Abstract
A general theoretical approach for the determination of the optimum pulsewidth used for TENS/EMS/FES is presented. Analysis based on the interaction between a stepup transformer-coupled nerve stimulator and an electrode-tissue load modelled as a simple lossy capacitive load results in excellent agreement between the predicted and measured performance. The analysis shows that, by adjusting the pulsewidth (PW) of a push-pull symmetric square waveform such that PW=4.5τ, the total charge delivered to the tissue load can be minimised without impacting the efficiency of the nerve stimulation. Additionally, by minimising the charge exchange supported portion of the current pulse, which is primarily responsible for the pH shift and subsequent tissue burning with longterm use, the latter can be reduced to an acceptable level.
Similar content being viewed by others
References
AAMI (1986): ‘American national standard for transcutaneous electrical nerve stimulators.’Assoc. Advanc. Med. Instrum., 3330 Washington Blvd., Suite 400, Arlington, VA 22201-4598, USA
Bing-He, Z., Baratta, R. andSolomonow, M. (1987): ‘Manipulation of muscle force with various firing rate and recruitment control strategies,’IEEE Trans.,BME-34, pp. 123–130.
Flanagan, W. M. (1986): ‘Handbook of transformer applications’ (McGraw-Hill, New York) pp. 1.5–1.6
Geddes, L. A. andBourland, J. D. (1985): ‘The strength-duration curve,’IEEE Trans.,BME-32, pp. 301–310
Malezic, M., Stanic, U. andKljajic, M. (1984): ‘Multichannel electrical stimulation of gait in motor disabled patients,’Orthopedics,7, pp. 1187–1195
Milner-Brown, H., Stein, R. B. andYemm, R. (1973): ‘The orderly recruitment of human motor units during voluntary isometric contraction,’J. Physiol. Lond.,230, pp. 339–370.
Petrofsky, J. S. andSmith, J. (1991): ‘A combined computer-controlled walking and exercise system,’J. Clin. Eng.,16, pp. 223–235
Ragheb, T. andGeddes, L. A. (1990): ‘Electrical properties of metallic electrodes,’IEEE Trans.,BME-37, pp. 323–329.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kolen, P.T. Pulsewidth optimisation for transformer-coupled transcutaneous current pulse generation. Med. Biol. Eng. Comput. 36, 471–474 (1998). https://doi.org/10.1007/BF02523216
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02523216