Abstract
In 38 adults undergoing cardiac surgery, 4 indirect blood pressure techniques were compared with brachial arterial blood pressure at predetermined intervals before and after cardiopulmonary bypass. Indirect blood pressure measurement techniques included automated oscillometry, manual auscultation, visual onset of oscillation (flicker) and return-to-flow methods. Hemodynamic measurements or calculations included heart rate, cardiac index, stroke volume index, and systemic vascular resistance index. Indirect and intraarterial blood pressure values were compared by simple linear regression by patient and measurement period. Measurement errors (arterial minus indirect blood pressure) were calculated, and stepwise regression assessed the relationship between measurement error and heart rate, cardiac index, stroke volume index, and systemic vascular resistance index. Indirect to intraarterial blood pressure correlation coefficients varied over time, with the strongest correlations often occurring at the first and last measurement periods (preinduction and 60 minutes after cardiopulmonary bypass), particularly for systolic blood pressure. Within-patient correlations between indirect and arterial blood pressure varied widely—they were consistently high or low in some patients. In other patients, correlations were especially weak with a particular indirect blood pressure method for systolic, mean, or diastolic blood pressure; in some cases indirect blood pressure was inadequate for clinical diagnosis of acute blood pressure changes or trends. The mean correlations between indirect and direct blood pressure values were, for systolic blood pressure: 0.69 for oscillometry, 0.77 for auscultation, 0.73 for flicker, and 0.74 for return-to-flow; for mean blood pressure: 0.70 for oscillometry and 0.73 for auscultation; and for diastolic blood pressure: 0.73 for oscillometry and 0.69 for auscultation. The mean measurement errors (arterial minus indirect values) for the individual indirect blood pressure methods were, for systolic: 0 mm Hg for oscillometry, 9 mm Hg for auscultation, -5 mm Hg for flicker, 7 mm Hg for return-to-flow; for mean: -6 mm Hg for oscillometry, and -3 mm Hg for auscultation; and for diastolic: -9 mm Hg for oscillometry and -8 mm Hg for auscultation. Mean measurement error for systolic blood pressure was thus least with automated oscillometry and greatest with manual auscultation, while standard deviations ranging from 9 to 15 mm Hg confirmed the highly variable nature of single indirect blood pressure measurements. Except for oscillometric diastolic blood pressure, a combination of systemic hemodynamics (heart rate, stroke volume index, systemic vascular resistance index, and cardiac index) correlated with each indirect blood pressure measurement error, which suggests that particular numeric ranges of these variables minimize measurement error. This study demonstrates that striking variability occurs in the relationship between indirect and arterial blood pressure measurements, and that the systemic hemodynamic state influences accuracy of indirect blood pressure measurements. When the reproducibility of repeated indirect blood pressure measurements appears unsatisfactory or inconsistent with other clinical observations, clinicians may find that an alternative indirect blood pressure method is a better choice. Of the methods tested, no single indirect blood pressure technique showed precision superior to the others, but two methods yielded data only for systolic pressure. These findings lend support to intraarterial blood pressure measurement in conditions of hemodynamic variability, and suggest the theoretical benefits of continuous indirect blood pressure measurements.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Borow KM, Newburger JW. Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow: comparative study of indirect systolic, diastolic, and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurements. Am Heart J 1982;103:879–886
Cohn JN. Blood pressure measurement in shock. Mechanism of inaccuracy in auscultatory and palpatory methods. JAMA 1967;199:118–122
Chyun DA. A comparison of intra-arterial and auscultatory blood pressure readings. Heart Lung 1985, 14:223–228
Finnie KJC, Watts DG, Armstrong PW. Biases in the measurement of arterial pressure. Crit Care Med 1984; 12:965–968
Whalen P, Ream AK. A quantitative evaluation of the Hewlett-Packard 78354A noninvasive blood pressure meter. J Clin Monit 1988;4:21–30
Green M, Paulus DA, Roan VP, van der Aa J. Comparison between oscillometric and invasive blood pressure monitoring during cardiac surgery. Int J Clin Monit Comput 1984;1:21–26
Snyder NC, Lichtor JL, Schreider B. Evaluation of an automatic blood pressure monitor in children. Anesthesiology 1984;61:A174
van Egmond J, Hasenbos M, Crul JF. Invasive v. non-invasive measurement of arterial pressure: comparison of two automatic methods and simultaneously measured direct intra-arterial pressure. Br J Anaesth 1985;57:434–444
Gourdeau M, Martin R, Lamarche Y, Tetreault L. Oscillometry and direct blood pressure: a comparative clinical study during deliberate hypotension. Can Anaesth Soc J 1986;33:300–307
Yelderman M, Ream AK. Indirect measurement of mean blood pressure in the anesthetized patient. Anesthesiology 1979;50:253–256
Hutton P, Dye J, Prys-Roberts C. An assessment of the Dinamap 845. Anaesthesia 1984;39:261–267
Bruner JMR, Krenis LJ, Kunsman JM, Sherman AP. Comparison of direct and indirect methods of measuring arterial blood pressure. Med Instrum 1981;15:11–21, 97–101, 182–188
Kirkendall WM, Feinleib M, Freis ED, Mark AL. Recommendations for human blood pressure determination by sphygmomanometers: subcommittee of the AHA Post-graduate Education Committee. Stroke 1981;4:555A-564A
Gardner RM. Direct blood pressure measurement— dynamic response requirements. Anesthesiology 1981;54: 227–236
Wood EH. Physical response requirements of pressure transducers for the reproduction of physiological phenomena. Trans Am Institute Electrical Engineers [part I] 1956;75:32–40
Rothe CF, Kim KC. Measuring systolic arterial blood pressure: possible errors from extension tubes or disposable transducer domes. Crit Care Med 1980;8:683–689
Gravlee GP, Wong AB, Adkins TG, et al. A comparison of radial, brachial, and aortic pressures after cardiopulmonary bypass. J Cardiothorac Anesth 1989;3:20–26
Runciman WB, Rutten AJ, Ilsley AH. An evaluation of blood pressure measurement. Anaesth Intesive Care 1981; 9:314–325
O’Callaghan WG, Fitzgerald DJ, O’Malley K, O’Brien E. Accuracy of indirect blood pressure measurement in the elderly. Br Med J 1983, 286:1545–1546
Park MK, Robotham JL, German VF. Systolic pressure amplification in pedal arteries in children. Crit Care Med 1983;11:286–289
Park MK, Menard SM. Accuracy of blood pressure measurement by the Dinamap monitor in infants and children. Pediatrics 1987;79:907–914
Gloyna DF, Huber P, Abston P, Arens JF. A comparison of blood pressure measurement techniques in the hypotensive patient (abstr). Anesth Analg 1984;63:222
Breit SN, O’Rourke MF. Comparison of direct and indirect arterial pressure measurements in hospitalized patients. Aust N Z J Med 1974;4:485–491
Davis RF. Clinical comparison of automated auscultatory and oscillometric and catheter-transducer measurements of arterial pressure. J Clin Monit 1985;1:114–119
Pereira E, Prys-Roberts C, Dagnino J, et al. Auscultatory measurement of arterial pressure during anaesthesia: a reassessment of Korotkoff sounds. Eur J Anaesthesiol 1985;2:11–20
Ramsey M. Noninvasive automatic determination of mean arterial pressure. Med Biol Eng Comput 1979; 17: 11–18
Venus B, Mathru M, Smith RA, Pham CG. Direct versus indirect blood pressure measurements in critically ill patients. Heart Lung 1985;14:228–231
Colan SD, Fujii A, Borow KM, et al. Noninvasive determination of systolic, diastolic, and end-systolic blood pressure in neonates, infants and young children: comparison with central aortic pressure measurements. Am J Cardiol 1983;52:867–870
Friesen RH, Lichtor JL. Indirect measurement of blood pressure in neonates and infants utilizing an automatic noninvasive oscillometric monitor. Anesth Analg 1981; 60:742–745
Loubser PG. Comparison of intra-arterial and automated oscillometric blood pressure measurement methods in postoperative hypertensive patients. Med Instrum 1986; 20:255–259
Nystrom E, Reid KH, Bennett R, et al. A comparison of two automated indirect arterial blood pressure meters: with recordings from a radial arterial catheter in anesthetized surgical patients. Anesthesiology 1985;62:526–530
Glantz SA. Primer of biostatistics. New York: McGraw Hill, 1981:210–217
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–310
Vardan S, Mookherjee S, Warner R, Smulyan H. Systolic hypertension. Direct and indirect BP measurements. Arch Intern Med 1983;143:935–938
Hausman D, Rommelsheim K. Automatic non-invasive blood pressure determination-comparative determinations and experiences with the DINAMAP Vital Data 845 XT monitor in the intensive care station. Anasth Intensivther Notfallmed 1983;18:93–97
Holland WW, Humerfelt S. Measurement of blood-pressure: comparison of intra-arterial and cuff values. Br Med J 1964;2:1241–1243
Hunyor SN, Flynn JM, Cochineas C. Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings. Br Med J 1978;2:159–162
Raftery EB, Ward AP. The indirect method of recording blood pressure. Cardiovasc Res 1968;2:210–218
Kimble KJ, Darnall RA, Yelderman M, et al. An automated oscillometric technique for estimating mean arterial pressure in critically ill newborns. Anesthesiology 1981; 54:423–425
Berliner K, Fujiy H, Lee DH, et al. The accuracy of blood pressure determinations: a comparison of direct and indirect measurements. Cardiologia 1960;37:118–128
Rutten AJ, Ilsley AH, Skowronski GA, Runciman WB. A comparative study of the measurement of mean arterial blood pressure using automatic oscillometers, arterial cannulation and auscultation. Anaesth Intensive Care 1986; 14:58–65
Van Bergen FH, Weatherhead DS, Treloar AE, et al. Comparison of indirect and direct methods of measuring arterial blood pressure. Circulation 1954;10:481–490
Stern DH, Gerson JI, Allen FB, Parker FB. Can we trust the direct radial artery pressure immediately following cardiopulmonary bypass? Anesthesiology 1985;62:557–561
Hutton P, Prys-Roberts C. The oscillotonometer in theory and practice. Br J Anaesth 1982;54:581–591
Sy WP. Ulnar nerve palsy possibly related to use of automatically cycled blood pressure cuff. Anesth Analg 1981; 60:687–688
Showman A, Betts EK. Hazard of automatic noninvasive blood pressure monitoring (letter). Anesthesiology 1981; 55:717–718
Author information
Authors and Affiliations
Additional information
Annual meeting of the American Society of Anesthesiologists, New Orleans, LA, Oct 1984.
Rights and permissions
About this article
Cite this article
Gravlee, G.P., Brockschmidt, J.K. Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlations. J Clin Monitor Comput 6, 284–298 (1990). https://doi.org/10.1007/BF02842488
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02842488