Abstract
A portable optoelectronic reflectometric photocolorimeter was designed. The photocolorimeter implements two operating modes (transmission and reflection), and its performance was tested in two systems of enzyme immunoassay for testosterone: microplate ELISA and membrane dot-ELISA. The detection threshold for microplate ELISA and membrane dot-ELISA was 0.1 and 0.6 ng/ml, respectively. The coefficient of correlation between the photocolorimeter readings and conventional photometric methods is r=0.999. Relative standard deviation of the results of photocolorimetric measurements (n = 4) within the optical density range of 0.03 to 1.00 ranged from 3.4 to 0.7%. The simple and versatile design of the photocolorimeter makes it appropriate for testing various substances both in laboratory settings and in the field.
Similar content being viewed by others
References
Dzantiev, B.B., Zherdev, A.V., Romanenko, O.G., Izumrudov, V.A., and Zezin, A.B.,Advances in Steroid Analysis, Budapest: Akademiai, 1994, pp. 119–125.
Zherdev, A.V., Byzova, N.A., and Dzantiev, B.B.,Anal. Lett. (in press).
Dzantiev, B.B., Choi, M.J., Park, J., Choi, J., Romanenko, O.G., Zherdev, A.V., Eremin, S.A., and Izumrudov, V.A.,Immunol. Lett., 1994, vol. 41, nos. 2-3, pp. 205–211.
Bordier, C. and Ryter, P.,Anal. Biochem., 1986, vol. 152, no. l, pp. 113–118.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Shishkin, Y.L., Zherdev, A.V., Dzantiev, B.B. et al. A portable reflectometric photometer for quantitative enzyme immunoassay. Appl Biochem Microbiol 36, 429–433 (2000). https://doi.org/10.1007/BF02738057
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02738057