Abstract
The preparation, principle and measurement results of the complex ISFET with Ag−AgCl reference electrode are presented in this paper. Through experiment and theory we endeavor to show the feasibility of using the Ag−AgCl electrode which is without solution contact as reference electrode and the way of overcoming its instability; it gives a rational explanation for phenomenon of the ion sensitive field effect transistor which does not conform to Nernstian response, when we measured cations of Na+, Ca++, etc. with the Ag−AgCl reference electrode.
Similar content being viewed by others
Reference
S. Shoji, M. Esashi, T. Masuo, Sensors and Actuators,5 (1984), 89–99.
Author information
Authors and Affiliations
About this article
Cite this article
Xingjun, H., Shouqi, W. Study of feasibility of using Ag−AgCl electrode as reference electrode of the isfet. J. of Electron.(China) 7, 285–288 (1990). https://doi.org/10.1007/BF02778432
Issue Date:
DOI: https://doi.org/10.1007/BF02778432