Summary and Conclusions
Of all the methods available for the serial determination of cardiac output, the one using Cardio-Green has proved the most suitable. Most other methods of cardiac output estimation do not lend themselves readily to many consecutive tests, with the exception perhaps of the ones using radioactive tracers. The suitability of the latter, however, depends upon the availability of highly sensitive counters so that the total dose of radioactive material injected during all these tests does not reach dangerous levels.
Résumé
Nous avons étudié les diverses méthodes à notre disposition actuellement pour mesurer le débit cardiaque et nous avons discuté de ľutilité de chacune pour pratiquer des déterminations en série.
On peut diviser en trois groupes les méthodes de mesurer le débit cardiaque: des méthodes physiques, gazométriques, et de dilution ďindicateur. Les diverses méthodes physiques comprennent ľanalyse du tracé du pouls, la ballistocardio-graphie, la pneumotachygraphie, et ľimpédance cardiographique. Les méthodes gazométriques comprennent la détermination directe de Fick et un certain nombre de procédés indirects de Fick. En ce qui concerne les méthodes de dilution ďindicateur, on emploie soit des substances radioactives, soit des tein-tures. On a fait usage à maintes reprises du Bleu Evans pour déterminer des débits cardiaques, mais son emploi n’est pas pratique pour faire de nombreux tests consécutifs. Le Vert-Cardio, une teinture tricarbocyanine, possède plu-sieurs avantages sur le Bleu Evans dont les plus importants sont de ne pas s’accumuler dans ľorganisme et de ne pas être toxique. On peut donc, en conséquence, chez le même individu, ľemployer à répétition à de courts intervalles. On obtient les courbes de dilution rapidement à ľaide ďun système moniteur oxymètre cuvette, car la courbe dèfinitive ne subit pas ľmfluence des variations de la saturation en oxygène du sang.
Nous avons décrit, en détail, la façon exacte ďem^loyer le Vert-Cardio car elle est devenue notre méthode de choix pour déterminer le débit cardiaque chez ľhumain. Le seule autre méthode convenable est celle de ľusage de matériel radioactif, à la condition ďavoir des compteurs très sensibles pour dépister la radioactivité de facon à ce que la radioactivité totale n’atteigne pas des proportions dangereuses.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Krogh, A., &Lindhard, J. Measurements of Blood Flow through Lungs in Man. Skandinav. Arch. f. Physiol.27 100 (1912).
Osborn, J. J. Rapid Measurement of the pCO2 of the Pulmonary Artery in Man: Clinical Use in the Diagnosis of Congenital Heart Disease in 75 Patients. Proc. 30th Scientific Sessions, Am. Heart Ass. Circulation16: 921 (Oct., 1957).
Veall, NT.;Pearson, J. D., Hanley, T., &Lowe, A. E. A Method for the Determination of Cardiac Output: Pieliminary Report. Proc. 2nd Radioisotope Conference, Oxford, July 19-23, 1954, pp. 183–192. London. Butterworth’s Scientific Publications (1954)
Huff, R. L.;Feller, D. D., Judd, O. J., &Bogardus, G. M. Cardiac Output of Men and Dogs Measured byin vivo Analysis of Iodinated (I131) Human Serum Albumin Circ. Res. 3(6). 564–569: (Nov., 1955).
Etsten, B. E. &Li, T. H. The Determination of Cardiac Output by the Dye Dilution Method: Modifications, Comparison with the Fick Method, and Application during Anaesthesia. Anaesthesiology 15(3) 217–230 (May, 1954).
Hamilton, W. F.;Riley, R. L., Aityah, A. M.;Cournand, A.;Fowell, D. M., HiMMELSTEIN, A.;NOBLE, R.P., REMINGTON, J.W.;Richards, D. W.,jrJr.,WHEELER, N. C.; &Witham, A.C. Comparison of Fick and Dye Injection Methods of Measuring Cardiac Output in Man. Am. J Physiol. 153(2): 309–321 (May, 1948).
Eliasch, H. Pulmonary Circulation at Rest and on Effort in Mitral Stenosis. Scandinav. J. Clm. & Lab. Invest4 (Suppl 4) (1952).
Fox, I.J., Brooker, L.G.S., Heselttne, D.W.;Essex, H.E. &Wood, E.H. A Tri- carbocyanine Dye for Continuous Recording of Dilution Curves in Whole Blood Independent of Variations in Blood Oxygen Saturation. Proc. Staff Meetings Mayo Clinic 32(18): 478–84 (Sept. 4, 1957).