Abstract
Purpose
To report our clinical experience with a new blood processing device which uses ultrafiltration. We assessed safety and efficacy by evaluating: I) the quality and the quantity of intraoperative shed blood processed and reinfused to the patient 2) homologous blood requirements 3) clinical status of the patient post-transfusion.
Methods
With Ethics Committee approval, the ultrafiltration device was used in six consenting patients undergoing major elective spinal surgery. Blood samples for haematology and biochemistry tests were collected from patients post-induction of anaesthesia (baseline). I hr and 24 hr post-autotransfusion. Volumes of blood collected and processed, and all autologous and homologous transfusions were recorded. Patients were assessed postoperatrvely for any adverse effects.
Results
Five patients had donated blood preoperatively. One patient required homologous blood products in addition to autologous blood. In two patients, the filtration cartndge became blocked and required changing midprocessing. No patient sustained device-related complications. One patient had postoperative haematuna which resolved spontaneously within two hours.
Conclusion
The ultrafiltration device was safe and effective in reducing homologous blood requirements in six patients undergoing elective spinal surgery. Further evaluation of the ultrafiltration device will be necessary, especially in view of the blockage of the filtration cartndge.
Résumé
Objectif
Décnre notre expénence clinique avec un nouveau dispositif de traitement du sang fonctionnant par ultrafiltration. Nous avons estimé sa sécunté et son efficacité en évaluant: 1) la qualité et la quantité peropératoires de sang perdu traité et reperfusé au patient 2) les besoins en sang homologue 3) l’état posttransfusionnel du patient.
Méthodes
Avec l’accord du comité d’éthique, nous avons utilisé le dispositif d’ultrafiltration chez six patients consentant soumis à une chirurgie rachidienne élective. Les patients ont fourni des échantillons destinés aux épreuves hématologiques et biochimiques après l’induction de l’anesthésie, une heure et 24 h après l’autotransfusion. Nous avons noté le volume du sang recueilli et traité ainsi que les transfusions homologues. Les effets secondaires ont été évalués en postopératoire
Résultats
Cinq patients avaient donné du sang à la période préopératoire. Un patient a eu besoin de produits sanguins homologues en plus du sang autologue. Chez deux patients, la cartouche filtrante s’est obstruée et a dû être remplacée en cours d’intervention. Aucun patient n’a présenté de complication en rapport avec l’appareil. Un patient a souffert d’une hématurie qui s’est dissipée en moins de deux heures.
Conclusion
Le dispositif à ultrafiltration a rédurt les besoins de sang homologue avec sécurité et efficacité chez six opérés pour une intervention rachidienne non urgente. Une évaluation plus complète du dispositif s’impose spécialement en raison de l’obstruction de la cartouche filtrante.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Weiskopf RB. More on the changing indications for transfusion of blood and blood components during anesthesia (Editorial). Anesthesiology 1996; 84: 498–501.
Bengtsson A, Lisander B. Anaphylatoxin and terminal complement complexes in red cell salvage. Acta Anaesthesiol Scand 1990; 34: 339–41.
A report by the American Society of Anesthesiologists Task Force on blood component therapy. Practice guidelines for blood component therapy. Anesthesiology 1996; 84: 732–47.
Williamson KR, Taswell HF. Intraoperative blood salvage: a review. Transfusion 1991; 31: 662–75.
Keeling MM, Gray LA Jr, Brink MA, Hillerich VK, Bland KI. Intraoperative auto-transfusion. Experience in 725 consecutive cases. Ann Surg 1983; 197: 536–40.
Kelley-Patteson C, Ammar AD, Kelley H. Should the cell saver autotransfusion device be used routinely in all infrarenal abdominal aortic bypass operations? J Vasc Surg 1993; 18: 261–5.
Goulet JA, Bray TJ, Timmerman LA, Benson DR, Bargar WL. Intraoperative autologous transfusion in orthopaedic patients. J Bone Joint Surg Am 1989; 71: 3–8.
Wilson WJ. Intraoperative autologous transfusion in revision total hip arthroplasty. J Bone Joint Surg Am 1989; 71: 8–14.
Flynn JC, Metzger CR, Csencsitz TA. Intraoperative autotransfusion (IAT) in spinal surgery. Spine 1982; 7: 432–5.
Johnson RG, Murphy M, Miller M, et al. Fusions and transfusions. An analysis of blood loss and autologous replacement during lumbar fusions. Spine 1989; 14: 358–62.
Kang Y, Aggarwal S, Virji M, et al. Clinical evaluation of autotransfusion during liver transplantation. Anesth Analg 1991; 72: 94–100.
Jurkovich GJ, Moore EE, Medina G. Autotransfusion in trauma. A pragmatic analysis. Am J Surg 1984; 148: 782–5.
Mah ET, Davis R, Seshadri P, Nyman TLM, Seshadri R. The role of autologous blood transfusion in joint replacement surgery. Anaesth Intensive Care 1995; 23: 472–7.
Author information
Authors and Affiliations
Additional information
This study was financially supported by Davol Inc.
Rights and permissions
About this article
Cite this article
Duffy, C.M., Manninen, P.H., Chung, F. et al. Assessment of a new ultrafiltration blood processing system. Can J Anesth 44, 1204–1207 (1997). https://doi.org/10.1007/BF03013346
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF03013346