Summary
A multifunctional suppressor for both anion and cation chromatography has been designed. The suppressor comprises five thin chambers—an anion eluent suppressive chamber, a cation eluent suppressive chamber, a cathode chamber, an anode chamber, and a common electrode chamber, all of which are clipped together. An electrochemical process—electrolysis of deionized water or detector effluent—is used to regenerate the suppressor for continuous operation. Two power sources are used to supply current. The device can work as an anion suppressor, a cation suppressor, or as both anion and cation suppressors, with high suppression capacity (60 mmol L−1) and good reproducibility (RSD=0.80–0.91%) and linearity (r=0.9992).
Similar content being viewed by others
References
Small, H.; Stevens, T.S.; Bauman, W.C.Anal. Chem. 1975,47, 1801.
Stevens, T.S.; Davis, J.C.; Small, H.Anal. Chem. 1998,53, 1482.
Franklin, G.O.Am. Lab. 1985,17(6), 65.
Tian, Z.W.; Hu, R.Z.; Lin, H.S.; Wu, J.T.J. Chromatogr. 1988,439, 159.
Rabin, S.; Stillian, J.; Barreto, V.; Friedman, K.; Toofan, M.J. Chromatogr. 1993,640, 97.
Saari-Nardhaus, R.; Anderson Jr, J.M.J. Chromatogr. 1997,782, 75.
Small, H.; Riviello, J.Anal. Chem. 1998,70, 2205.
Saari-Nardhaus, R.; Anderson Jr, J.M.J. Chromatogr. 2002,956, 15.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Hu, R.Z., Weng, Y.H., Lai, L.M. et al. Electrochemically generated multifunctional suppressor for ion chromatography. Chromatographia 57, 471–474 (2003). https://doi.org/10.1007/BF02492543
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02492543