Poor resolution and sensitivity are not inherent in thin layer chromatography (TLC). They merely re-flect the fact that TLC, as normally practiced, is not a high performance (efficient) technique. When efforts are made to improve the efficiency of TLC, the re-sulting improvements in resolution and sensitivity make high performance TLC fully comparable to high performance liquid chromatography (HPLC). Two approaches to improved TLC performance are the use of small-particle TLC plates and the technique of Programmed Multiple Development (PMD). Prelimi-nary PMD methods for limonin in citrus juice and for caffeine in cola beverages clearly illustrate the use of PMD.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Jupille, T.H., and Perry, J.A.; Paper No. 213, presented at the 27th Pitts. Conf. on Anal. Chem. and Appl. Spect., (1976).

  2. Snyder, L.R., in “Chromatography: A Laboratory Handbook of Chromatographic and Electrophoretic Methods,” 3rd edition, edited by E. Heftmann, Van Nostrand Reinhold, New York, 1976, p. 73.

    Google Scholar 

  3. Snyder, L.R., “Principles of Adsorption Chromatography,” Dekker, New York, 1968, p. 120.

    Google Scholar 

  4. Riphahn, J., and H. Halpaap, J. Chromatogr. 112:81 (1975).

    Article  Google Scholar 

  5. Jupille, T.H., and J.A. Perry, J. Chromatogr. Sci. 13:163 (1975).

    CAS  Google Scholar 

  6. Jupille, T.H., and J.A. Perry, J. Chromatogr. 99:231 (1974).

    Article  CAS  Google Scholar 

  7. Perry, J.A., T.H. Jupille, and A.B. Curtice, Sept. Sci. 10:571 (1975).

    Article  CAS  Google Scholar 

  8. Thoma, J.A., Anal. Chem. 35:214 (1963).

    Article  CAS  Google Scholar 

  9. Perry, J.A., K.W. Haag, and L.J. Glunz, J. Chromatogr. Sci. 11:447 (1973).

    CAS  Google Scholar 

  10. Jupille, T.H., and A.B. Curtice, Chromatographia 8:193 (1975).

    Article  CAS  Google Scholar 

  11. Maier, V.P., and E.R. Grant, J. Agr. Food Chem. 18:250 (1970).

    Article  CAS  Google Scholar 

  12. Tatum, J.H., and R.E. Berry, J. Food Sci. 38:1244 (1973).

    Article  Google Scholar 

  13. Perry, J.A., and L.J. Glunz, J. Ass. Offic. Anal. Chem. 57:832 (1974).

    Google Scholar 