2000.

    Google Scholar 

  10. Suenaga A., Ono M., Hirata H., Susaki H.: International Symposium on in vivo ESR Spectroscopy. In conjunction with 5th Conference on in vivo ESR, p. 48. Fukuoka, Japan 2000.

    Google Scholar 

  11. Hyde J.S.: Rev. Sci. Instrum.43, 629–631 (1972)

    Article  ADS  Google Scholar 

  12. Dalai D.P., Eaton S.S., Eaton G.R.: J. Magn. Reson.44, 415–428 (1981)

    Google Scholar 

  13. Sueki M., Rinard G.A., Eaton S.S., Eaton G.R.: J. Magn. Reson.118, 173–188 (1996)

    Article  Google Scholar 

  14. Ito T., Yokoyama H., Ogata T.: Appl. Magn. Reson.20, 257–263 (2001)

    Article  Google Scholar 

  15. Hirata H., Walcak T., Swartz H.M.: Rev. Sci. Instrum.68, 3187–3191 (1997)

    Article  ADS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ito, T., Yokoyama, H., Sato, T. et al. Influence of the lens effect in a sample with large dielectric constant in a loop-gap resonator on the EPR signal intensity at 700 MHz. Appl. Magn. Reson. 21, 97–103 (2001). https://doi.org/10.1007/BF03162444

Download citation

  • Received:

  • Revised:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03162444

Keywords

Navigation