Abstract
We present the results of tests of five Hygrolog WSX-160 moist gas generators. We note that the generators reproduce five (out of six according to the documentation) values of the dew point (frost point) with the error indicated in the documentation. We consider methods for elimanation of the uncertainty arising at an operating pressure different from normal atmospheric pressure. We indicate additional operating conditions for the generators which assure the claimed technical and metrological characteristics.
Similar content being viewed by others
Additional information
Translated from Izmeritel'naya Tekhnika, No. 11, pp. 65–66, November, 1997.
Rights and permissions
About this article
Cite this article
Dubovikov, N.I., Podmurnaya, O.A. Tests of hygrolog WSX-160 moist gas generators. Meas Tech 40, 1124–1125 (1997). https://doi.org/10.1007/BF02505138