Abstract
We present an algorithm that permits automated acquisition, display, and disk storage of single or dual channel sensory evoked potential waveforms using a Pathfinder II evoked response monitor. Once initiated, data acquisition and storage continue at user-specified intervals without additional user input. A cascade screen display provides trend monitoring. The file for data storage is established and searched using standard Pathfinder commands. The algorithm is written in MECOL, the Pathfinder specific programming language. This programming code imposed a number of limitations that had to be overcome to create a versatile and user friendly algorithm. During clinical use in our operating rooms, the program has been found effective and easy to use by both veteran and novice Pathfinder operators. The “handsoff” trend monitoring permitted by this algorithm has reduced the labor-intensive aspect of sensory evoked potential monitoring.
Similar content being viewed by others
References
Grundy BL: Intraoperative monitoring of sensory-evoked potentials. Anesthesiology 1983;58:72–87
Raudzens PA: Intraoperative monitoring of evoked potentials. Ann NY Acad Sci 1982;388:308–326
Author information
Authors and Affiliations
Additional information
The opinions expressed herein are those of the authors and are not to be construed as reflecting the views of the Navy Department, the Naval Service at large, or the Department of Defense.
Rights and permissions
About this article
Cite this article
Schubert, A., Zornow, M.H. & Drummond, J.C. A program designed to facilitate the repetitive acquisition, display, and disk storage of sensory evoked responses. J Clin Monitor Comput 1, 207–212 (1985). https://doi.org/10.1007/BF02832263
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02832263