Skip to main content
Log in

A spectral overlap method for self monitoring quench correction in double isotope liquid scintillation counting

  • Published:
Journal of Radioanalytical Chemistry Aims and scope Submit manuscript

Abstract

A method for quench correction of samples with double radioactive labelling is described. Each nuclide makes a contribution to the counting rate of three channels of a liquid scintillation counter. This channel overlap is an essential requirement of the calibration procedure rather than a limitation, and allows more freedom in the choice of counting conditions. After calibration with suitable standards the method will tolerate wide variations in the ratio of one isotope to the other extending to single label samples of either isotope. This is the outstanding advantage over the channel ratio method which requires a statistically significant counting rate for the higher energy isotope. The method takes advantage of the facilities offered by a computer which may be on line or remote.14C and tritium are used to demonstrate the utility of the method.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. R. J. HERBERG, Anal. Chem., 36 (1964) 1079.

    Article  CAS  Google Scholar 

  2. F. N. HAYES, Advances in Tracer Methodology. Vol. 3, Plenum, New York, 1966, p. 95.

    Google Scholar 

  3. D. S. GLASS, Intern. J. Appl. Radiation Isotopes, 21 (1970) 531.

    Article  CAS  Google Scholar 

  4. L. A. BAILLIE, Intern. J. Appl. Radiation Isotopes, 8 (1960) 1.

    Article  CAS  Google Scholar 

  5. C. A. BRUNO, J. E. CHRISTIAN, Anal. Chem., 33 (1960) 650.

    Article  Google Scholar 

  6. E. T. BUSH, Anal. Chem., 35 (1963) 1024.

    Article  CAS  Google Scholar 

  7. J. K. WELTMAN, D. W. TALMAGE, Intern. J. Appl. Radiation Isotopes, 14 (1963) 541.

    Article  CAS  Google Scholar 

  8. R. W. HENDLER, Analyt. Biochem., 7 (1964) 110.

    Article  CAS  Google Scholar 

  9. R. NINOMIYA, Intern. J. Appl. Radiation Isotopes, 17 (1966) 355.

    Article  CAS  Google Scholar 

  10. B. D. CADDOCK, P. T. DAVIES, J. H. DETERDING, Intern. J. Appl. Radiation Isotopes, 18 (1967) 209.

    Article  CAS  Google Scholar 

  11. D. L. HANSEN, C. O. CARROLL, Intern. J. Appl. Radiation Isotopes, 22 (1971) 677.

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ayrey, G., Evans, K.L. A spectral overlap method for self monitoring quench correction in double isotope liquid scintillation counting. J. Radioanal. Chem. 25, 59–73 (1975). https://doi.org/10.1007/BF02519538

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02519538

Keywords

Navigation