Abstract
Procedures for the determination of bromine by the reactions81Br(n, γ)82Br (T=35.4 h) and79Br(n, γ)80mBr (T=4.4 h). In the case of82Br a flat coaxial Ge(Li) crystal is used to measure the 619 keV photopeak. For80mBr a planar Ge(Li) detector is applied to measure the 39 keV γ-ray. The agreement between the data obtained with both techniques for some Standard Reference Materials is satisfactory.
Similar content being viewed by others
References
C. SHENBERG, et al., Anal. Chem., 39 (1967) 780.
R. A. NADKARNI, G. H. MORRISON, Anal. Chem., 45 (1973) 1957.
R. SCHELENZ, J. F. DIEHL, in Accuracy in Trace Analysis, Spec. Publ. NBS No. 422, (1976), Vol. 2, p. 1173.
R. A. NADKARNI, Radiochem. Radioanal. Letters, 30 (1977) 329.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Hoede, D., Zonderhuis, J. & Das, H.A. The determination of bromine in dry biological material by instrumental neutron activation analysis. J. Radioanal. Chem. 56, 199–202 (1980). https://doi.org/10.1007/BF02516951