Experience with unipolar pectoral defibrillation

Klinische Erfahrungen mit unipolaren pectoralen Defibrillationssystem

  • Topic Review
  • Published:
Herzschrittmachertherapie und Elektrophysiologie Aims and scope Submit manuscript

Summary

With simple, single lead unipolar pectoral defibrillators, ICD technology has reached a level of ease and safety comparable to pacemaker implantation. It will be difficult to further decrease the morbidity associated with ICD implantation; just as it will be difficult to improve upon current device treatment of sudden cardiac death. Even as further incremental improvements in devices and leads will undoubtedly occur, at this point in ICD evolution, it is investigating the expanded use of this therapy as a prevention tool that is likely to have the largest overall impact on cardiac arrest survival.

Zusammenfassung

Ziel dieses Manuskriptes ist es, über die Wirksamkeit, Komplikationen und den klinischen Verlauf mit dem unipolaren pectoralen Defibrillationssystem zu berichten. Für die Implantation eines unipolaren pectoralen Defibrillationssystem ist eine ausreichende chirurgische Erfahrung erforderlich, um die transvenöse Elektrode in die Spitze des rechten Ventrikels und das Defibrillatorgehäuse in der Loge des Pectoralismuskels zu positionieren. Mit Ausnahme des größeren Elektrodendurchmessers und des größeren Gehäuses ist die pectorale Implantation derzeit vergleichbar mit einer konventionellen Schrittmacherimplantation. Für die Implantation eines Defibrillatorsystems ist es üblich, eine Defibrillationsschwelle zu ermitteln. Die max. verfügbare Energie der derzeitigen Defibrillatoren mit einem Volumen von 70–80 cc beträgt 34 Joule. Eine Defibrillationsschwelle unter 24 Joule ist ein allgemein akzeptiertes Kriterium zur Implantation, wenn man eine Sicherheitsmarge von 10 Joule einhalten möchte. In einer Studie bei 24 Patienten mit einem aktiven Defibrillatorgehäuse mit einem Volumen von 40 cc, 60 cc und 80 cc wurden die folgenden Defibrillationsschwellen ermittelt: 9,5±4,8 J; 8,7±5,8 J und 8,1±4,7 J. Aufgrund dieser Ergebnisse wurde die Schlußfolgerung gezogen, daß bis zu einer Verkleinerung des aktiven Defibrillatorgehäuses auf 40 cc mit keiner Einschränkung der Defibrillationsschwelle gerechnet werden muß. In einer kürzlich veröffentlichten Publikation über 473 Patienten von 74 teilnehmenden Zentren konnten 98% der eingschlossenen Patienten mit einem unipolaren pectoralen Defibrillationssystem versorgt werden. Die mittlere Defibrillationsschwelle betrug in dieser Serie 11,4 J zum Zeitpunkt der Implantation und 9,9 j drei Monate nach Implantation. Von diesen 473 Patienten hatten lediglich 6,1% der Patienten eine Defibrillationsschwelle über 24 J. Trotz dieser Tatsache erhielten 4,2% dieser Patienten dennoch ein unipolares Defibrillationssystem, welches im Verlauf erfolgreich die spontan aufgetretenen Tachyarrhythmien beenden konnte. Eine pectorale Defibrillatorimplantation mit 2 bzw. 3 transvenösen bzw. subkutanen Elektroden wurden bei 1,5% der Patienten durchgeführt. Bei 0,4% der Patienten war eine transvenöse Defibrillatorimplantation aufgrund mangelnder Defibrillationsschwellen nicht möglich. In einer jüngst veröffentlichten Studie wurde kein klinischer Parameter gefunden, der einen Patienten identifizieren konnte, bei dem eine unipolare pectorale Implanatation nicht möglich war.

Operationszeit und Krankenhausaufenthalt

Die durchschnittliche Operationszeit für die Implantation und Testprozedur betrug für das unipolare pectorale Defibrillationssystem 96±45 Min. Diese Operationszeit ist vergleichbar mit der bei einer konventionellen Schrittmacherimplantation und ist signifikant niedriger als die erforderlichen 120 min, die für eine konventionelle transvenöse Defibrillatorimplantation erforderlich ist. Die verkürzte operative Eingriffszeit führt zu einer signifikanten Reduktion der perioperativen Morbidität und somit zu einer vorzeitigen Entlassung des Patienten aus dem Krankenhaus. Die durchschnittliche Verweildauer im Krankenhaus nach Implantation betrug in den US-amerikanischen Zentren 2,6 Tage.Komplikationen: Seit Einführung der transvenösen Defibrillationssystem ist die perioperative Mortalität und Morbidität signifikant gesunken. In der Multizenter-Untersuchung betrug die Komplikationsrate 6% bei den 473 eingeschlossenen Patienten. Die meisten dieser Komplikationen waren elektrodenbezogene Probleme (2%) wie Elektrodendislokation, Oversensing und Elektrodenbruch. Zur Behebung dieser Komplikationen war eine Revisionsoperation erforderlich. In 0,5% der Fälle traten intraoperative Komplikationen wie Pneumothorax bzw. Hämatothorax auf. In der Multizenter-Untersuchung wurden über keine perioperativen Todesfälle berichtet. Ferner traten keine Hirninsulte, Herzinfarkte, Tascheninfektionen bzw.-erusionen bei den 464 erfolgreich implantierten Patienten auf. Diese beachtliche Reduktion der perioperativen Komplikationen ist sicherlich auf die signifikant weniger invasive, kürzere und einfache Implantationsprozedur mit der pectoralen Implantationstechnik zurückzuführen.Überlebensrate: Die Ergebnisse der Multizenter-Studie belegen anhand der Kaplan-Meier-Kurve eine Inzidenz des plötzlichen Herztodes von 1% in einem Verlauf von 18 Monaten. Von den 2061 zu behandelnden ventrikulären Tachyarrhythmien wurden 98,7% erfolgreich in diesem Zeitraum terminiert. Die Gesamtmortalität betrug in dem Zeitraum von 18 Monaten 94,4% auf der Grundlage einer Intention-to-Treat Analyse.Inadäquate Schocks: Inadäquate Schocktherapien wurden in der Multizenter-Studie bei 15% der Patienten beobachtet. Die Mehrheit der inadäquaten Schocktherapien wurden durch Sinustachykardien bzw. Vorhofflimmern verursacht. Durch die zusätzliche Anwendung weiterer Diskriminationsalgorithmen bzw. Implementierung von Zweikammer-Systemen kann die Zahl der inadäquaten Schocktherapien in naher Zukunft sicherlich signifikant reduziert werden.

Schlußfolgerungen

Die transvenöse, unipolare pectorale Defibrillatorim-plantation hat einen hohen technologischen Standard erreicht und ist nahezu so einfach durchzuführen wie eine konventionelle Schrittmacherimplantation. Es wird schwierig sein in naher Zukunft die Morbidität, die mit einer Defibrillatorimplantation verbunden ist, weiter zu senken. Genauso schwierig wird es sein mit einer zukünftigen Gerätegeneration die Überlebenskurve für den plötzlichen Herztod zu verbessern. Nach 17 Jahren der Defibrillatorimplantation ist die Elektrotherapie von klinischen Experiment zur Standardtherapie für Patienten mit lebensbedrohlichen Herzrhythmusstörungen geworden. Zukünftige Studien werden zeigen, ob diese Elektrotherapaie auch für prophylaktische Indikationen geeignet sein wird.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Mirowski M, Reid PR, Mower MM, Watkins L, Gott VL, Schauble JF, Langer A, Heilman MS, Kolenik SA, Fischell RE, Weisfeldt ML (1980) Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N Engl J Med 303:322–324

    PubMed  CAS  Google Scholar 

  2. Marchlinski FE, Flores BT, Buxton AE, Hargrove WCD, Addonizio VP, Stephenson LW, Harken AH, Doherty JU, Grogan EW, Jr., Josephson ME (1986) The automatic implantable cardioverter-defibrillator: efficacy, complications, and device failures. Ann Intern Med 104:481–488

    PubMed  CAS  Google Scholar 

  3. Kelly PA, Cannom DS, Garan H, Mirabal GS, Harthorne JW, Hurvitz RJ, Vlahakes GJ, Jacobs ML, Ilvento JP, Buckley MJ, Ruskin JN (1988) The automatic implantable cardioverter-defibrillator: efficacy, complications and survival in patients with malignant ventricular arrhythmias. J Am Coll Cardiol 11:1278–1286

    PubMed  CAS  Google Scholar 

  4. Borbola J, Denes P, Ezri MD, Hauser RG, Serry C, Goldin MD (1988) The automatic implantable cardioverter-defibrillator. Clinical experience, complications, and follow-up in 25 patients. Arch Intern Med 148:70–76

    Article  PubMed  CAS  Google Scholar 

  5. Gartman DM, Bardy GH, Allen MD, Misbach GA, Ivey TD (1990) Shortterm morbidity and mortality of implantation of automatic implantable cardioverter-defibrillator implantable cardioverter-defibrillator [see comments]. J Thorac Cardiovasc Surg 100: 353–357

    PubMed  CAS  Google Scholar 

  6. Kim SG, Fisher JD, Furman S, Gross J, Zilo P, Roth JA, Ferrick KJ, Brodman R (1991) Exacerbation of ventricular arrhythmias during the postoperative period after implantation of an automatic defibrillator. J Am Coll Cardiol 18:1200–1206

    PubMed  CAS  Google Scholar 

  7. Kuppermann M, Luce BR, McGovern B, Podrid PJ, Bigger JT, Jr., Ruskin JN (1990) An analysis of the cost effectiveness of the implantable defibrillator. Circulation 81:91–100

    PubMed  CAS  Google Scholar 

  8. O’Donoghue S, Platia EV, Brooks-Robinson S, Mispireta L (1990) Automatic implantable cardioverter-defibrillator: is early implantation cost-effective? [see comments]. J Am Coll Cardiol 16: 1258–1263

    PubMed  Google Scholar 

  9. Camphell RW (1990) Life at a price: the implantable defibrillator [editorial] [see comments]. Br Heart J 64:171–173

    Article  Google Scholar 

  10. Yee R, Klein GJ, Guiraudon GM, Jones DL, Sharma AD, Norris C (1990) Initial clinical experience with the pacemaker-cardioverter-defibrillator. Can J Cardiol 6:147–156.

    PubMed  CAS  Google Scholar 

  11. Fromer M, Schlapfer J, Fischer A, Kappenberger L (1991) Experience with a new implantable pacer-, cardioverter-defibrillator for the therapy of recurrent sustained ventricular tachyarrhythmias: a step toward a universal ventricular tachyarrhythmia control device. Pace Pacing Clin Electrophysiol 14:1288–1298

    Article  PubMed  CAS  Google Scholar 

  12. Bardy GH, Hofer B, Johnson G, Kudenchuk PJ, Poole JE, Dolack GL, Gleva M, Mitchell R, Kelso D (1993) Implantable transvenous cardioverter-defibrillators [see comments]. Circulation 87:1152–1168

    PubMed  CAS  Google Scholar 

  13. Saksena S, Parsonnet V (1988) Implantation of a cardioverter/defibrillator without thoracotomy using a triple electrode system. Jama 259:69–72

    Article  PubMed  CAS  Google Scholar 

  14. McCowan R, Maloney J, Wilkoff B, Simmons T, Khoury D, McAlister H, Morant V, Castle L (1991) Automatic implantable cardioverter-defibrillator implantation without thoracotomy using an endocardial and submuscular patch system. J Am Coll Cardiol 17: 415–421

    PubMed  CAS  Google Scholar 

  15. Saksena C, Mehta D, Krol RB, Tullo NG, Saxena A, Kaushik R, Neglia J (1991) Experience with a third-generation implantable cardioverter-defibrillator. Am J Cardiol 67:1375–1384

    Article  PubMed  CAS  Google Scholar 

  16. Bardy GH, Allen MD, Mehra R, Johnson G (1990) An effective and adaptable transvenous defibrillation system using the coronary sinus in humans. J Am Coll Cardiol 16:887–895

    Article  PubMed  CAS  Google Scholar 

  17. Bardy GH, Allen MD, Mehra R, Johnson G, Feldman S, Greene HL, Ivey TD (1990) Transvenous defibrillation in humans via the coronary sinus. Circulation 81:1252–1259

    PubMed  CAS  Google Scholar 

  18. Bardy GH, Troutman C, Poole JE, Kudenchuk PJ, Dolack GL, Johnson G, Hofer B (1992) Clinical experience with a tiered-therapy, multiprogrammable antiarrhythmia device. Circulation 85: 1689–1698

    PubMed  CAS  Google Scholar 

  19. Yee R, Klein GJ, Leitch JW, Guiraudon GM, Guiraudon CM, Jones DL, Norris C (1992) A permanent transvenous lead system for an implantable pacemaker cardioverter-defibrillator. Nonthoracotomy approach to implantation. Circulation 85:196–204

    PubMed  CAS  Google Scholar 

  20. Raitt MH, Johnson G, Dolack GL, Poole JE, Kudenchuk PJ, Bardy GH (1995) Clinical predictors of the defibrillation threshold with the unipolar implantable defibrillation system. J Am Coll Cardiol 25:1576–1583

    Article  PubMed  CAS  Google Scholar 

  21. Jones GK, Poole JE, Kudenchuk PJ, Dolack GL, Johnson G, DeGroot P, Gleva MJ, Raitt M, Bardy GH (1995) A prospective randomized evaluation of implantable cardioverter-defibrillator size on unipolar defibrillation system efficacy. Circulation 92:2940–2943

    PubMed  CAS  Google Scholar 

  22. Bardy GH, Johnson G, Poole JE, Dolack GL, Kudenchuk PJ, Kelso D, Mitchell R, Mehra R, Hofer B (1993) A simplified, single-lead unipolar transvenous cardioversion-defibrillation system. Circulation 88:543–547

    PubMed  CAS  Google Scholar 

  23. Bardy GH, Yee Y, Jung W, for the Active Can Investigators (1996) Multicenter Experience With a Pectoral Unipolar Implantable Cardioverter-Defibrillator. JACC 28:400–410

    PubMed  CAS  Google Scholar 

  24. Strickberger SA, Hummel JD, Daoud E, Niebauer M, Williamson BD, Man KC, Horwood L, Schmittou A, Kalbfleisch SJ, Langberg JJ (1994) Implantation by electrophysiologists of 100 consecutive cardioverter defibrillators with nonthoracotomy lead systems. Circulation 90:868–872

    PubMed  CAS  Google Scholar 

  25. McDaniel WC, Schuder JC (1987) The cardiac ventricular defibrillation threshold: inherent limitations in its application and interpretation. Med Instrum 21:170–176

    PubMed  CAS  Google Scholar 

  26. Malkin RA, Burdick DS, Johnson EE, Pilkington TC, Swanson DK, Ideker RE (1993) Estimating the 95% effective defibrillation dose. IEEE Trans Biomed Eng 40:256–265

    Article  PubMed  CAS  Google Scholar 

  27. Clinical outcome of patients with malignant ventricular tachyarrhythmias and a multiprogrammable implantable cardioverter-defibrillator implanted with or without thoracotomy: an international multicenter study. PCD Investigator Group. J Am Coll Cardiol 1994; 23:1521–1530

    Article  PubMed  Google Scholar 

  28. Zipes DP, Roberts D (1995) Results of the international study of the implantable pacemaker cardioverter-defibrillator. A comparison of epicardial and endocardial lead systems. The Pacemaker-Cardioverter-Defibrillator Investigators. Circulation 92:59–65

    PubMed  CAS  Google Scholar 

  29. Grimm W, Flores BF, Marchlinski FE (1993) Complications of implantable cardioverter defibrillator therapy: follow-up of 241 patients. Pace Pacing Clin Electrophysiol 16:218–222

    Article  PubMed  CAS  Google Scholar 

  30. Winkle RA, Mead RH, Ruder MA, Gaudiani VA, Smith NA, Buch WS, Schmidt P, Shipman T (1989) Longterm outcome with the automatic implantable cardioverter-defibrillator. J Am Coll Cardiol 13:1353–1361

    Article  PubMed  CAS  Google Scholar 

  31. Manolis AS, Tan-DeGuzman W, Lee MA, Rastegar H, Haffajee CI, Huang SK, Estes NA (1989) Clinical experience in seventy-seven patients with the automatic implantable cardioverter defibrillator. Am Heart J 118:445–450

    Article  PubMed  CAS  Google Scholar 

  32. Echt DS, Armstrong K, Schmidt P, Oyer PE, Stinson EB, Winkle RA (1985) Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator. Circulation 71: 289–296

    PubMed  CAS  Google Scholar 

  33. Lehmann MH, Steinman RT, Schuger CD, Jackson K (1988) The automatic implantable cardioverter defibrillator as antiarrhythmic treatment modality of choice for survivors of cardiac arrest unrelated to acute myocardial infarction. Am J Cardiol 62:803–805

    Article  PubMed  CAS  Google Scholar 

  34. Tchou PJ, Kadri N, Anderson J, Caceres JA, Jazayeri M, Akhtar M (1988) Automatic implantable cardioverter defibrillators and survival of patients with left ventricular dysfunction and malignant ventricular arrhythmias. Ann Intern Med 109:529–534

    PubMed  CAS  Google Scholar 

  35. Fogoros RN, Elson JJ, Bonnet CA, Fiedler SB, Burkholder JA (1990) Efficacy of the automatic implantable cardioverter-defibrillator in prolonging survival in patients with severe underlying cardiac disease. J Am Coll Cardiol 16:381–386

    Article  PubMed  CAS  Google Scholar 

  36. Akhtar M, Avitall B, Jazayeri M, Tchou P, Troup P, Sra J, Axtell K (1992) Role of implantable cardioverter defibrillator therapy in the management of high-risk patients. Circulation 85 (1 Suppl):I131-I139

    PubMed  CAS  Google Scholar 

  37. Gottlieb CD, Schwartzman D, Callans D, Sarter B, Dillan SM, Marchlinski FE (1996) Effects of high and low shock energies on sinus electrograms recorded via integrated and true bipolar nonthoracotomy lead systems. J Cardiovasc Electrophysiol 7:189–196

    Article  PubMed  CAS  Google Scholar 

  38. Marchlinski FE, Callans DJ, Gottlieb CD, Schwartzman D, Preminger M (1995) Benefits and lessons learned from stored electrogram information in implantable defibrillators. J Cardiovasc Electrophysiol 6:832–851

    Article  PubMed  CAS  Google Scholar 

  39. Marchlinski FE, Gottlieb CD, Sarter B, Finkle J, Hook B, Callans D, Schwartzman D (1993) ICD data storage: value in arrhythmia management. Pacing Clin Electrophysiol 16:527–534

    Article  PubMed  CAS  Google Scholar 

  40. Gold MR, Shorofsky SR (1996) Transvenous defibrillation lead systems. J Cardiovasc Electrophysiol 7(6): 570–580

    Article  PubMed  CAS  Google Scholar 

  41. Gold MR, Foster AH, Shorofsky SR (1996) Effects of an Active Pectoral-Pulse Generator Shell on Defibrillation Efficacy with a Transvenous Lead System. Am J Cardiol 78:540–543

    Article  PubMed  CAS  Google Scholar 

  42. Haffajee C, Martin D, Bhandari A, Bardy GH, DeSouza C, Krehlkamp V, Church T, for the Active Can Investigators (1996) A Multicenter, Randomized Trial Comparing an Active Can Implantable Defibrillator to a Passive Can. PACE (In Press)

  43. Marks ML, Johnson G, Hofer BO, Bardy GH (1994) Biphasic waveform defibrillation using a three-electrode transvenous lead system in humans. J Cardiovasc Electrophysiol 5(2): 103–108

    Article  PubMed  CAS  Google Scholar 

  44. Bardy GH, Dolack GL, Kudenchuk PJ, Poole JE, Mehra R, Johnson G (1994) Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode. Circulation 89:1090–1093

    PubMed  CAS  Google Scholar 

  45. Jorgenson DB, Haynor DR, Bardy GH, Kim Y (1995) Computational studies of transthoracic and transvenous defibrillation in a detailed 3-D human thorax model. IEEE Trans Biomed Eng 42: 172–184

    Article  PubMed  CAS  Google Scholar 

  46. Jorgenson DB, Schimpf PH, Shen I, Johnson G, Bardy GH, Haynor DR, Kim Y (1995) Predicting cardiotheracic voltages during high energy shocks: methodology and comparison of experimental to finite element model data. IEEE TRans Biomed Eng 42:559–571.

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Reddy, R.K., Bardy, G.H. Experience with unipolar pectoral defibrillation. Herzschr Elektrophys 8, 32–38 (1997). https://doi.org/10.1007/BF03042475

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF03042475

Key words

Navigation